Comprehensive practice vocabulary flashcards covering wastewater laboratory procedures, chemical storage, holding times, and microbiological testing standards.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
Sulfuric Acid
The titrant used in the determine of alkalinity in water.
Calcium Bicarbonate
The chemical in terms of which alkalinity is usually expressed as an equivalent amount.
Alkalinity
The acid-neutralizing capacity of water.
Phenolphthalein alkalinity titration endpoint
The point at which a sample of suitable size is titrated with a standard acid to a pH of 8.3.
Total alkalinity titration endpoint
The point at which a sample of suitable size is titrated with standard Acid solution to a pH of 4.5.
Methyl orange
An indicator used in the laboratory to test for alkalinity.
Desiccator
An apparatus used in the lab for maintaining a dry environment.
Volumetric glassware calibration temperature
The temperature at which volumetric glassware is generally calibrated, which is 20∘C.
Autoclave (complex sugars)
Used to sterilize media containing complex sugars by operating at 121∘C.
Chlorine residual determination timing
A test that must be run immediately after sampling to avoid dissipation.
Coliform organisms (prevalence)
Microorganisms most prevalent in the intestinal tract of man and other warm-blooded animals.
Sodium thiosulfate (timing)
Added to bacteriological sample bottles before sterilization.
Bacteriological sample bottles
Should be sterilized by the lab and ready to use when received.
Sodium Thiosulfate (purpose)
A chemical added to bacteriological sample bottles to remove residual chlorine.
Grab Sample
A single sample taken without considering the time or the rate of flow.
Sample refrigeration preservation temperature
The most accepted method of sample presentation, listed at 121∘C.
Dechlorinate
The process of adding Sodium Thiosulfate to chlorinated samples.
Chlorinated bacteriological sample bottles
Bottles used for collection that usually contain Sodium Thiosulfate.
Composite Sample
A sample built up from a series of samples taken at intervals during a fixed sampling period.
Grab samples (representation)
These represent instantaneous conditions at the time the sample was collected.
Nitric Acid (HNO3)
The chemical that should be added to sample bottles before collecting for iron and manganese determinations.
Non-preserved DO sample holding time
Requires the analyst to Analyze Immediately.
Total Dissolved Solids drying temperature
A constant weight is achieved by drying the sample at 180∘C±2.0∘C.
Total Dissolved Solids drying oven duration
Sample is placed in a drying oven set at 180∘C for at least 1 hour.
Conductivity
The lab procedure from which total dissolved solids can be estimated.
Hardness values
Expressed as equivalent to Calcium Carbonate.
Fecal coliform membrane filter incubation
Incubated in a water bath at 44.5∘C±0.02∘C.
Total coliform cultures (incubation)
Cultures should be incubated at 35.0±0.5∘C for 24HRS.
Bacteriological test units
Results are expressed as Coliforms per 100ml.
EC Medium
The medium used in the multiple tube fermentation technique for fecal coliforms at 44.5±0.2∘C.
Pathogenic
Bacteria that can cause disease.
De-ionized water conductivity
The conductivity should be less than 0.2μmhos / cm.
Chemical storage safety
All chemicals should be stored according to manufacturers recommendations.
Sulfuric acid storage
Should not be stored with Potassium Permanganate.
Oil and Electrical fires (extinguisher)
Carbon Dioxide is the best type of fire extinguisher to use.
Phenolphthalein
Used as an indicator in the process of titration.
Buffer
Solutions that oppose a change of composition, especially of hydrogen-ion concentration.
Normality
Gram equivalent-weights of reagent chemical per liter of solution.
Sodium hydroxide
A chemical that will make water more alkaline.
Chlorine residual maintenance
Maintained to protect the bacteriological quality of the water.
Total suspended solid oven temperature
The oven should be between 103∘C−105∘C.
Glass Fiber
The type of filter used in the total suspended solids lab procedure.
MPN
Most Probable Number of Organisms.
Oxidizer
A category of chemical exemplified by HNO3 (Nitric Acid).
Phosphorus pipet cleaning
Wash with HCl then water before using.
Nitrification
The process described as Ammonia $\rightarrow$ Nitrites $\rightarrow$ Nitrates (NH3, NO2, NO3).
Cadmium Reduction
Involves the conversion of Nitrates to Nitrites.
Stock solution quantity
The correct amount to keep is 700 of the stock.
Manhole gas safety
Check for gases before opening.
Conductivity (measures)
Measures electrical current and also Total dissolved solids.
Specific gravity (1.20g/cm3)
Indicates that a compound is 20% heavier than water.
H2SO4
The titrant specifically used in alkalinity determinations.
Ammonia release pH
Ammonia is released at a pH>11.
Digestion
Decomposing a sample using high heat and acidic conditions.
Nephelometer
An instrument where light is measured at an angle of 90∘.
Chemical storage (standard)
Should be kept in their appropriate container as indicated.
Control curve points
Three points are needed to construct a control curve.
Oil and grease analysis sample
Utilizes a Grab sample.
Eye wash and shower station maintenance
Analysts need to check these stations in intervals.
Acid dilution safety
Always add acids to water.
Phosphorus test digestion product
This process produces Orthophosphates.
Membrane Electrode
The most common type of portable DO Meter.
BOD Incubation temperature
Incubated at 20∘C±1∘C.
Type C fire extinguisher
Uses Carbon Dioxide to fight oil and electrical fires.
Ammonia borate buffer adjust
Ammonia is adjusted to a pH of 9.5 with this buffer.
OSHA
The organization that provides regulation for confined spaces.
Alkalinity phenolphthalein pH endpoint
The pH endpoint is 8.3.
Code of Federal Regulations 40 136
The place one should look to find approved methods for measuring.
Total solids temperature
Total solids are ran at 180∘C.
Ammonia holding time
A 28-day hold.
COD holding time
A 28-day hold.
Metals holding time
An 80-day hold for most metals.
Fecal holding time
A 6 Hours hold.
Nitrate-Nitrite holding time
A 28-day hold.
Oil & Grease holding time
A 28-days hold.
Phenyl arsine oxide
The titrant used for amperometric titration of free chlorine.
Coliforms (indicator)
Indicators for Pathogenic Organisms.
Dosage Calculation
Dosage=Demand+Residual.
Silver Sulfate
Used as a catalyst in COD reactions.
Filamentous bacteria (microscope)
Read using a magnification of 100x.
TKN
Total kjeldahl nitrogen, composed of Organic Nitrogen and Ammonia.
Amines (Ammonia & Chlorine)
The most common form of interference in ammonia readings.
Distilled water pre-treatment
Important because it minimizes build up on equipment.
Lab grade water filter size
A 0.2μm filter should be used to produce contaminant free water.
Fecal coliform incubation temperature
Incubated at 44.5∘C±0.2.
Water bath (fecal coliform purpose)
Used during incubation to provide a consistent incubation.
95% Ethanol
The purity of ethanol used to prepare some standards.
Residual chlorine hold time
A hold time of 15min.
Fume hood volatile threshold
Utilized when a concentration of volatiles X>150ppm is present.
Portable air blowers
Equipment used to help ventilate man holes prior to entry.
1000 mg
Equal to 1g.
Incineration
An example of a temperature dependent reaction.
Activated carbon
Works through the process of Absorption.
Broken control point protocol
If an analyst breaks a control point, they should identify potential problems and reanalyze.
Total metals preparation
Acid Wash (1:1HCl); Rinse with reagent grade DI water; Air Dry.
Analytical reagent grade
The primary indicator that should be look for when buying stock chemicals.
Buffers (buying concern)
The primary concern is the shelf life.
Chemical box labelling
Must indicate the date received and the date opened.
BOD sample setup
Set up in a BOD bottle of 60ml or greater.
Ammonia testing stirring
Samples should be tested for ammonia with constant stirring.