What determines the result in urinalysis automation and what is the most important when determining? (2)
Color
Accuracy of the timing
What is the ultimate goal of the urine analysis?
Improve reproducibility and color discrimination while increasing productivity and standardization for reporting results
Automated reagent strip readers use a spectrophotometric measurement of ___________
Reflectance photometry
Semi-Automated Analyzers (lab size, calibration, abnormal result, checks)
Small and medium-volume size
Self-calibrating
Abnormal result are flagged and stored sent to the LIS
Auto-check
Fully-Automated Analyzers (lab size, abnormal result)
High-volume laboratory
Abnormal ranges to be selected so confirmatory testing and sent to the LIS
Advantge of Microscopy Analyzers (3)
Cost-effective
Improve TAT’s
Flow cytometry and digital imaging
Sysmex UF-1000 Fully Automated (methodology, urine, result, others)
Flow cytometry
1.2 mL of uncentrifuged urine
Results in quantitative cells per microliter and cells per high- or low-powered field
Independently or integrated with an automated urine chem strip
Sysmex UN-2000 Automated (methodology, core, surface, result)
Urine cytometry with digital image
Core channel: nuclear materials
Surface channel: anucleate elements
result reporting and reflex rule setting
Beckman’s Coulter’s iQ200 Microscopy (methodology, urine, result, others)
digital flow morphology and particle recognition
1.3 mL of urine or bf
results are reviewed or auto-released to the LIS
Classification categories in Beckman’s Coulter’s iQ200 Microscopy (5)
RBC’s, WBC’s, WBC clump
Epithelial cells
Casts
Crystals
Bacteria, yeast, mucus, and sperm
Electonika UriSed 2 and 3 Pro (methodology, urine, others)
automated microscopy with digital imaging
2mL of urine
3 Pro imporves in differentiation of elements
FUS-100 and FUS-200 (methodology, others)
Flat-flow cell digital imaging and neural network
Classifies 12 visible components based on shape, contrast, texture, and frequency
Clinitek AUWi System (methodology, urine, result, others)
Automated urinalysis
5mL urine
Results are verified automatically
No centifugation is required
iRICELL (methodology, urine, result)
urine chem and microscopy analyzer
4 mL urine
Results transmitted to LIS
LaUMat 2 with UriSed 2 and 3 (methodology, urine)
physical and chemical tests
4mL urine
COBAS 6500 (methodology, urine, others)
Urine chemistry analyzer
2.8mL urine
Particle recognition software determines the identification of cells
UX-2000 (methodology, urine, others)
flow cytometry
5mL urine
Physicial examination + microscopic examination
How to do body fluid analysis automation
Hemocytometer for WBC differentials
Advantages of BF analysis automation (3)
Precision
Faster TATs
Standardization of results
ADVIA 2120i (methodology, cell count results)
Flow cytometry, light scattering, and absorbance to count RBCs and WBC
WBC differential available
Sysmex XN-series (methodology, cell count results, others)
Enumerations and flow cytometry
Reports BF WBCs and total nucleated cell counts
Identifies Malignant tumor cells
Beckman Coulter’s iQ200 (cell count results)
Total cells, nucleated cells, and RBCs