Pacemakers – Key Vocabulary

Indications for Permanent Pacemakers

  • Central governing concept: (CO = HR \times SV). Any pathology that drops heart rate (HR) enough that stroke volume (SV) can no longer compensate becomes an indication for artificial pacing.
  • High-yield exam list (know verbatim):
    • Sick-sinus syndrome (brady–tachy, sinus arrest, sinus pauses)
    • AV-conduction delays/blocks (1°, 2°, 3°)
    • Post-cardiopulmonary bypass surgery bradyarrhythmias

Cardiac-Muscle & Electrical Basics (Why the Device Works)

  • Both ventricles contain exactly the same number of muscle fibres; LV fibres are thicker (↑diameter) because they chronically work against higher afterload.
  • EACH individual fibre—regardless of ventricle—needs:
    • A fixed amount of current (\text{mV} or \text{mA})
    • For a fixed time interval (ms)
    • Delivered electricity above the threshold ➔ ventricular systole ("capture").

Pacemaker Types

  • Synchronous ("demand")
    • Fires only when needed; goal = let intrinsic conduction work whenever possible.
  • Asynchronous ("fixed-rate")
    • Fires at a programmed rate regardless of native activity ➔ risk of competition arrhythmias.

Pacemaker Coding System (3 letters you must decode)

  1. Chamber paced: V (ventricle) / A (atrium) / D (dual—A+V)
  2. Chamber sensed: V / A / D (lead(s) monitor intrinsic activity here)
  3. Response to sensing:
    • I = Inhibit (withhold pulse if intrinsic beat sensed)
    • T = Trigger (fire when intrinsic event detected)
    • D = Dual (can do either)

Implantation & Threshold-Testing Workflow (high-yield sequence)

  1. Lead advanced via subclavian v., across tricuspid valve, tip lodged in RV endocardium.
  2. Rep turns generator ON ➔ increases output until 1st capture; record value.
  3. Gradually decrease output until capture LOST; record value.
  4. WAIT ~5 min (fibres stretch/relax, local irritation subsides).
  5. Re-increase output ➔ 2nd capture (almost always at a LOWER threshold).
  6. Program this 2nd (lower) threshold into generator; generator ↔ tip continuously communicate.

How Synchronous Pacing Actually Decides to Fire

  • SA node → AV node → His → Purkinje → reaches RV-lead tip.
  • Lead measures incoming electricity:
    • E{sensed} \ge E{programmed} ➔ Inhibit (native beat adequate)
    • E{sensed} < E{programmed} ➔ Trigger (generator emits pulse)

Classic Pacemaker Modes for the Exam

VVI Mode

  • Letters = V (paced) / V (sensed) / I (inhibit).
  • ONE lead in RV → implies underlying AV-block problem.
  • Rhythm strip clues:
    • Narrow, tall positive pacer spike immediately before a wide, negative QRS (resembles PVC).
    • Test items: “How many leads?”, “Where is the lead?”, “Identify the pacer spike vs QRS.”

DDD Mode

  • Letters = D / D / D ➔ TWO leads (RA + RV).
  • Covers both atrial & ventricular dysfunction.
  • Test usually shows strip and asks location/number of leads.

DVI Mode (mentioned, rarely tested since 4 yrs)

  • Know it exists; low probability.

Pacemaker Terminology & Malfunctions

  • Capture – EVERY pacer spike followed by the intended chamber’s depolarisation (P-wave or QRS).
  • Sensing – Device evaluates intrinsic beats and decides whether to fire.
  • Failure to Fire – No spike when one was needed.
    • Etiology: generator end-of-life (EOL) / depleted battery.
  • Failure to Capture – Spike present, but no chamber response.
    • Etiology: lead fracture, lead dislodgement, poor tip-myocardium contact.
  • Failure to Sense – Generator ignores intrinsic activity; spikes appear haphazardly.
    • Etiology: sensitivity setting too low/high, lead issues, electromagnetic interference.

Biventricular (Bi-V) / CRT Devices

  • Purpose: Re-establish synchrony between RV & LV contractions.
    • Physiologic delay ≈ 3{-}4\text{ ms} = 0.003!–!0.004\ \text{s} (RV then LV).
    • Pathology elongates or abolishes this delay ➔ dyssynchrony, ↓CO, pulmonary backup.
  • Lead set-up (3 leads):
    • RA lead ➔ pace RA
    • RV lead ➔ pace RV
    • CS (coronary sinus) lead ➔ pace LV

Rapid-Fire Exam Q&A (Likely One-liners)

  • Single lead in atrium only ➔ problem = SA-node disease.
  • Single lead in ventricle only ➔ problem = AV-block.
  • In an A-V pacer, the 1st spike triggers atrial contraction.
  • Leads are introduced via the subclavian vein (NOT the IJ).
  • Why do paced QRS complexes look like PVCs (negative deflection)?
    • Activation travels bottom→top (Purkinje → myocardium) instead of the normal top→bottom.
  • Asynchronous pacing vs demand pacing – know definitions.
  • Can you defibrillate a paced patient? YES, just do NOT place paddle directly over generator.

Practical / Ethical / Real-World Points

  • Device = "an incredibly intelligent computer" that adapts to changing thresholds over the years.
  • Waiting 5–10 min during implant respects myocardial irritation & immune response—reduces need for high chronic outputs, prolongs battery life (ethical use of resources).
  • Bi-V/CRT devices improve HF quality-of-life—ties to evidence-based medicine and healthcare economics.