kilovoltage peak (kVp)
exposure factor in the diagnostic range controlling beam quality or beam penetrability
controls contrast
as it increases fewer x rays are absorbed by the patient
can increase compton scattering when increased
15% rule
Changing the kVp by 15% has the same effect as doubling the mAs or reducing the mAs by 50% which decreases patient dose
milliampere seconds (mAs)
exposure factor in diagnostic range that determines beam quantity
patient dose follows a direct proportion
controls patient dose more than kVp
protective aprons
made of 0.5 mm of lead
(equivalent to 2 HVL, reduces occupational exposure to 10%)
half value layer (HVL)
the amount of absorbing material that will reduce the intensity of the primary beam to one-half its original value
tenth value layer (TVL)
the amount of absorbing material that will reduce the intensity of the primary beam to one-tenth its original value
3.3 HVL
1 TVL =
collimate
to reduce patient dose during imaging procedures, like KV and MV, do what to the field sizes?
thicker
thicker/thinner cause more scatter
peds patients
exposure time many need to be reduced to reduce patient motion for which group of patients?
mA
to compensate for shorter exposure time, __ may need to be increased which will increase patient dose
radon
varies based on where the person lives, naturally occuring
#1 source of background radiation
cosmic radiation
radiation that comes from the sun and stars
exposure increased with increasing altitudes
type of background radiation?
internal radioactive elements
main source in our body is K-90 which emits beta and gamma rays
type of background radiation?
As Low As Reasonably Achievable
ALARA
time, distance, shielding
most important aspects of radiation protection
shielding
best form of protection in radiation therapy to reduce exposure
10
state and federal regulations require monitoring of anyone who may receive more than __% of the maximum permissible dose (MDP)
records of employee dose
institutions must keep indefinitely
when occupationally exposed workers change employment, exposure records should be transferred to new employment
restricted/controlled areas
supervised by a radiation safety officer
limited to radiation workers
occupational limit of radiation of no more than 1 rem per year (
unrestricted/uncontrolled areas
not under supervision
limited to the general public
cannot receive more than 0.5 rem/year (
radiation area sign
posted in areas where >0.05 mSv/hr at 30 cm from source
high radiation area sign
posted in areas where >1mSv/hr at 30 cm from source
very high radiation area sign
posted in areas where >5 Gy/hr at 1 m from source
caution: radioactive materials sign
posted in areas where sources of radiation are used and/or stored (ex: I-131)
specific and sensitive
low levels of radiation detection devices should be:
primary barrier
directly hit by the useful part of the beam
concrete
most common and cheapest barrier
2 meters
average thickness of a primary concrete barrier
1 meter
the average thickness of a secondary concrete barrier
secondary barrier
hit by scatter or leakage radiation
heat and burn the patient
interactions between biomedical implants and the MRI RF coils can cause the device to
fill out a pre-mri screening form and an interview
prior to the MRI the patient must (2 things)
MR safe
nonferromagnetic or weakly ferromagnetic materials are considered
medical equipment must be labeled with this
aneurysm clips
can be made of pure titanium, elgiloy, MP35N, are MR compatible
neurostimulators
patients with implanted ___ should not have an MRI performed
newer devices that are MR safe for only MR's with 2T or less
implantable spinal fusion stimulators
MR safe
must have radiograph performed prior to ensure leads are not broken (cause burning)
1.5T or less
implantable drug infusion pumps with magnetic switches
not MR safe
heart valve prostheses containing metal
MR safe
ocular implants
not MR safe
radiographs of orbits should be performed prior to MR to determine metal presence
dental implants
MR safe
except those with magnetically activated peices
cochlear implants
not MR safe due to damage to the device
vascular access ports and catheters
MR safe if made of stainless steel, titanium, silicone, or plastic
IV stents, filters, coils
made of non magnetic material and are MR safe
breast tissue expanders and implants
not MR safe
penile implants
not MR safe
welders, metal workers
profession where the risk of metal in the eye can occur leading to blindness if MR is performed, radiograph is needed prior to MR
pregnancy
should only be performed on this group when other diagnostic studies are not sufficient
tattoos
patients with ___ can experience skin reaction including tingling sensations, skin irritation, swelling, heating, burning
can cause image artifact
missle effect
ferromagnetic object is throw into MR system
handling and disposal of radioactive materials
ALARA, PPE (forceps, gloves, gown, mask)n needed when
nuclear regulartory commission
regulates the safe use of radioactive materials and protects the public health in regards to radioactive materials
department of transportation
regulate the transportation of hazardous radioactive material
environmental protection agency
regulates the removal, storage, and usage of nuclear waste
klystron, waveguide, circulator, cooling system
the drive stand contains four parts:
water cooling system
prevents the linear accelerator from overheating
gantry, collimator, and couch
three parts of the linear accelerator that move around isocenter
power supply
provides direct current tot eh modulator
modulator
powers the microwaves source and electron gun
microwaves
is injected into the waveguide to accelerate electrons
klystron
amplifies microwaves
used for energies above 12MV
magnetron
generates microwaves
low energies
circulator
stops microwaves from being reflected and reentering the klystron
located between the klystron and waveguide
electron gun
generates electrons and sends them into the accelerator
made of tungsten
accelerator structure (wave guide)
tube where electrons flow to the gantry under a vacuum
copper tube
uses microwaves from klystron to accelerate electrons to the target
contains SF6 to prevent arcing
positioned horizontally for higher energy production
magnent system
used to bend electrons 90 to 270 degrees to aim towards the target
270 creates a more confined beam
treatment head
where the beam is formed and beam output is monitored
target > primary collimator > beam flattening filter > ion chambers > secondary jaws
photon therapy linear accelerator components
flattening filters
made of lead tungsten uranium steel or aluminum
photon therapy
ion chambers
small chambers located in path of radiation beam and check the beams dose rate and symmetry
target > primary collimator > beam scattering filter > ion chambers > secondary jaws
electron therapy linear accelerator components
scattering foil
made of lead
spread out 3 mm pencil electron beam into larger area
volumetric
CT scanner obtains (volumetric/geometric) images of the patient, which are then used for tx planning
rotating x ray tube with a fixed ring of imaging detectors
internal components of CT scanner
controls for couch movements, gantry tilt, emergency off buttons, lasers
external components of CT scanner
helical
current CT scanners utilize ___ scanning, continuous x ray rotation while the couch moves
larger (70-85 cm)
CT scanners for therapy have a smaller/larger bore size than diagnostic
ionization chamber
gas filled chamber with two electrodes with an applied voltage and a meter that measures the electrical signal
measures the production of rad therapy equip
most common way to measure output
inc/direct
ion chamber:
larger chamber has dec/inc sensitivity?
exposure is in/directly prop to the current?
temp, pressure, volume
when calibrating the ionization chamber, certain correction factors are needed (3) to cause ionizations
pocket dosimeter
tyoe of ion chamber that is low energy personnel dosimeter
cutie pie
Ionization chamber-type portable survey meter
farmer chamber
type of thimble chamber
used to calibrate external beam radiation machines
Geiger-Muller detector
portable instrument used to detect radiation of low levels and radioactive contamination
very sensitive, not dose measuring
used to locate misplaced sources
thermoluminescent dosimeter
made of lithium fluoride
electrons are released and visible light photons are emitted when they are excited by heat
commonly used for in vivo measurements to record entrance or skin dose
optically stimulated luminescence
made of aluminum oxide
electrons are released and visible light photons are emitted when they are excited by light
most common detector used to measure occupational (personal) exposure
diodes
solid-state detector
sensitive
real-time reading
equipment of choice for patient dosimetry
used for a daily QA check on LA beam output
neutron detectors
important in proton and photon therapy above 10 MV
ionization chambers can be used to detect
boron trifluoride (BF3)
proportional counters are used for neutron dosimetry are filled with what gas?
door interlock
radiation beam is unable to turn on when tx room door is ope
daily functional check
collision interlock
stops motion of gantry when too close
daily functional check
laser guard interlock
checked monthly
electron applicator itnerlock
checked annually
safety lights
outside the treatment room
checked daily (at the console and door)
the light bulb can be burned out, the reason for malfunction
emergency off button
terminate the radiation beam and shut down the machine
checked monthly
laser alignment
can affect geometric parameters
daily functional check
must intersect at iso
2 mm
non imrt tolerance
1.5 mm
imrt tolerance
1 mm
srs/sbrt tolerance
40 +/- 5 C
temperature of the waveguide should be