What serves as the primary protective barrier in fluoroscopic imaging?
Image intensifier or a flat panel detector
A primary protective barrier of _____ lead equivalent is required for a fluoroscopic unit.
2 mm
The input phosphor is typically composed of _______ and converts ______ to _______.
cesium iodide; x-rays; light
A thicker input phosphor layer has a (higher or lower) absorption efficiency and (Increaseses or decreases) patient dose.
higher; decreases
What effect does a thicker input phosphor have on spatial resolution?
decreases
The photocathode converts ____ to _____
light to electrons
What charge do the electrostatic lenses have?
negative
A potential difference of ____ is maintained between the photocathode and the anode.
25,000 volts
The output phosphor is typically composed of _____ and converts _____ to ______.
zinc cadium sulfide; electrons to light
How much the image is minified from the input phosphor to the output phosphor.
flux gain
brightness gain
minification gain
minification gain
The ability of the image intensifier to convert xrays into light and increase brightness.
flux gain
brightness gain
minification gain
brightness gain
The ratio of light photons at the output phosphor to the number of light photons emitted from the input phosphor (converting a few xrays into many light photons)
flux gain
brightness gain
minification gain
flux gain
Magnification results in (higher/lower) patient dose, (better/worse) contrast resolution, (better/worse) spatial resolution
higher; better; better
_____ is a reduction of brightness at the periphery of the image.
vignetting
______ distortion is a result of the inaccurate control of electrons released at the periphery of the photocathode and the curved shape of to photocathode.
pincushion
Compared with the image on the input phosphor, the image on the output phosphor of the image intensifier is ( select three):
magnified
distorted
brighter
dimmer
inverted
minified
brighter, inverted, minified
As field of view (FOV) decreases in image intensification (select three):
output screen image is magnified
output screen image has more noise
mA increases
output screen image has improved resolution
the fical point is closer to the output phosphor
voltage to the focusing lenses is decreased
output screen image is magnified, mA increases, output screen image has improved resolution
Which of the following will improve the spatial resolution of image intensified images?
a very thin input phosphor
a larger diameter input screen
increased brightness pain
1 only
Which of the following systems functions to compensate for changes in patient/part thickness, FOV, and OID during fluoroscopic procedures?
automatic exposure rate control
electronic magnification
automatic resolution control
flux gain
automatic exposure rate control
Which of the following modes of a triple-field image intensifier will result in the highest patient dose?
25 inch mode
17 inch mode
12 inch mode
diameter does not affect patient dose
12 inch mode
All of the following are components of the image intensifier, except:
the photocathode
the focusing lenses
the TV monitor
the accelerating anode
the TV monitor
Which of the following is used in fluoroscopy, replacing the image intensifiers’ television camera tube?
solid-state diode
charge coupled device
photostimulable phosphor
vidicon tube
charge coupled device
In fluoroscopy, the automatic brightness control is used to adjust the:
kilovoltage (kV) and the milliamperage (mA)
backup timer
milliamperage (mA) and time
kilovoltage (kV) and time
kilovoltage (kV) and the milliamperage (mA)
The image-intensifier tube’s input phosphor functions to convert:
kinetic energy to light
x-rays to light
electrons to light
fluorescent light to electrons
x-rays to light
The total brightness gain of an image intensifier is the product of:
flux gain
minification gain
focusing gain
1 and 2

Which number in the figure to the right represents magnification mode?
1
2
7
8
8

Which number in the figure represents electrostatic focusing lens?
5
6
4
2
4
Advantages of CCDs over the use of television cameras in image intensification include which of the following?
compact size
improved resolution
higher DQE
1, 2, and 3
The image intensifier’s input phosphor generally is composed of?
cesium iodide
zinc cadmium sulfide
gadolinium oxysulfide
calcium tungstate
cesium iodide
The brightness level of the fluoroscopic image can vary with:
milliamperage
kilovoltage
patient thickness
1, 2, and 3
Moving the image intensifier closer to the patient during fluoroscopy:
decreases the SID
decreases patient dose
improves image quality
1, 2, and 3
Which of the following is/are associated with magnification fluoroscopy?
higher patient dose than non-magnification fluoroscopy
higher voltage to the fucusing lenses
image intensifier focal point closer to the input phosphor
1, 2, and 3
The device that receives the remnant beam, converts it into light, and then increases the brightness of that light is the:
charge coupled device (CCD)
spot image device
image intensifier
television monitor
image intensifier
When using the smaller field in a dual-field image intensifier:
the image is magnified
the image is brighter
a larger anatomic area is viewed
1 only
Advantages of smaller diameter input screens in image intensification include which of the following?
improved resolution
availability of large FOV
image magnification
1 and 3
To maintain image clarity in an image-intensifier system, the path of electron flow from the photocathode to the output phosphor is controlled by:
the accelerating anode
electrostatic lenses
the vacuum glass envelope
the input phosphor
electrostatic lenses

What is A?
indirect flat panel

What is B?
direct flat panel

What is C?
cesium iodide

What is D?
amorphous silicon

What is E?
amophous selenium
The ____ signal is the electronic signal which carries information from the charged coupled device to the TV monitor?
uptake
coupling
video
monitor
video
The diameter or size of the _____ is the value used to describe the size of the image intensifier.
output phosphor
photocathode
input phosphor
anode
input phosphor
Which one of the following modes is an image intensifier would produce an image with the most magnified image?
5 inch mode
9 inch mode
12 inch mode
14 inch mode
5 inch mode
The reduction of field of view (FOV) size, 24 cm to 12 cm, in a dual field image intensifier can be accomplished by:
increasing the size of the input phosphor
decreasing the size of the output phosphor
increasing the voltage on the electrostatic lenses
decreasing the potential difference across the tube
increasing the voltage on the electrostatic lenses
Which mode of a triple field image intensifier will have the highest minification ratio gain factor?
5 inch mode
9 inch mode
12 inch mode
14 inch mode
14 inch mode
The reduction of brightness at the periphery of image display of an image intensified fluoroscope is termed:
pincuchion reduction
vignetting
fluctuation reduction
brightness falloff
vignetting
The principal purpose of employing image-intensified fluoroscopy is to:
enlarge the image for better viewing
raise the level of image brightness
increase the patient dose during fluoroscopy
act as a means of storing the image
raise the level of image brightness
The image intensifier tube is a/an ______ tube.
gas filled
gas evacuated
oil filled
fluid filled
gas evacuated
During a conventional fluoroscopic examination, the fluoroscopic milliamperage is:
less than 5 mA
10 to 20 mA
50 to 75 mA
100 to 150 mA
less than 5 mA
What is the total brightness gain for an image intensifier tube with a minification gain of 50 and a flux gain of 75?
500
1020
1550
3750
5000
3750
The device in the image intensifier tube that converts x-rays into light is the:
input phosphor
photocathode
accelerating anode
output phosphor
input phosphor
Most modern image intensifier tubes utilize an output phosphor composed of:
barium lead sulfate
zinc cadmium sulfide
calcium tungstate
cesium iodide
zinc cadmium sulfide
At least how many monitors are required for digital fluoroscopy?
1
2
3
4
2
A charge-coupled device used in digital fluoroscopy provides high:
signal-to-noise ratio
spatial resolution
detective quantum efficiency
patient dose
1, 2, and 3
Dynamic radiography is:
radiography which produces images of the body at one point in time
recording on a radiograph only selected layers within the body
radiography in motion
procedures which produce depth perception
radiography in motion
THe increased brightness of an image due to accelerated electrons traveling across to the output phosphor (and convertinga few xrays into many light photons) is called:
image gain
minification gain
total brightness gain
flux gain
flux gain
The detector used for direct capture in digital radiography/fluoroscopy is made up of:
cesium iodide + silicon photodiodes
amorphous selenium
CCD sensors
barium fluorohalide
amorphous selenium
Brightness gain for an image intensifier is typically in the range of:
50 - 75
100 - 1,000
3,000 - 4,000
5,000 - 30,000
5,000 - 30,000
Which of the following components found in digital fluoroscopy is NOT found in conventional fluoroscopy?
a video monitor
analog-to-digital converter (ADC)
automatic brightness control (ABC)
an image intensifier
analog-to-digital converter (ADC
The purpose of a “scintillator” in a flat-panel receptor is to:
absorb light photons and emits x-ray photons
absorb x-ray photons and emit light photons
absorb electrons and emit light photons
absorb light photons and emit electrons
absorb x-ray photons and emit light photons
What directly influences the number of electrons collected in a detector element (DEL)?
exposure to image receptor
exposure field recognition
threshold detection level
histogram analysis
exposure to image receptor
How magnifies is the image of a 30/20/10 in the 20 cm mode compared to the 30 cm mode?
0.75 times
1.65 times
1.5 times
3.0 times
1.5 times
How much higher is the patient dose of a 30/20/10 in the 20 cm mode compared to the 30 cm mode?
0.13 times higher
2.25 times higher
5.3 times higher
7.7 times higher
2.25 times higher
A high signal-to-noise ratio (SNR) provides an image with:
poor spatial resolution
higher spatial resolution
poor contrast
higher distortion
higher spatial resolution
During digital fluoroscopy the x-ray beam is typically:
continuous and at higher mA
continuous and at lower mA
pulsed
produced by a TV camera tube
pulsed
The use of an image receptor with ____ DQE can result in _____ patient dose.
low; lower
high; lower
high; higher
low; tripled
high; lower
Which one of the following modes in an image intensifier would produce an image with the best/highest spatial resolution?
15 cm
23 cm
30 cm
43 cm
15 cm
During digital fluoroscopy:
the x-ray beam is continuous and at higher mA
the x-ray beam is continuous and at lower mA
the x-ray beam is pulsed
the x-ray beam is produced by a TV camera tube
the x-ray beam is pulsed
In the image intensifier tube, total brightness gain is a product of what?
total light emitted at photocathode
diameter of input phosphor times diameter of output phosphor
intensification factor: brightness without an image intensifier divided by brightness with an image intensifier
flux gain times minification gain
flux gain times magnification gain
Which of the following measures the efficiency of a system to convert the x-ray input signal into a useful output signal?
modulation transfer function
detective quantum efficiency
exposure latitude
spatial resolution
detective quantum efficiency
The minification gain of an image intensifier increases with increasing:
output phosphor size
kVp
input phosphor size
mAs
input phosphor size
The _____ signal is the electronic signal which carries information from the charged coupled device to the TV monitor.
uptake
coupling
video
monitor
video
The ____ in the image intensifier emits electrons when it is stimulated by light photons.
input phosphor
output phosphor
photocathode
electron gun
photocathode
Compared with the image on the input phosphor, the image on the output phosphor of the image intensifier is (select 3):
magnified
brighter
distorted
dimmer
inverted
minified
2, 5, and 6
A field of view (FOV) decreases in image intensification (select 3):
output screen image is magnified
output screen image has more noise
mA increases
output screen image has improved resolution
the focal point is closer to the output phosphor
voltage to the focusing lenses is decreased
1, 3, and 4
Which of the following will improve the spatial resolution of image-intensified images?
a very thin input phosphor layer
a larger diameter input screen
increased total brightness gain
1 only
Which of the following systems functios to compensate for changes in patient/part thickness, FOV and OID during flat-panel detector fluoroscopic procedures?
automatic exposure rate control
electronic magnification
automatic resolution control
flux gain
automatic exposure rate control
All of the following are components of the image intensifier, except:
the photocathode
the focusing lenses
the TV monitor
the accelerating anode
the TV monitor
Which of the following is used in digital fluoroscopy, replacing the image intensifiers television camera tube?
solid-state diode
charge coupled device
photostimulable phosphor
vidicon tube
charge coupled device
In fluoroscopy, the automatic brightness control is used to adjust the:
kV and mA
backup timer
mA abd time
kV and time
kV and mA
The image-intensifier tubes input phosphor functions to convert:
kinetic energy to light
x-rays to light
electrons to lige
fluorescent light to electrons
x-rays to light
The total brightness gain of an image intensifier is the product of:
flux gain
minification gain
focusing gain
1 and 2
Advantages of CCDs over the use of television cameras in image intensification include which of the following?
compact size
improved resolution
higher DQE
1, 2, and 3
The image intensifiers input phosphor generally is composed of?
cesium iodide
zinc cadmium sulfide
gadolinium oxysulfide
calcium tungstate
cesium iodide
The brightness level of the fluoroscopic image can vary with:
milliamperage
kilovoltage
patient thickness
1, 2, and 3
Moving the image intensifier closer to the patient during fluoroscopy:
decreases the SID
decreases patient dose
improves image quality
1, 2 and 3
Which of the following is/are associated with magnification fluoroscopy?
higher patient dose than non-magnification fluoroscopy
higher voltage to the focusing lenses
image intensifier focal point closer to the input phosphor
1,2, and 3
The device that received the remnant beam, converts it into light, and then increases the brightness of that light is the:
charge coupled device
spot image device
image intensifier
television monitor
image intensifier
When using the smaller field in a dual-field image intensifier:
the image is magnified
the image is brighter
a larger anatomic area is viewed
1 only
Advantages of smaller diameter input screens in image intensification include which of the following?
improved resolution
availability of large FOV
image magnification
1 and 3
To maintain image clarity in an image-intensifier system, the path of electron flow from the photocathode to the output phosphor is controlled by:
the accelerating anode
electrostatic lenses
the vacuum glass envelope
the input phosphor
electrostatic lenses
The fixed spatial resolution characteristic of direct digital systems is determined by the:
TFT and DEL size
sampling frequency
laser scanner
PSP thickness
TFT and DEL size
The components of a detector element (DEL) include which three of the following?
sensing area
TFT
capacitor
ADC
CCD
1, 2 and 3
The part of the detector element (DEL) that stores electrical charges is the:
capacitor
TFT
sensing area
CCD
capacitor
The part of the detector element (DEL) that functions as a switch/gate to release charges for readout is the:
capacitor
TFT
sensing area
CCD
TFT
As window width increases:
contrast scale increases
contrast scale decreases
brightness increases
brightness decreases
contrast scale increases
In digital imaging, thin film transistor (TFT) and detector element (DEL) size is related to:
contrast
brightness
spatial resolution
plate size
spatial resolution
The ratio of the pixel/DEL’s sensing area of the pixel/DEL itself refers to:
fill factor
DQE
spatial reolution
sign-to-noise ratio
fill factor
The main diffrence between a direct-capture and an indirect-capture DR panel is that:
direct capture has no scintillator
direct capture uses a photostimulable phosphor
with direct capture, light is detected by CCDs
with direct capture, light is detected by TFTs
direct capture has no scintillator