1/195
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
T/F the isocenter of a given treatment unit is not a fixed point and therefore can be modified as needed
false
T/F the isocenter is a point in space at a specified distance from the source of radiation around which the gantry rotates
true
T/F the isocenter is a point of intersection of the gantry, collimator and table axes of rotation
true
T/F the treatment room lasers are aligned to the isocenter
true
what are the factors influencing the dose delivered to a patient
beam energy, tissue density, calculation depth, field size, distance
determine the fraction dose for a prescription which calls for a total dose of 45 Gy delivered in 25 fractions
180 cGy
a conventional fractionation schedule delivers a fraction dose of
180-200 cGy
the division of the body into equal left and right segments and can also be referred to as midsagittal
midline
defined as the measure of energy transferred from radiation photons to charged particles (electrons) in matter
KERMA
the distance beneath the skin surface where the prescribed dose is to be delivered
depth
A patient has lost over 7 pounds during the course of treatment. a therapist is asked to acquire an AP/PA separation measurement while the patient is on the treatment table, but calipers are not available in the department. what is another effective approach to obtaining the patient's separation measurement
use the ODI to obtain SSD readings and use the readings to determine separation
the anterior field of an isocentric setup has an SSD reading of 89.5 cm, based on this information, what is the depth to isocenter for this field
10.5 cm
corresponds to the field defined by the light localizer
geometric field size
defined as the area included within the 50% isodose curve at the depth of maximum dose
physical field size
solve for the equivalent square field size of a rectangular field measuring 12cm x 22cm
15.5cm x 15.5cm
based on a standard configuration of field dimensions as shown in the image, which axis reflects field length
Y2-Y1
T/F the radiation therapy beam is comprised of primary radiation and scatter radiation
true
radiation that is scattered back toward the surface after entering tissue is termed
backscatter radiation
as beam energy increases, the depth of Dmax
increases
the dose at Dmax, also called given dose, is always ___ ____ the prescribed dose for single fields
greater than
the effect whereby megavoltage photon beams reach 100% dose at a depth greater than zero thus reducing dose to superficial tissue best defines
skin sparing
the amount of radiation exposure produced by a treatment machine at a specified reference field size and distance best defines
output
as field size decreases, output or dose rate
decreases
as distance increases, output or dose rate
decreases
as attenuating medium thickness increases, output or dose rate
decreases
the mathematical relationship between radiation intensity and beam divergence whereby the intensity us inversely proportional to the square of the distance from the source of radiation best defines
inverse square law
the relationship between field size and distance is described as
direct
a lung cancer patient treated on a 100cm SAD unit has a AP/PA separation measuring 25 cm. if the isocenter is to be placed at midplane, what is the depth of isocenter
12.5cm
the intensity of a beam is 5mR/h at a distance of 10 cm. what is the intensity of the beam at 5 cm
20mR/hr
if the dose rate of a coblat-60 unit is 80 cGy/minut at 100cm SSD, what is the dose rate at 80cm SSD
125cGy/minute
a patient is treated with an isocentric technique on a 100cm SAD unit. the patient's separation is 23cm. if the isocenter is placed at a depth of 11cm from the anterior surface, at what depth is the isocenter from the posterior surface
12cm
if the collimator setting is 10cm x 10cm for a 100cm isocentric setup in which of the isocenter is placed at a depth of 8cm, what is the field size on the patient's skin surface
9.2cm x 9.2cm
T/F beam energy is an influencing factor for output
false
Dmax: 10 mV
2.5cm
Dmax 6 MV
1.5cm
Dmax 18MV
3.5cm
Dmax 4MV
1cm
T/F the depth of Dmax is influence primarily by beam quality
true
T/F Dmax is a point in space around which the linear accelerator rotates
false
T/F Dmax is the depth at which electron equilibrium occurs for photon beams
true
T/F Dmax is a point where the percent depth dose is normalized to 50%
true
units of measurements for absorbed dose
J/kg, cGy, Gy, rad
calculate the exit calculation point for a patient with a separation of 20cm being treated with a 6MV beam
18.5
Cerrobend blocks must achieve less than 5% transmission and therefore typically consist of ____ half value layers
5
the rotational axis of the gantry, collimator and treatment table
isocenter
pertaining to or involving right angles or perpendiculars
orthogonal
involves the use of reference points at a more stable location for alignment purposes and shifts are made to arrive at the planned isocenter
pinning
concept in which the isocenter is located relative to three setup coordinates
triangulation
indicates that points or beams lie within the same two-dimensional geometric plane
coplanar
what may be employed to achieve a uniform dose distribution among adjacent photon fields
half beam block, geometrically calculated skin gap, beam splitter
if a field measures 15 cm in length at 95cm SSD what will it measure at 120 cm SSD
19cm
refers to changing the location of the match line or the skin gap on a scheduled basis so that possible points of over or under dose do not occur at one spot throughout the entire course of treatment
feathering
two fields are planned to abut at a depth of 9cm. calculate the skin separation required if the field lengths are 40cm and 32cm respectively. Both fields are treated at 100cm SSD
3.2cm
what is the magnification factor if the actual collimator readout is 12cm x 12cm and the field size on the image measures 16cm x 16cm
1.33
photon blocks must be a minimum of ____cm from the patient's skin surface to reduce scatter from reaching the patient and contributing a non-therapeutic dose
15-20
the open collimator equivalent square field size is used for the inverse square correction factor
false
the open collimator equivalent square field size is used for the collimator scatter factor
true
the open collimator equivalent square field size is used for the phantom scatter factor
false
the open collimator equivalent square field size is used for the patient attenuation factor
false
as beam energy decreases, the PDD will
decrease
as depth decreases, the PDD will
increase
as field size decreases, the tissue maximum ratio will
decrease
as beam energy increases, the tissue phantom ratio will
increase
back scatter factor (BSF) becomes negligible around ____ MeV
8
a calculated value that identifies how long the machine is "on" to deliver a specific dose under specified conditions beset defines
MU
the relationship between dose and PDD is described as being
direct
the relationship between dose and MU is described as being
direct
a wedge factor of .831 indicates that _% of the dose is being transmitted through the wedge
83.1
radiation that is scattered 180 degrees from its original path is termed
backscatter
when comparing two arrangements in which everything remains equal except the use of a wedge, which field will have the higher monitor unit setting
wedge field
for a non-isocentric treatment setup, _ is the patient attenuation factor used in MU calculations to demonstrate the central axis at a given depth
PDD
a lung patient has a chest separation measuring 24cm. if the isocenter is to be placed at midplane, what is the depth of isocenter
12cm
a special application of the inverse square law used in special cases of non-isocentric MU calculations, for the purpose of converting the PDD at a reference distance to the PDD at a non-reference distance is the
mayneord's factor
when comparing two arrangements in which all elements remain equal except field size, will a smaller or larger field size have a lower MU setting
larger
tissue air ratio is dependent on distance
false
tissue maximum ratio is dependent on distance
false
tissue phantom ratio is dependent on distance
false
PDD is dependent on distance
true
as distance increases, the PDD will
increase
as field size increases beyond the established reference field size for output, the beam output will
increase
the ratio of absorbed dose at a given depth in phantom (tissue) to the absorbed dose at the same point in free space (air) best defines
tissue air ratio
the ratio of absorbed dose at depth to absorbed dose at Dmax is termed
PDD
represents the ratio of dose at a given depth in phantom (tissue) to the dose in phantom (tissue) at a depth of Dmax
tissue maximum ratio
the angle between the slanted isodose line and a line perpendicular to the central axis of the beam best defines
wedge angle
treatment planning volume encompasses the palpable, visible, and/or demonstrable extent of malignant disease and areas of possible microscopic or subclinical disease
CTV
to reduce surface dose resulting from scatter, compensators and wedges must be a minimum of _Cm from the patient's surface
15-20
treatment planning volumes encompass the CTV plus a margin to account for physiologic movement (such as heartbeat, respiration, and peristalsis) and changes in the size, shape, or position of the CTV
ITV
treatment techniques refers to an intensity-modulated radiation therapy (IMRT) treatment characterized by the gantry rotating while the treatment beam is on, along with multiple MLC positions and changing dose rate?
intensity modulated arc therapy
The treatment planning volume that encompasses the internal target volume plus a margin for geometric uncertainties such as patient motion, beam penumbra, and treatment setup differences is the
PTV
the ratio of normal tissue tolerance dose to tumor lethal dose is referred to as the
therapeutic ratio
generally, the dose at any depth is on the central axis of the beam and gradually toward the beam edges
highest, decreases
a beam modification device that modifies the distribution of dose by progressively decreasing beam intensity across the irradiated field
wedge
a _ field arrangement may be used to treat small, unilateral tumors to depths up to 7cm
wedge pair
what is the wedge angle needed to achieve improved dose homogeneity for two beams which have a hinge angle of 90 degrees
45 degrees
wedge fields commonly demonstrate a hot spot under the ____ end of the wedge
toe
small positional errors may affect the treatment outcome of IMRT more so than when conventional or 3D-CRT is used
true
the difference between the max and min intensity of the central 80% of the beam profile or the evenness of dose across a beam profile is termed
flatness
beam compensators such as wedges have been used in radiation therapy to attenuate the beam in an effort to produce a more dose distribution within the target volume
homogeneous
the volume of tissue receiving a significant percentage of the specified target dose (in relation to normal tissue tolerance values) is termed the
irradiated volume
an extremely objective, premliminary treatment plan evaluation tool which plots target or normal structure volume as a function of dose is termed
dose volume histogram