Flashcards testing vocabulary, concepts, formulas, artifacts, and dosimetric indices from the lecture on CT performance physics, image quality, and quality assurance.

Last updated 9:07 PM on 8/27/26
Name
Mastery
Learn
Test
Matching
Spaced
Call with Kai
Chat

No analytics yet

Send a link to your students to track their progress

49 Terms

1
New cards

CT Internal Alignment Lights

Patient positioning lights inside the CT scanner gantry that explicitly identify the tomographic plane.

2
New cards

CT External Alignment Lights

Laser or tungsten lights that identify the sagittal and coronal planes, plus a transaxial plane positioned at a fixed distance from the tomographic plane.

3
New cards

Scan Projection Radiography (SPR)

A planar scout scan used in CT to define the beginning and end boundaries of the target scanned volume.

4
New cards

Image Noise

The random fluctuation of CT numbers around a mean value when imaging a uniform material, caused primarily by the statistical nature of X-ray photon production and interaction.

5
New cards

Quantum Noise

X-ray photon noise modeled as a Poisson process where signal is proportional to NN, noise is proportional to \root%1%2{%1}, and signal-to-noise ratio equals \root%1%2{%1}, where NN is the number of absorbed photons.

6
New cards

Normalized Standard Deviation (SS)

A normalized parameter defined in IAEA CT QA guidelines to compare noise across CT scanners with different contrast scales: S=σwaterCTscale×100%S = \frac{\sigma_{water}}{CT_{scale}} \times 100\%, where CTscale=CTwaterCTairCT_{scale} = CT_{water} - CT_{air}.

7
New cards

Contrast to Noise Ratio (CNR)

A quantitative parameter representing low-contrast detail detectability, defined as CNR=CT1CT2σ1+σ22CNR = \frac{CT_1 - CT_2}{\frac{\sigma_1 + \sigma_2}{2}}.

8
New cards

Spatial Resolution

The capability of an imaging system to distinguish small, high-contrast objects, evaluated subjectively via visual phantom inspection or objectively via Modulation Transfer Function (MTF).

9
New cards

Modulation Transfer Function (MTF)

An objective measure of spatial resolution derived from the point spread function (PSF) or edge spread function (ESF), typically evaluated at 50%50\% and 10%10\% frequency values.

10
New cards

Streak Artifact

Single or multiple black or white line patterns that partially or completely obscure a CT image.

11
New cards

Windmill Artifact

A rotating star-like streak artifact in multislice helical CT that appears around high-density objects due to helical interpolation and reconstruction processes.

12
New cards

Shading Artifact

A CT artifact appearing as a dark band between dense bone structures, such as in the base of the skull, across the shoulders, or between femoral heads.

13
New cards

Cupping Artifact

An image artifact characterized by artificially reduced CT voxel values near the centre of a homogenous object.

14
New cards

Ring Artifact

A hardware-related artifact displaying single or multiple concentric rings centered on the rotation axis due to miscalibration or failure of CT detector elements.

15
New cards

Motion Artifact

A patient-based artifact resulting from voluntary or involuntary movement during scanning, manifesting as blurring, streaking, or shading.

16
New cards

Metal Artifact

Severe streaking artifacts caused by metal objects (e.g., dental fillings, prosthetics) whose extreme density exceeds the attenuation handling range of the computer algorithm.

17
New cards

Beam Hardening

The shift of a polychromatic X-ray beam toward higher average energy as lower-energy photons are preferentially attenuated when passing through matter.

18
New cards

Bowtie Filter

A specialized beam filter that pre-hardens the beam edges to match the thinner peripheral paths through a patient's body.

19
New cards

Compton Scatter

X-ray photon scattering proportional to 1E\frac{1}{E} that causes photons to change direction and impart excess unscattered signal to detectors, producing dark bands and cupping artifacts.

20
New cards

Partial Volume Artifact

Shading or inconsistent pixel values occurring when an anatomical detail is only partially in the beam path or when materials of widely different absorption occupy the same voxel.

21
New cards

Photon Starvation

An artifact occurring when an X-ray beam is severely attenuated (such as laterally through the shoulders), causing detector noise to be magnified into horizontal streaks.

22
New cards

Ray Aliasing

Under-sampling within a projection that occurs when the effective detector size is large compared to the anatomical detail.

23
New cards

View Aliasing

Angular undersampling in frequency space resulting from an excessively large projection interval during fast gantry rotation.

24
New cards

Zebra Artifact

Faint stripe artifacts appearing on multiplanar and 3D reformatted images due to z-axis noise inhomogeneity produced by helical interpolation.

25
New cards

Stairstep Artifact

Distortions along structural edges in sagittal and coronal reformatted images caused by wide collimation and non-overlapping reconstruction intervals.

26
New cards

Acceptance Testing

Testing carried out jointly by the vendor and medical physicist upon installation to verify that equipment fulfills purchase contract specifications.

27
New cards

Commissioning Tests

Baseline performance testing conducted by a medical physicist before putting newly installed radiological equipment into clinical service.

28
New cards

Constancy Tests

Routine quality control tests conducted at regular intervals (e.g., monthly or annually) to confirm ongoing satisfactory performance of equipment.

29
New cards

Catphan 600

A standard modular performance phantom used for objective CT quality assurance evaluations including spatial resolution, CT number linearity, and slice thickness.

30
New cards

Hounsfield Unit (HU)

The standard metric for CT numbers defined as HU=μmaterialμwaterμwater×1000HU = \frac{\mu_{material} - \mu_{water}}{\mu_{water}} \times 1000, where μ\mu is the linear attenuation coefficient.

31
New cards

CT Number Uniformity

The maximum difference between the mean CT number of a central ROI and peripheral ROIs across an image of a homogenous phantom.

32
New cards

Z-Sensitivity

The effective imaged slice thickness along the z-axis, measured as the full width at half maximum (FWHM) of CT numbers across an inclined phantom test ramp.

33
New cards

Irradiated Slice Thickness

The full width at half maximum (FWHM) of the radiation dose profile along the z-axis from a single scan, which is always equal to or greater than the imaged slice thickness.

34
New cards

Mean Absorbed Dose (DTD_T)

The mean energy imparted by ionizing radiation per unit mass in a specific organ or tissue, measured in Gray (1Gy=1J/kg1\,Gy = 1\,J/kg).

35
New cards

Equivalent Dose (HTH_T)

The absorbed dose to a tissue multiplied by the radiation weighting factor wRw_R (HT=wRDT,RH_T = \sum w_R D_{T,R}), measured in Sievert (SvSv).

36
New cards

Effective Dose (EE)

The sum of tissue-weighted equivalent doses across all body organs (E=wTHTE = \sum w_T H_T), measured in Sievert (SvSv), representing total stochastic biological risk.

37
New cards

Computed Tomography Dose Index (CTDI)

The integral of the single-scan radiation dose profile along the z-axis normalized to nominal slice thickness TT: CTDI=1TD(z)dzCTDI = \frac{1}{T} \int_{-\infty}^{\infty} D(z)\,dz.

38
New cards

CTDI100, air

A CTDI measurement performed free-in-air at the gantry center of rotation using a 100-mm pencil ionization chamber.

39
New cards

Weighted CTDI (CTDIwCTDI_w)

An estimate of average single-slice cross-sectional dose in standard 16-cm or 32-cm PMMA phantoms: CTDIw=13CTDI100,c+23CTDI100,pCTDI_w = \frac{1}{3} CTDI_{100,c} + \frac{2}{3} CTDI_{100,p}.

40
New cards

Volume CTDI (CTDIvolCTDI_{vol})

A CT radiation dose index representing average dose across a scanned volume in helical CT, calculated as CTDIvol=CTDIwpitchCTDI_{vol} = \frac{CTDI_w}{pitch}.

41
New cards

Multiple Scan Average Dose (MSAD)

The average radiation dose along the z-direction over the central scan in a series of contiguous axial CT exposures.

42
New cards

Planar Average Equilibrium Dose (DeqD_{eq})

The dose metric proposed in AAPM Report 111 to replace CTDIvolCTDI_{vol} using direct measurement in wide phantoms with small ion chambers.

43
New cards

Overbeaming

The extension of the physical X-ray beam along the z-axis beyond the active detector array width in multidetector CT to deliver uniform radiation intensity across all active detector rows.

44
New cards

Overscanning

The additional gantry rotation required at the start and end of a helical scan volume to provide sufficient data for tomographic reconstruction.

45
New cards

Z-Axis Geometric Efficiency

The ratio of the dose profile area integrated over the imaged slice range to the total z-axis dose profile area, expressed as a percentage (GE=AAtot×100%GE = \frac{A}{A_{tot}} \times 100\%).

46
New cards

Dose-Length Product (DLP)

A measure of total radiation exposure for a CT examination, defined as DLP=CTDIvol×scan lengthDLP = CTDI_{vol} \times \text{scan length} and expressed in units of mGycmmGy \cdot cm.

47
New cards

Size-Specific Dose Estimate (SSDE)

A patient dose metric introduced in AAPM Reports 204 and 220 that tailors CTDIvolCTDI_{vol} using patient anatomical width or water equivalent diameter conversion factors.

48
New cards

Diagnostic Reference Level (DRL)

An investigation level tool used in medical imaging to support dose optimization without functioning as a legal upper safety limit or optimum dose standard.

49
New cards

DRL Value

A benchmark standard established at the 75th percentile of distribution medians for a measured DRL quantity collected across broad regional or national facility surveys.