RTE 006: Chapter 13 Screen-Film Radiographic Technique

0.0(0)
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
Card Sorting

1/99

flashcard set

Earn XP

Description and Tags

P2 Coverage •Exposure Factors ‒Kilovolt Peak ‒Milliamperes ‒Exposure Time ‒Distance •Imaging System Characteristics ‒Focal-Spot Size ‒Filtration ‒High-Voltage Generation •Patient Factors ‒Thickness ‒Composition ‒Pathology

100 Terms

1
New cards
common (SID) at 180 cm (72 in)

currently, 300 cm
The source-to-image distance (SID) from chest radiography is
2
New cards
common (SID) fixed at 100 cm (40 inches)

currently, 120
For most radiographic examinations, the source-to-image distance (SID) from the tabletop radiography is
3
New cards
Screen-film Radiographic Technique
The combination of settings selected on the control panel. Its purpose is to produce a high-quality image.
4
New cards

Exposure factors

The factors that influence and determine the quantity and quality of x-radiation to which the patient is exposed are called

5
New cards
Radiation Quantity
Refers to radiation intensity measured in mGya or mGya/mAs (mR or mR/mAs)
6
New cards
Radiation Quality
Refers to x-ray beam penetrability, best measured by the half-value layer (HVL) of the x-ray beam
7
New cards
kVp, mAs, Exposure time, SID (source-to-image receptor distance)
What are the primary exposure factors?
8
New cards
Kilovolt Peak (kVp)
Meaning of kVp
9
New cards
Focal-spot size, Distance, Filtration
What are the secondary factors of exposure factors?
10
New cards

kVp and mAs

The factors principally responsible for x-ray quality and quantity.

11
New cards
kVp
The primary control of x-ray beam quality (beam penetrability)
12
New cards
Controls radiographic contrast, Influences beam quantity, and Determines the average OD
Functions of kvp
13
New cards

mA/s (milliamperes / per second)

It determines the number of x-rays produced (radiation quantity) 1 A = 1 C/s = 6.3 x 1018 per second

14
New cards
Higher kVp:
-Higher beam quality
-Greater beam penetrability
-More scatter radiation
• Rationale: more Compton effect interaction
- Less differential absorption
-Result: reduced image contrast
What does high kVp do?
15
New cards
Higher mA:
-More electrons produced
-Higher x-ray quantity
-Higher patient dose
-No change in x-ray quality
What higher mA do?
16
New cards
1 A = 1 C/s (coulomb per second) = 6.3 x 10^18 electrons per second
Formula in getting 1 A (ampere)
17
New cards
Inversely proportional
The relationship between low exposure time and high mAs are
18
New cards
Kept as short as possible (ALARA)
Exposure time needs to be
19
New cards
To minimize the motion blur due to patient motion because it provides the required x-ray intensity
The purpose for keeping the exposure time short is to
20
New cards
Also the patient dose
With a constant exposure time, mA controls the x-ray quantity and therefore
21
New cards
mA
X-ray quality remains fixed with a change of what
22
New cards
mAs = mA x time

(don't forget to convert to seconds)
Formula of mAs
23
New cards
mA (second exposure) time (second exposure)
------------------------=
mA (first exposure) time (first exposure)
Other formula for getting the mAs value?
24
New cards
mA (first exposure) x time (first exposure)
mA (second exposure) x time (second exposure)
25
New cards

Distance

It determines the intensity of the x-ray beam at the image receptor.

26
New cards

Inverse Square Law

Distance affects exposure of the image receptor according to

27
New cards

600 mA

Inexpensive radiographic imaging systems designed for private physician's offices normally have a maximum capacity of

28
New cards

1500 mA

Interventional radiology imaging systems may have the capacity of

29
New cards
Motion Blur
Short exposure time reduces
30
New cards

mAs

What is the key factor on the control of OD (optical density) on the radiograph?

31
New cards

mAs value

It is a measure of the total number of electrons conducted through the x-ray tube for a particular exposure

32
New cards
mAs (milliampere per second)
One measure of electrostatic charge
33
New cards
Direct Square Law
It allows a radiologic technologist to calculate the required change in mAs after a change in SID to maintain constant OD.
34
New cards
To minimize the cardiac magnification
What is the purpose of Inverse Square Law?
35
New cards
<p>(Note that both the original mAs value and the original SID are in the denominator rather than reversed, as in the inverse square law.)</p>

(Note that both the original mAs value and the original SID are in the denominator rather than reversed, as in the inverse square law.)

Inverse Square Law Formula

36
New cards

Distance (SID)

What affects OD?

37
New cards

Distance

____________ has no effect on radiation quality

38
New cards
What are the uses of Large focal spot?
- It is used in general imaging
- Ensures that sufficient mAs can be used to image thick or dense body
parts.
- Provides for a shorter exposure time, which minimizes motion blur.
- Produce more x-rays compared with small focal spot
39
New cards
What are the uses of Small focal spot?
-It is used for fine-detail radiography, in which the quantity of x-rays is
relatively low.
-Always used for magnification radiography
- These are normally used during extremity radiography and in
examination of other thin body parts in which higher x-ray
quantity is not necessary.
40
New cards

The x-ray quantity or quality

Changing the focal spot for a given kVp/mAs setting does not change

41
New cards
<p>(Note that both the original mAs value and the original SID are in the denominator rather than reversed, as in the inverse square law.)</p>

(Note that both the original mAs value and the original SID are in the denominator rather than reversed, as in the inverse square law.)

Direct Square Law Formula

42
New cards
Small Focal Spot and Large Focal Spot
What are 2 focal spot?
43
New cards
Bi focal X-ray Tube
Another name for 2 Focal Spot
44
New cards
0.3 mm/1.0 mm focal spots
X-ray tubes used in interventional radiology procedures or magnification radiography may have, how many focal spot?
45
New cards
0.1 mm/0.3 mm focal spots
Mammography x-ray tubes have, how many focal spot?
46
New cards
Microfocus Tube
Are designed specifically for imaging very small microcalcifications at
relatively short SIDs?
47
New cards
Inherent filtration, Added filtration, and Compensating filters
Three types of filtration
48
New cards

Inherent filtration

It is made of glass or metal envelope.

49
New cards

approximately 0.5 mm Al equivalent

For general-purpose tubes, the value of inherent filtration is

50
New cards
1.0 mm Al equivalent
In light-localizing collimator, it provides an additional of
51
New cards

Added filtration

1-mm Al filter is inserted between the x-ray tube housing and the collimator

52
New cards
Compensating filters
Are shapes of aluminum mounted onto a transparent panel
that slides in grooves beneath the collimator.

These filters also balance the intensity of the x-ray beam so as to
deliver a more uniform exposure to the image receptor.
53
New cards
Single phase, Three phase, and High Frequency
Three basic types of high-voltage generators
54
New cards
the type of high-voltage generator that is used.
The radiation quantity and quality produced in the x-ray tube are influenced by
55
New cards
Half-wave rectification
Results in the same radiation quality as is produced by full-
wave rectification, but the radiation quantity is halved.
- Used in mobile and dental x-ray imaging systems
56
New cards
Full-wave rectification
Same as half-wave rectification except there is no dead time.
- X-rays are emitted continually as pulses.
- The required exposure time for full-wave rectification is only
half that for half-wave rectification.
57
New cards

Three-phase power 6p (pulse) /12p (pulse)

  • Results in higher x-ray quantity and quality.

  • more efficient than singlephase power.

  • More x-rays are produced for a given mAs setting

  • The average energy of those x-rays is higher.

  • The x-radiation emitted is nearly constant rather than pulsed.

58
New cards

High-frequency generators

The voltage waveform is nearly constant, with less than 1% ripple.

  • High-frequency generation results in even greater x-ray quantity and quality.

  • Used increasingly with dedicated mammography systems, computed tomography (CT) systems, and mobile x-ray imaging systems.

59
New cards
(Anatomical) Thickness, (Body) Composition, and Pathology
First group of patient factors
60
New cards

Body habitus

The general size and shape of a patient is called

61
New cards

Sthenic

Meaning “strong, active”—patients are average patients.

62
New cards

Hyposthenic

Thin but healthy appearing; these patients require less radiographic technique.

63
New cards

Hypersthenic

Big in frame and usually overweight.

64
New cards

Asthenic

Small, frail, sometimes emaciated, and often elderly

65
New cards

the more x-radiation is required to penetrate the patient to expose the image receptor

The thicker the patient?

66
New cards

Sthenic, Hyposthenic, Hypersthenic, Asthenic

What are the four general states of body habitus?

67
New cards

Sthenic patients

Radiographic technique charts are based on

68
New cards

Calipers

Are available to the radiologic technologist for use to measure the thickness of the anatomy being imaged

69
New cards

This is due to the fact that the primary X-ray beam has not yet been attenuated by the tissue at this point, and also, that area is exposed by some of the scattered radiation from the body.

Why should we not guess the patient thickness?

70
New cards

They are image quality factors that consist of OD (optical density), contrast, detail, distortion

Second group of patient factors?

71
New cards

They are exposure technique factors that includes kVp, milliamperage, exposure time, SID, Grids, screens, focal-spot size, and filtration

Final group of patient factors?

72
New cards

The patient’s size, shape, and physical condition

What influences greatly the radiographic technique?

73
New cards

High subject contrast

The chest has

74
New cards

Low subject contrast

The abdomen has

75
New cards

Increased radiolucency or radiopacity

Pathology can appear with what?

76
New cards

Radiolucency or Radiopacity

Body tissues are often described by their degree of what?

77
New cards

Mass density

The radiologic technologist must estimate the __________ of the anatomical part and the range of ___________ involved. (Only one answer)

78
New cards

Low kVp and High mAs

Soft tissue requires ____ kVp and _____ mAs

79
New cards

Low kVp, thin

Extremity consists of soft tissue and bone, ___ kVp is used because the body part is ____

80
New cards

Low, high, intermediate

Lung tissue has very ___ mass density, the bony structures have ____ mass density, and the mediastinal structures have ____________ mass density.

81
New cards

It provides in an image with satisfactory contrast and low patient radiation dose

High kVp and low mAs can be used to good advantage as a result?

82
New cards

Type of pathology, size, and composition

The ______________, its ____, and its ___________ influence radiographic technique.

83
New cards

Destructive pathology

Causing the tissue to be more radiolucent.

84
New cards

Constructive pathology

Increasing mass density or composition, causing the tissue to be more radiopaque.

85
New cards

Patient examination request form and previous images.

(The radiologic technologist should also not hesitate to more information from the referring physician, the radiologist, or the patient regarding the suspected pathology.

In pathology, what do we need to help evaluate our patient before examination?

86
New cards

Increase

An increase in kVp will result in ________ x-ray quantity

87
New cards

Increase

An increase in mA will result in ________ x-ray quantity

88
New cards

Increase

An increase in exposure time will result in ________ x-ray quantity

89
New cards

Increase

An increase in mA/s will result in ________ x-ray quantity

90
New cards

Decrease

An increase in distance will result in ________ x-ray quantity

91
New cards

Decrease

An increase in voltage ripple will result in ________ x-ray quantity

92
New cards

Decrease

An increase in filtration will result in ________ x-ray quantity

93
New cards

Increase

An increase in kVp will result in ________ x-ray quality

94
New cards

No change

An increase in mA will result in ________ x-ray quality

95
New cards

No change

An increase in exposure time will result in ________ x-ray quality

96
New cards

No change

An increase in mA/s will result in ________ x-ray quality

97
New cards

No change

An increase in distance will result in ________ x-ray quality

98
New cards

Decrease

An increase in voltage ripple will result in ________ x-ray quality

99
New cards

Increase

An increase in filtration will result in ________ x-ray quality

100
New cards

manipulation

Secondary exposure factors require ____________ for particular examinations