1/116
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
What is fluoroscopy?
A specialized radiographic equipment used to visualize the motion of internal structures and fluids
What is fluoro referred to as?
Dynamic (moving)
What is plain radiography referred to as?
Static
Who invented the first fluoroscope and what year?
Thomas Edison in 1896
What ten things make up a fluoro unit?
fluoroscopic x-ray tube and image receptor
arm
intage intensification (II) tube
control panel
spot imaging or other imaging capability
foot exposure switch (deadman’s type)
television monitor
protective curtain
bucky slot cover
timer
Fluoroscopic x-ray tube and image receptor
Will be mounted on a C-arm to maintain their alignment at all times
What are the two types of fluoroscopic x-ray tubes?
Under-table unit
Over table unit
What is the arm that supports the equiptment over the table called?
Carriage
What in included in the control panel of the fluoro unit?
Power drive, shutters, image selection, etc
The timer on the fluoro machine makes an audible sound at what time?
5 minutes
What must be engaged for fluoro to work?
Carriage
Fluoroscopic x-ray tubes are very similar to diagnostic tubes except they are what?
Designed to operate for longer periods of time at a much lower mA
Diagnostic tubes operate at _____ mA
50-1200 mA
Fluorocopic tube range is _____ mA
.5-5 mA
Why is patient dose higher for fluoro?
Due to the decreased SOD
UGI is approximately _____ mR
700
BE is approximately _____ mR
1300
Can you use fluoro as a positioning aid?
No
What is the kVp range for barium studies (UGI, BE, etc)?
High kVp (90-120).
What is the kVp range for iodine studies (gallbladder, angiography, orthopedics, etc)?
Low kVp (60-80)
What are the two types of technology being used for fluoro today?
conventional
digital conversion
Conventional fluoroscopy
In the old days, radiologists used to look directly at the fluoroscopic screen and it caused a lot of exposure to the radiologist
What was used after people realized the damange convertional fluoroscopy had on the radiologist?
A mirror-optic system transmitted the optical image to the viewer
What year were image intensification tubes developed?
1948
With image intensification tubes, where does the image go?
To a TV monitor, spot film device, or cine camera
What does conventional fluoroscopy use to produce an analog (visible) image?
An x-ray tube, image intensifier (II), and a television system
The radiologist can obtain a permanent image with little interruption of the study by using ______ (or DR now)
Spot film device
When using a photospot camera, what is the mm film?
105
Image intensification
Utilizes the radiation that passes through the patient and enhances the image by electronically making the image 1000 to 6000 times brighter
For the maximum detail to be visible, the image brightness must be ______
High
What is the main reason that image intensifiers were created?
For increased image brightness
What are the units for illumination (brightness)?
Lumen per square meter or lux
Radiographs are visualized under illumination levels of ____-____ lux, fluoroscopy is performed at similar levels
100-1000
Before II, conventional fluoroscopy fluoroscopic screens used what phosphor?
Zinc cadmium sulfide (ZnCdS)
The image intensifier tube receives the _____ x-ray beam, converts it into _____, and _____ the light intensity
Remnant, light, increases
What is the process of image intensification?
vacuum tube with glass enelope
x-rays pass through the glass envelope and hit the input phosphor
What is the input phosphor made out of?
Cesium iodide (CsI)
What does the input phosphor do?
Converts x-rays to visible light
For image intensification, the light strikes the ________
Photocathode
What is the photocathode connected to?
The input phosphor by an adhesive layer
The photocathode is a thin metal layer made up of _____ and ______
Cesium (Cs) and antimony (Sb)
What absorbs the light energy recieved from input phosphor and emits electrons?
Photocathode
What is the process of photoemission?
When the photocathode turns light energy into electrons
The number of electrons emitted through photoemission is _____ proportional to the intensity of the light reaching it
Directly
A potential difference between the photocathode and the anode of about _______ volts is applied
25,000
Anode
A circular plate with a hole in the middle which allows the electrons to pass through the output phosphor
What do electrostatic focusing lenses do?
Focus the electrons to keep them going in the right direction
When electrons strike the ________ (located just after the anode) light is produced
Output phosphor
What is the output phosphor made out of?
Zinc cadmium sulfide (ZnCdS) crystals
What is the light (image) directed to after it is created?
A monitoring system (TV monitor, spot-film camera, etc)
What are the several occasions where information is transferred within the II?
from the x-ray beam to the input screen,
from the input to the photocathode,
from the photocathode to the electron beam,
from the electrons to the output screen,
from the output to the human eye
Each photoelectron that hits the output phosphor produces ___-___ times as many light photons
50-75
The gain in brightness achieved by the II is a result of the ___________ and __________
Electron acceleration, image minification
A typical II has a flux gain of about _____
50
What is flux gain?
The ratio of the number of light photons at the output phosphor to the number of x-rays at the input phosphor
What is the equation for flux gain?
number of light photons output/ number of x-ray photons input
The image produced on the input screen is reproduced as a ______ image on the output screen because the output screen is so much smaller than the input screen
Minified
What is minification gain?
The increase in brightness caused by the electrons being focused into a small area
What is the equation for minification gain?
(input screen diameter/output screen diameter)²
The ____ increases the illumination level- “brightness gain”
II
Brightness gain=
minification gain x flux gain
THe brightness gain of most IIs is _____ to _____-this decreases with tube age
5,000-30,000
As your tube gets older, patient dose ______ to maintain brightness
Increases
Input phosphors (screens) range from ____” (15 cm) to ____” (30 cm) in diameter
6, 12
A typical output phosphor has a diameter of ___” (2.5 cm)
1
Most image intensifiers are of the multifield type, what can they be called?
dual-focus image intensification
dual-field inage intensification
tri-focus image intensification
tri-field image intensification
What is the most common design for multi-field image intensification?
10/7” (25/17) dual focus
On a 6” mode for multi-field image intensification, the voltage of the focusing lenses is increased and only the inner ____ inches of the input phosphor is used
6
The electron focal point is moved further from the output phosphor, causing what?
Magnification
Using the 6” mode reduces the field of view and thereby ______ the image
Magnifies (mag mode)
On a 9/6 II, the ^” mode is ______ times larger than the 9” mode
1.5
Mag mode will _____ magnification, ______ resolution (spatial and contrast), minification gain is _____ (less brightness), and mA will need to ______
Increase, increase, reduced, increase
The area of the input phosphor not used is still hit with x-rays, this creates a loss of brightness at the periphery of the image called?
Vignetting
For fluoroscopic control, what does ABC stand for?
Automatic Brightness Control
For fluoroscopic control, what does ABS stand for?
Automatic Brightness Stabilization
What does Automatic Brightness Control or Automatic Brightness Stabilization do?
Maintains the brightness of the image by automatically adjusting the exposure factors (mA and kVp) according to the density and contrast of the subject
What are the two types of digital fluoroscopy being used today?
charged coupled device (CCD)
flat panel
What is the advantages of digital fluoroscopy over conventional fluoroscopy?
no cassettes are required
requires less dose
speed of image acquisition and the ability to post process
In digital fluoro, the image-intensifier output screen image is coupled via?
Charged-coupled device
What does a CCD do?
Converts visible light to an electrical charge that is sent to the ADC for processing
The CCD’s rapid discharge time virtually eliminates image _____ and is particularly useful in high-speed imaging procedures such as cardiac catheterizations
Lag
CCD camera have replaced what?
Analog cameras
CCDs are more sensitive to light emitted by the output phosphor and are associated with less ______
Noise
CCD still uses an II, and uses a layer of ______ silicon
Crystalline
CCD is mounted where?
On the output phosphor of the II
Flat panels are ______ and _____ than the II and CCD for easier manipulation
Smaller and lighter
What are flat panels composed of?
Cesium iodide
For image monitoring and recording, what are the systems?
optical monitoring (mirrors)
cinefluoroscopy
spot filming
television (video) monitoring
video recording
Optical (mirror-optics) system
system of lenses and mirrors
magnifies the image from the output phosphor onto a viewing glass
the field of view is small
only one person can use it at a time
a significant amount of light is lost
Cinefluoroscopy (cinecamera)
a movie camera records the images on film
primarily used in cardiac catheterization
patient dose is very high
image quality is high
What mm film cameras are used for cinefluoroscopy?
16 or 35 mm film cameras
35 mm cameras in cinefluoroscopy has _______ quality and ______ dose than 16 mm
Higher
Cine cameras are driven by _________ motors (60 Hz)
Synchronous
What are the framing frequencies in cinefluoroscopy?
7.5, 15, 30, 60, and 90 frames per second
Cine cameras are different from a roll film photospot camera; the major difference are that the film is smaller (35 mm wide) and longer (400 ft on a roll) and that the camera is capable of operating at up to _______ images per second
90
Increased framing frequency will _______ patient dose
Increase
Spot filming
uses the same technology as conventional screen-film radiography
a cassette with regular radiographic film/CR is positioned between the patient and II
when fluoro is on, the cassette is positioned away from the II
Spot film devices usually allow _____ than one image to be obtained on a single film
More
Spot film devices produce ______ patient dose
High
What is a disadvantage of conventional spot film devices?
The delay involved in moving the cassette into position for exposure