1/32
Flashcards covering essential vocabulary related to pacemaker indications, components, functions, coding, and malfunctions.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
Pacemaker (Cardiac)
An implanted device that delivers electrical impulses to stimulate heart contractions when intrinsic rhythm is inadequate.
Cardiac Output Formula
Cardiac Output = Heart Rate × Stroke Volume; pacemakers address inadequate heart rate to restore output.
Sick Sinus Syndrome
Disorder in which the SA node malfunctions, often requiring pacemaker support.
Atrioventricular (AV) Conduction Delay
Slowed or blocked impulse conduction through the AV node; common indication for ventricular pacing.
Post-Bypass Surgery Bradycardia
Slow heart rate following cardiac bypass surgery that may necessitate temporary or permanent pacing.
Millivolt / Milliamp (mV / mA)
Units measuring the amplitude of electrical energy delivered by a pacemaker lead to myocardial tissue.
Millisecond (ms)
Unit of time describing the duration of pacemaker stimulus delivery.
Synchronous (Demand) Pacemaker
Device that senses intrinsic cardiac activity and delivers impulses only when natural beats are inadequate.
Asynchronous (Fixed-Rate) Pacemaker
Device that paces continuously at a preset rate regardless of the heart’s own activity.
Pacer Coding – 1st Letter
Indicates the chamber(s) being paced: V (ventricle), A (atrium), or D (dual—both).
Pacer Coding – 2nd Letter
Indicates the chamber(s) being sensed: V, A, or D (dual).
Pacer Coding – 3rd Letter
Indicates pacemaker response to sensing: I (inhibit), T (trigger), or D (dual—both functions).
Capture
Successful depolarization of the cardiac chamber following a pacer spike, producing a contraction.
Sensing
Pacemaker’s ability to detect intrinsic cardiac electrical activity and decide whether to pace.
Failure to Fire
Malfunction in which the generator does not deliver an impulse when needed, often due to battery end-of-life.
Failure to Capture
Impulse is delivered but does not depolarize myocardium; causes include lead fracture or dislodgement.
Failure to Sense
Pacemaker does not properly detect intrinsic beats, resulting in inappropriate pacing activity.
VVI Pacemaker
Single-lead device that paces (V), senses (V), and inhibits (I) in the ventricle; used for AV block.
DDD Pacemaker
Dual-lead device that paces and senses both atrium and ventricle with dual responses (trigger/inhibit).
DVI Pacemaker
Dual-lead system that paces both chambers, senses ventricle only, and inhibits pacing as needed.
Biventricular (BiV) Device
Three-lead system pacing right atrium, right ventricle, and left ventricle via coronary sinus to resynchronize contraction.
Lead (Pacemaker)
Insulated wire that delivers electrical energy from the generator to cardiac tissue and senses activity.
Generator
Implanted power source and computer that controls pacing rate, output, and sensing thresholds.
Pacer Spike
Vertical line on ECG indicating delivery of a pacemaker impulse.
Negative QRS in Paced Rhythm
Appearance of a downward QRS deflection because ventricular stimulation originates from the lead tip upward.
Sinoatrial (SA) Node
Primary cardiac pacemaker whose failure may necessitate atrial or dual-chamber pacing.
Atrioventricular (AV) Node
Conduction relay between atria and ventricles; block here often treated with ventricular pacing.
Purkinje Fibers
Ventricular conduction network that delivers impulses to myocardium and interacts with pacing leads.
Trigger (Pacemaker Response)
Generator action that delivers an impulse when sensed activity is inadequate.
Inhibit (Pacemaker Response)
Generator action that withholds an impulse when sufficient intrinsic activity is sensed.
Capture Threshold
Minimal electrical energy required to consistently depolarize myocardium; determined during implantation.
End-of-Life (EOL) Battery
Stage when generator battery capacity is nearly depleted, risking failure to fire.
Coronary Sinus Lead
Electrode placed in coronary sinus to pace the left ventricle in biventricular devices.