Coulometry
total charge
primary use - absolute quantification of substance amount
chloride in bodily fluid
Voltammetry
current as function of varying voltage
primary use - identification and quantification of multiple species
heavy metal analysis, redox behavior studies
Potentiometry
voltage between electrodes
primary use - direct measurement of ion activity and concentration
pH measurement, ion selective electrodes for blood electrolytes
Ion Selective Electrodes
membrane-based electrochemical transducers capable of responding to specific ion
Direct ISE
requires no sample dilution
determines activity of ions in plasma water
lipids and proteins have no effect on electrolyte measurement
Indirect ISE
requires sample dilution
interference by hyperlipidemia and hyperproteinemia
reference range 7% higher than direct ISE
Na+ electrode
silicate glass membrane that is permeable to Na ions and excludes other cations
K+ electrode
valinomycin antibiotic membrane that binds K+ and effectively eliminates interference by Na+ and other ions
Ca2+ electrode
calcium di-phosphate membrane that selectively binds calcium ions
Mg2+ electrode
N'N"N-imn-6,1-hexandiyl-tris membranes that specifically interact with Mg2+ ions
pH Measurement
Potential difference is measured and related to ion activity by the Nernst equation
pO2 Measurement
amperometric Clark electrodes
pCO2 Measurement
CO2 is measured by detecting pH changes as CO2 gas diffuses across a gas-permeable membrane