311 numbers

0.0(0)
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
Card Sorting

1/67

flashcard set

Earn XP

Description and Tags

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

68 Terms

1
New cards

5 to 10 degrees but not less than 15 degrees

Target angulation

2
New cards

100%

Voltage ripple of single phase generator

3
New cards

14%

Voltage ripple of 3phase 6pulse generator

4
New cards

4%

Voltage ripple of 3phase 12pulse generator

5
New cards

Less than 1%

Voltage ripple of high frequency generator

6
New cards

Greater than 3 hrs

Rest for xrt

7
New cards

3, 200 rpm and 10, 000 rpm

Rotor/anode rotations

8
New cards

10, 000 rpm

Anode rotation for high capacity tubes

9
New cards

1-2% thorium

Additional for filament

10
New cards

80%

Maximum exposure limit for single exposure

11
New cards

25s

Intervals between spot films

12
New cards

450kg

Floating tables

13
New cards

150kw (150kvp, 1000mA)

Power rating of stationary radiography unit

14
New cards

1 mGy/hr at 1 m

Limit of scatter radiaion

15
New cards

4 thickness of 5/8 gypsum board

Walls of secondary protective barriers

16
New cards

0.04-320 mAs

MAs for mobile units

17
New cards

40-130 kVp

KVp for mobile units

18
New cards

15-25kw

Power rating for mobile units

19
New cards

Less than 1ms

Solid state detector can measure exposure times at

20
New cards

600mAs greater than or equal to 51kVp

Termination of back up time in AEC

21
New cards

50 to 75 times light photons

Photoelectrons arrives in output phosphor

22
New cards

12, 17, 25 cm

Most popular trifield tubes

23
New cards

2.5 or 5cm

Standard size of output phosphor

24
New cards

10-40cm

Input phosphor size

25
New cards

15mm 25cm

Diameter and length of cylindrical housing of television camera

26
New cards

120 films/hr

First automatic processor could process

27
New cards

40 minutes

Total cycle time for processing film in first automatic processor

28
New cards

10 ft 3 tons

Size of first roller transport system

29
New cards

90 seconds

Dry-to-drop time of rapid processing

30
New cards

45 seconds

Processing cycle time for automatic processor introduced by Konica

31
New cards

1 inch

Diameter of transport rollers

32
New cards

3 inches

Diameter of of master/solar rollers

33
New cards

10-20 rpm

Reduced shaft of drive motor

34
New cards

35*C or 95*F

Temperature of developer

35
New cards

3*C or 5*F

Temperature of wash water

36
New cards

12L/minute or 3gal/minute

Wash water in circulation sytem

37
New cards

60 to 70 ml

Replenishment rate for developer

38
New cards

100 to 110 ml; 14 inches or 35cm of film

Replenishment rate for fixer; for every

39
New cards

3 heating coils and 2, 500w

Used for dryer system

40
New cards

22s

Developing time for automatic processor

41
New cards

22s

Fixing time for automatic processor

42
New cards

26s

Drying time for automatic processor

43
New cards

90s

Total cycle time for automatic processing

44
New cards

15s

Wetting time for manual processing

45
New cards

30s

Rinsing time for manual processing

46
New cards

15min

Fixing time for manual processing

47
New cards

20min

Washing time for manual processing

48
New cards

30min

Drying time for manual processing

49
New cards

1hr 10min 45s

Total cycle time for manual processing

50
New cards

1mm x 4mm = 4mm2

Size of stationary target

51
New cards

15cm diameter, 17cm radius

Size of rotating target

52
New cards

2 PI 70mm x 4mm = 1760mm2

Total target area

53
New cards

3 cassette per second

Maximum exposure rate of ACC

54
New cards

24 to 48 hrs

Maintenance for perfect film-screen radiograph result in fuzzy radioraph

55
New cards

16mm and 35mm

Spot film sizes

56
New cards

100mA (usually 10-40mA)

Exposure factors for conventional tomography

57
New cards

3-6 seconds

Exposure amplitude

58
New cards

5%

Change in kVp for visible exposure difference in image for conventional tomography

59
New cards

100%

Exposure amplitude of narrow angle tomography

60
New cards

1mm solid state cadmium telluride

Size of multiple detector array in BMD

61
New cards

30mm metal brass

Height of detector collimator in BMD

62
New cards

72mm x 8mm metal brass

Size of detector collimator in BMD

63
New cards

18mm x 2.5mm lead

Size of source collimator in BMD

64
New cards

4mm lead

Size of shutter in BMD

65
New cards

35kev, 40;100kev, 40;70kev

Energies of gamma radiation emitted by the xray source in BMD

66
New cards

140kVp-high and 70 to 100kVp-low

Alternating kVp emitted by x-ray generator in BMD

67
New cards

70kev-high and 40kev-low

Separate energy emitted by rare earth filter

68
New cards

Every 3 or 6 months

Periodic maintenance