1/98
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
the pulser of an instrument automatically reduces the PRF for deeper imaging to avoid the ___ artifact.
range ambiguity
if an echo arrives 143 microseconds after the pulse that produced was emitted, it should be located at a depth of ____ cm. if a second pulse was emitted 13 microseconds before the arrival of this echo, it will be placed incorrectly at a depth of ___ cm.
11 cm, 1 cm
if the propagation speed in a soft tissue path is 1.60 mm/microsecond, a diagnostic instrument assumes a propagation speed too ____ and will show reflectors too ____ the transducer.
low, close to
true or false? mirror image can occur with only one reflector.
false
the most common artifact encountered in Doppler ultrasound is ___.
aliasing
which of the following can reduce or eliminate aliasing?
-increased PRF
-increased doppler angle
-increased operating frequency
-use of continuous wave mode
increased PRF, increased doppler angle, use of continuous wave mode
true or false? the fine texture in soft tissue indicates the excellent resolution that actually exists there.
false
the fact that a beam, as it scans through tissue, has some nonzero width perpendicular to the scan plane results in the ____ _____ artifact.
slice thickness
which of the following can cause improper location of objects on a display?
-shadowing
-enhancement
-speed error
-mirror image
-refraction
-grating lobe
speed error, mirror image, refraction, grating lobe
true or false? refraction can cause shadowing.
true
unreal structure displayed
reverberation
structure displayed with improper brightness
shadowing, enhancement
structure improperly positioned
propagation speed error, refraction
structure improperly shaped
propagation speed error, refraction
reverberation results in added reflectors being imaged with equal ___.
separation
in reverberation, subsequent reflections are ____ than previous ones.
weaker
enhancement is caused by a ____.
weakly attenuating structure
which of the following can correct or eliminate aliasing?
-decreased PRF
-decreased doppler angle
-increased operating frequency
-baseline shifting
baseline shifting
true or false? shadowing results in a decreased echo amplitudes.
true
propagation speed error results in improper ____ position of a reflector on the display.
axial
to avoid aliasing, a signal voltage must be sampled at least ____ time(s) per cycle.
2
if the highest doppler shift frequency present in a signal exceeds ____ the PRF, aliasing will occur.
one half
when doppler gain is set too high, which artifact is likely to occur?
spectrum mirror image
which artifact should be suspected if one observes twin gestational sacs when scanning through the rectus abdominis muscle?
refraction
range ambiguity can occur in which of the following?
-imaging instruments
-duplex instruments
-pulsed wave doppler instruments
-color flow instruments
-all of the above
all of the above
if the PRF is 4 kHz, which of the following doppler shifts will cause aliasing?
1 kHz
2 kHz
3 kHz
4 kHz
3 kHz, 4 kHz
if the PRF is 10 kHz, which of the following doppler shifts will cause aliasing?
1 kHz
2 kHz
3 kHz
4 kHz
none of the above
none of the above
there is no problem with aliasing as long as the doppler shifts are ____ half the PRF.
less than
true or false? if doppler shift is 2.6 kHz, no aliasing would result with a PRF of 10 kHz.
true
if there were a problem in question 27 (true: if doppler shift is 2.6 kHz, no aliasing would result with a PRF of 10 kHz.), ____ could be used to avoid it.
continous-wave, power
if red represents a positive doppler shift and blue represents a negative one, what color is seen for normal flow toward the transducer? what color is seen for aliasing flow toward the transducer? what colors are seen for normal flow away and for aliasing flow away from the transducer?
red, blue, blue, red
when a pulse is emitted before all the echoes from the previous pulse have been received, which artifact occurs?
range ambiguity
when a strong reflector is located in the scan plane, which of the following artifactsi s likely to occur?
-aliasing
-range ambiguity
-spectrum mirror image
-location mirror image
-speckle
location mirror image (in class, he said the answer was “mirror image”)
increasing PRF to avoid aliasing can cause which of the following?
-baseline shift
-range ambiguity
-spectrum mirror image
-location mirror image
-speckle
range ambiguity
which of the following decreases the likelihood of range-ambiguity artifact?
-decreasing operating frequency
-decreasing PRF
-decreasing doppler angle
-baseline shift
-increasing pulser output
decreasing PRF
range ambiguity produces which error in spectral doppler studies?
incorrect gate location
range ambiguity produces which error in anatomic imaging?
range too short
if the maximum imaging depth is 5 cm, the frequency is 2 MHz, and the doppler angle is zero, what is the maximum flow speed that will avoid aliasing and range ambiguity?
300 cm/s
true or false? solving aliasing by decreasing operating frequency increases the possibility of range ambiguity artifact.
true
true or false? if operating frequency is increased to decrease the possibility of range ambiguity (by increasing attenuation), the possibility of aliasing increases.
true
if a pulsed wave Doppler sample volume is located at a depth of 8 cm, the sampled echoes arrive at what time following the emission of the pulse?
104 microseconds
a tissue equivalent ____ has an attenuation of approximately 0.5 dB/cm-MHz and a propagation speed of 1.54 mm/microsecond. a ____ does not mimic tissue but provides a means for measuring some aspect of instrument performance.
phantom, test object
match the parameters measured with the items used.
nylon fibers
attenuating scattering material
simulated cysts
hyperechoic and hypoechoic simulated lesions
thin scattering layer
axial resolution =
nylon fibers, simulated cysts
match the parameters measured with the items used.
nylon fibers
attenuating scattering material
simulated cysts
hyperechoic and hypoechoic simulated lesions
thin scattering layer
lateral resolution =
nylon fibers, simulated cysts
match the parameters measured with the items used.
nylon fibers
attenuating scattering material
simulated cysts
hyperechoic and hypoechoic simulated lesions
thin scattering layer
range accuracy =
nylon fibers
match the parameters measured with the items used.
nylon fibers
attenuating scattering material
simulated cysts
hyperechoic and hypoechoic simulated lesions
thin scattering layer
caliper accuracy =
nylon fibers
match the parameters measured with the items used.
nylon fibers
attenuating scattering material
simulated cysts
hyperechoic and hypoechoic simulated lesions
thin scattering layer
contrast resolution =
hyperechoic/hypoechoic lesions, simulated cysts
match the parameters measured with the items used.
nylon fibers
attenuating scattering material
simulated cysts
hyperechoic and hypoechoic simulated lesions
thin scattering layer
compensation=
attenuating scattering material
match the parameters measured with the items used.
nylon fibers
attenuating scattering material
simulated cysts
hyperechoic and hypoechoic simulated lesions
thin scattering layer
sensitivity =
attenuating scattering material
match the parameters measured with the items used.
nylon fibers
attenuating scattering material
simulated cysts
hyperechoic and hypoechoic simulated lesions
thin scattering layer
beam profile =
thin scattering layer
match the parameters measured with the items used.
nylon fibers
attenuating scattering material
simulated cysts
hyperechoic and hypoechoic simulated lesions
thin scattering layer
section thickness =
thin scattering layer
match the parameters measured with the correlating types of observation modes.
gain settings =
compensation, dynamic range
match the parameters measured with the correlating types of observation modes.
deepest scattering material imaged =
compensation, sensitivity, dynamic range
match the parameters measured with the correlating types of observation modes.
fiber distances from the transducer or from each other on the display
range accuracy, caliper accuracy
match the parameters measured with the correlating types of observation modes.
minimum spacing of separately displayed fibers
axial resolution
match the parameters measured with the correlating types of observation modes.
lateral smearing of fibers =
lateral resolution
true or false? test objects and phantoms can be used by the instrument operator.
true
a moving ___ test object is useful in checking the accuracy of doppler spectral displays.
string
a ___ phantom is useful in simulating physiologic _____ conditions for a doppler instrument.
flow, flow
which of the following is used for doppler sensitivity measurements?
-cyst phantom
-profile test object
-string test object
-contrast phantom
-none of the above
none of the above
tissue equivalent phantoms attempt to represent some acoustic property of ____.
tissues
true or false? the string test object measures volumetric flow rate.
false
with the use of a hydrophone, which of the following can be measured or calculated?
-impedance
-amplitude
-period
-pulse duration
-pulse repetition period
amplitude, period, pulse duration, pulse repetition period
true or false? all hydrophones consist of a small element mounted on the end of a needle.
false
a needle hydrophone contains a small ____ element.
transducer, piezoelectric
true or false? because of its small size, a hydrophone can measure spatial details of a sound beam.
true
a hydrophone ____.
produces a voltage
heating depends most directly on ____.
SATA intensity
conditions under which cavitation may occur are best described by ____.
peak rarefactional pressure
bioeffects have been observed in experimental animals with intensities greater than ___.
100 mW/cm² SPTA
bioeffects have been observed in experimental animals with focused intensities greater than ____.
1 W/cm² SPTA
focal lesions have been observed in experimental animals with intensities greater than ____.
10 W/cm² SPTA
true or false? the available epidemiologic data are sufficient to make a final judgement on the safety of diagnostic ultrasound.
false
exposure is minimized by using ultrasound _____.
only when indicated, with minimum intensity, with minimum time
which of the following is (are) used currently to indicate output on the display?
-percent
-decibel
-SPTA intensity
-mechanical index
-all of the above
all of the above
which of the following affect(s) exposure of a fetus?
-intensity at the transducer
-distance to the fetus
-frequency
-gain
intensity at the transducer, distance to the fetus, frequency
true or false? there is no possible hazard involved in the use of diagnostic ultrasound.
false
true or false? ultrasound should not be used as a diagnostic tool because of the bioeffects it can produce.
false
no independently confirmed, significant bioeffects in mammalian tissues have been reported at intensities below ____.
100 mW/cm² SPTA
yes or no? is there any known risk with the current use of diagnostic ultrasound?
no
yes or no? are there any bioeffects that ultrasound produces in small animals under experimental conditions?
yes
which of the following are mechanisms by which ultrasound can produce bioeffects?
-direction ionization
-absorption
-photoelectric effect
-cavitation
-compton effect
absorption, cavitation
which of the following relates to heating?
-impedance
-sound speed
-absorption
-refraction
-diffraction
absorption
which of the following endpoints is documented well enough in the scientific literature to allow a risk assessment for diagnostic ultrasound to be based on it?
-fetal weight
-sister chromatid exchange
-fetal abnormalities
-carcinogenesis
-none
none
on which of the following endpoints has more than one epidemiologic study shown a statistically significant effect of ultrasound exposure?
-fetal activity
-birth weight
-fetal abnormalities
-dyslexia
-none
none
which of the following acoustic parameters has (have) been documented in ultrasound epidemiologic studies published thus far?
-frequency
-exposure time
-intenstiy and pulsing conditions
-scanning patterns
-none
none
a device commonly used to measure the output of diagnostic ultrasound instruments is a(n) ____
hydrophone
a typical output intensity (SPTA) for an ultrasound imaging instrument is ____.
1 mW.cm²
which of the following typically has the highest output intensity?
-fetal monitor doppler
-duplex pulsed doppler
-color doppler shift
-color power doppler
-phase array, gray scale
duplex pulsed doppler
as far as we know now, which of the following is the most correct and informative response to a patient’s question, “Will this hurt me or my baby?”
-No.
-Yes.
-We don’t know.
-The risks are well understood, but the benefits always outweigh them.
-There is no known risk with ultrasound imaging as it is applied currently.
There is no known risk with ultrasound imaging as it is applied currently.
to minimize whatever risk there may be with ultrasound imaging, which of the following should be done?
-scan to produce pictures for the family album
-scan to determine fetal sex
-minimize exposure time
-scan for medical indications only
-minimize exposure intensity
minimize exposure time, scan for medical indications only, minimize exposure intensity
which of the following controls affect instrument output intensity?
-dynamic range, compression
-transmit, output
-near gain, far gain
-overall gain
-slope, TGC
transmit, output
which of the following are correct for a duplex, pulsed wave doppler instrument?
-tissue anywhere in the doppler beam is exposed to ultrasound
-tissue anywhere in the imaging plane is exposed to ultrasound
-imaging intensities are higher than for conventional gray scale instruments
-doppler intensities are higher than for continuous wave fetal monitoring
tissue anywhere in the doppler beam is exposed to ultrasound, tissue anywhere in the imaging plane is exposed to ultrasound, tdoppler intensities are higher than for continuous wave fetal monitoring
the tissue of greatest concern regarding bioeffects in an abdominal scan is the ___.
fetus
would it be wise to substitute a duplex, pulsed wave doppler device for an inoperative fetal monitor for long term (e.g. 24 hour) monitoring in labor?
no
which of the following is (are) likely to be exposed to ultrasound during a diagnostic study?
-patient
-sonographer
-sonologist
-observers in the room
patient
no bioeffects have been observed in nonhuman mammalian tissues at thermal index values of less than ___.
2
no bioeffects have been observed in nonhuman mammalian tissues at mechanical index values of less than ___
0.4
no bioeffects have been observed in nonhuman mammalian tissues at peak rarefactional pressure values (megapascals) of less than ___.
0.4