1/65
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
grid ratio
h/w
higher grid ratio
more lead, more xray absorbed, need more mAs
grid cutoff
loss of remant radiation absorbed by lead strips and loss of density on film/receptor exposure caused by grid errors or CR not centered
off level grid
grid supposed to be flat but is slanted or tube angled (parallel or focused grids). cut off across entire image and looks underexposed
off center/lateral decentering grid
grid shifted laterally and CR not centered, loss of density across entire image and loss of receptor exposure
off focus/distance decentering
focused grids need specific SID, cut off severe at edges of image
upside down focused grid
center of image good. complete loss of exposure on lateral aspects of image
uniformity
all lead strips uniform height/width
manufacturing defect or dropping on edges
uniformity test
image a homogenous phantom, process and take density readings at different points
every point should be within ±0.1
CR and DR measure pixel brightness and should be within 20%
density readings #
every point within ±0.1 or 10%
CR/DR pixel brightness
within 20%
alignment
proper centering and distance
why would we xray a grid
to check for barium on it
misaligned grid
misaligned grid lead can cause an increase in attenuation of primary xray beam
loss of image quality and increased pt dose
scatter caused by
high kVp levels, more matter, more FOV, more filtration
reduce scatter production
less kVp, less matter, collimation
reduce scatter reaching image receptor
grids, air gap
grid conversion formula
MAS1/MAS2=GCF1/GCF2
grid conversion factors
None=1, 5:1=2, 6:1=3, 8:1=4, 10/12:1=5, 16:1=6
flouro is
the single most acute dose to pt and personnel
flouro spatial resolution
1-3 lp/mm
flouro tubes have _____ compared to general xray tubes
heat capacity
flouro QC tests should be done
every 6 months
QC tests for flouro
same as diagnostic: focal spot, filtration, grid, generator tests
image intensifier
digital flat panel detector
a vacuum tube with flow of electrons
function of II
brighten image
II increase in brightness due to
increase in xray photons in the xray tube hitting the intensifier
increase in kvp and ma
input phosphor II
cesium iodide - xray to light
aluminum II
placed in front of input to reflect light
photocathode II
cesium and antimony - light to electrons
output phosphor II
zinc cadmium sulfide - electrons to light
what will cause loss of detail/signal in flouro
large FOV, less photons/mAs in beginning, vignetting/curved input phosphor, far pt to II tube
what decreases brightness in II
magnification mode, increase in pt thickness/tissue density
minification gain
input diameter²/output diameter²
image from larger input onto a smaller output causes the image to be brighter
flux gain
high voltage accelerates electrons increasing their kinetic energy from photocathode to output phosphor
causes an increase in the amount of light photons releeasd at output phosphor
brightness gain formula
minification gain x flux gain
brightness gain
amount of brightness increased due to intensifier vs non intensified image
brightness gain in II tube #s
5k-30k candela/m² and decreases 10% a year
brightness in II is affected by
kVp and mA
automatic brightness control (ABC)
automatic exposure rate control (AERC)
magnification
smaller the input phosphor size the more magnification
more detail to see smaller objects
magnification disadvantages
less input phosphor being used= decrease in minification gain causing image to be lighter
decreased FOV
increase mA will increase skin dose
magnification factor
Input phosphor 1/ input phosphor 2
magnification dose increase
input phosphor 1²/ input phosphor 2²
conversion factor
the amount of light produced by the output phosphor per unit of xradiation incident on the input phosphor
output intensity measured in candelas- unit of luminous intensity
CF
50-300 candela/m²
decreases 10% a year, pt dose increases from this
veil glare/flare
light reflects from the window of the output phosphor reducing image contrast
occurs most often when moving from one body part to another
moving from chest to abdomen can cause a sudden increase in brightness
pincushion distortion
caused by projecting an image from a curved surface (input phosphor) onto a flat surface (output phosphor)
distortion and decrease in brightness is greater toward the lateral edges
vignetting
a decrease in image brightness at lateral portions of the image
flouro visual checks happen when
every 6 months
tower and table locks
power assist- tower should be able to move freely over tabletop
bucky slot cover should move to cover opening
bucky slot cover measure radiation
on both sides of cover. amt outside should be 10% of input exposure
deadman switch
must have this
radiation only operates when on switch
tested on acceptance
frequent inspection for sticking or malfunction
timer
must have a 5-min reset timer
timer accuracy done with stopwatch
park
should have a lock for park position
when locked in position should not be able to energize flouro mode
collimation
light field to radiation field should not exceed 3% of SID
table
should move freely and stop at appropriate spot
table table indicator and actual angle should be within 2 degrees
table table indicator and actual angle should be within
2 degrees
AERC and how it works
7.5 mm thick phantom.
dosimeter between phantom and xray source
expose for 10 seconds and record reading
add another 7.5 mm and expose again.
dosimeter reading should be double
radiation exposure limits tabletop
should not exceed 10R/,min or 88 mGy/min
in fixed flouro equipment source to tabletop
15 in (38cm)
in portable flouro equipment source to tabletop
12 in (30cm)
air kerma
can be used instead of exposure or intensity to measure amt of radiation reaching patient
measured in rads or Gy
measured in air at where the center of the useful beam enters the patient
air kerma max should not exceed 44 mGy/min or 5 rad/min
best spot for II and xray tube for pt
II closer, xray tube further
DAP meter
dose area product measures dose to pt in mGy/cm²
exposure should be
constant in a single room each time tested.
when comparing different rooms should be within 25%
tested annually
ways to reduce pt exposure
collimation, last frame/image hold - save, shortest pt to II distance, highest possible kVp, pulsed flouroscopy, using largest input phosphor, longest tube to pt distance
image lag
continuation or persistence of an image, blurring the object, when the II is moved over the pt