Mobile Radiography, Fluoroscopy and Digital Fluoroscopy

0.0(0)
studied byStudied by 0 people
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
Card Sorting

1/64

flashcard set

Earn XP

Description and Tags

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

65 Terms

1
New cards
for portables, they automatically engage the highest ___ available
mA
2
New cards
For 100 cm (40"), average person can use
fingertip-to-opposite axilla
3
New cards
From 100 cm (40"):
125 cm (50"):
150 cm (60"):
180 cm (72"):
125 cm: 1 ½ times increase in technique
150 cm (60"): Double technique
180 cm (72"): Triple technique
4
New cards
For mobile chests, always check the angle of the tube from ____ perspectives, while standing back as far as possible
two
5
New cards
If the anatomy of interest includes joint spaces, it is always more important to keep the CR perpendicular to the ____ rather than to the cassette or IR
anatomy
6
New cards
do not ever angle CR if looking at
fluid levels
7
New cards
if your pt is sitting at 20 degrees from vertical and back of bed is 10 degrees from vertical, how much angle for CR if looking for fluid levels
angle CR horizontal
8
New cards
For mobile radiography of large body parts, use ___ ratio grids for wide positioning latitude when grids are used
low
9
New cards
recommended approach to technique with portables is to have
manual technique chart
10
New cards
fluoroscopy was developed to radiologists could view ____ images
dynamic
11
New cards
image intensifier improved fluoroscopy by
• Improved fluoroscopy by increasing image brightness
Dramatically reduced the mA required, saving patient dose
12
New cards
fluoroscopic tubes operate between
1-5 mA
13
New cards
Due to low exposure rate, fluoroscopic images are statistically inferior to overhead images and demonstrate more
noise
14
New cards
ABC
automatic brightness control
15
New cards
purpose of image intensifier tube
Convert the remnant radiation beam into an amplified light image
16
New cards
The output phosphor (left, arrow) is always ___ cm in diameter
2.5
17
New cards
Concentration of the beam from the input to the output phosphor increases ___ by thousands of times
brightness
18
New cards
Increasing the physical diameter of the input phosphor results in increased
image resolution
19
New cards
the input phosphor converts x-rays into
light
20
New cards
Upon absorbing light energy, the photocathode emits
electrons
21
New cards
Concentrating the electrons increases output intensity. This process is called
minification
22
New cards
A thin photoemissive metal layer, made of a combination of antimony and cesium compounds
photocathode
23
New cards
A layer of fluorescent crystals of cesium iodide
input phosphor
24
New cards
A series of positively-charged metal rings that become narrower toward the anode end of the tube
The Electrostatic Focusing Lens
25
New cards
Made of cesium iodide
Converts the kinetic energy of the electrons striking it into light
output phosphor
26
New cards
The ratio of the number of light photons emitted by the output phosphor to the number of X-rays striking the input phosphor
flux gain
27
New cards
as output phosphor gets smaller, what happens to minification gain?
decreases
28
New cards
The product of the minification gain and the flux gain
brightness gain
29
New cards
a smaller region of the input phosphor is projected onto the output phosphor by moving the what where?
focal point of the electron beam closer to the input phosphor
30
New cards
Magnification mode can be used to improve what feature in the image?
spatial resolution or sharpness in the image
31
New cards
To avoid the appearance of noise from a poor SNR, the what increases the what to compensate for the loss of minification gain
Automatic Brightness Control
mA
32
New cards
Patient exposure increases by the ratio of the
change in area of the active input phosphor used
33
New cards
Using magnification mode in fluoro has what bad quality?
increases patient dose
34
New cards
As the patient rolls onto her side, the increased tissue thickness must be_____ to maintain fluoroscopic image brightness
compensated for
35
New cards
brightness of a fluoroscopic is _____ to the mA
directly proportional
36
New cards
brightness of a fluoroscopic image is to the what power of kvp
5th power
37
New cards
Generally, what kind of technique is preferred with fluoro?
high kVp / low mA
38
New cards
No form of electronic intensification can improve the image above the
statistical level of the absorbed x-ray photons
39
New cards
no amount of electronic intensification can overcome the mottle or noise produced when an
insufficient amount of x-rays is reaching the image intensifier
40
New cards
Small, momentary defects in the distribution of brightness ("twinkling") in a dynamic image

This noise pattern appears to be drifting across the image, creating the appearance of "crawling ants" or "snow"
scintillation
41
New cards
what is the best solution to improve scintillation
raise the mA
42
New cards
Image intensifiers are generally poor at
preserving high contrast in the transmitted image
43
New cards
Contrast deteriorates with aging of the intensifier -deterioration can be as high as ____% per year
10
44
New cards
Slight magnification and increased blur around the periphery of the image
pincushion distortion
45
New cards
Scattered random light from the output phosphor
veiling glare
46
New cards
Loss of brightness toward the periphery of the image
vignetting
47
New cards
A contributing factor to vignetting is
pincushion distortion
48
New cards
High-resolution TV cameras that use an ADC to turn the electronic signal into digital numbers
Digital Photospot Cameras
49
New cards
what piece of equipment in the OR allows for significant reductions in patient exposure
Carbon fiber tables
50
New cards
A series of short visual checks, rather than continuous fluoroscopy, dramatically reduces cumulative exposure to the patient
Intermittent Fluoroscopy
51
New cards
This is the most important way to limit patient exposure during fluoroscopic studies
Intermittent Fluoroscopy
52
New cards
Keeping the last image displayed on the monitor
"Last Image Hold" Devices:
53
New cards
feature that automatically turns the x-ray beam on and off during operations
pulsed fluoroscopy
54
New cards
with Pulsed-Progressive Mode, the fluoro tube is being run in
radiographic mode
55
New cards
refers to the length of each pulse
pulse width
56
New cards
refers to how many pulses occur per second of operation
pulse rate
57
New cards
if pulse width is reduced from 6 to 3, how much is patient dose decreased?
1/2
58
New cards
The _____ detector uses a cesium iodide phosphor coupled to an active matrix array (AMA) of amorphous silicon thin-film transistors (TFTs)
indirect
59
New cards
The ____ detector uses an AMA of amorphous selenium TFTs
direct
60
New cards
Dual-Use FPD systems convert from radiographic to fluoroscopic mode by grouping sets of 4 small pixels together to form a larger effective pixel
binning
61
New cards
DFPDs can be operated either in
continuous or pulsed fluoroscopy mode
62
New cards
• Changing from fluoroscopic mode to DF mode requires an increase in tube current of 20 to 100 times the mA

• Nonetheless, a net decrease in patient exposure can be achieved compared to continuous fluoroscopy because why
the extremely short pulse times for exposures
63
New cards
The time it takes for the x-ray tube to be switched on
Interrogation Time:
64
New cards
The time it takes for the x-ray tube to be switched off
Extinction Time
65
New cards
Grids for DFPDs have their lead lines running in which direction to prevent aliasing artifacts
diagonal