254: Unit 4 Fluoro / QC

0.0(0)
studied byStudied by 0 people
0.0(0)
full-widthCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/94

encourage image

There's no tags or description

Looks like no tags are added yet.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

95 Terms

1
New cards

fluoroscopy primary function

  • provides real-time dynamic viewing of anatomic structures

    • motion of circulation

    • motion of internal structures

2
New cards

conventional fluoro

  • less than 5 mA

  • high patient dose

  • long exposure time

  • continuous exposure

3
New cards

ABC

automatic brightness control

4
New cards

other names for ABC

  • ADC

  • ABS

  • AERC

5
New cards

ADC

automatic dose control 

6
New cards

ABS

automatic brightness stablization

7
New cards

image intensifier purpose

reduce patient dose by decreasing the amount of x ray photons needed for imaging purposes

8
New cards

image intensifier

  • intercept the X-ray photons and convert them into visible light

    • amplify or intensify this light signal

  • reduce patient dose by decreasing x ray photons needed

  • contains cathode & anode

9
New cards

output phosphor

  • a fluorescent screen

  • receives the electrons from the photocathode

  • absorbs electrons and emits light photons

  • made a thin layer of silver activated zinc cadmium sulfate phosphor

10
New cards

anode

  • positively charged end

  • serves to attract the electrons emitted from the photocathode

  • in front of the output phosphor

    • a hole in the center allows electrons to pass and interact with the output phosphor 

11
New cards

how do the electrons pass through to interact with the output phosophor? 

a hole in the center of the anode 

12
New cards

electrostatic lenses

  • a series of negatively charged electrodes

  • focuses and accelerates the electrons emitted from the photocathode toward the anode

  • uses electrostatic repulsion

    • pushes the electron stream toward the middle 

13
New cards

input phosphor

  • first element of the image intensifier

  • scintillator screen

  • absorbs X-ray photons and emits light

  • constructed of cesium iodide crystals

14
New cards

flux gain

  • energy gained during acceleration from the photocathode to the output phosphor

  • light photons / x ray photons

15
New cards

flux gain equation

number of output light photons / number of input x ray photons

16
New cards

minifcation gain

  • the concentration of photoelectrons from the input phosphor to the output phosphor

  • (input screen)2 / (output screen)2

17
New cards

brightness gain

  • the product of the flux gain and the minifcation gain

  • product of X-ray photons coming in to how many light photons come out 

18
New cards

brightness gain equation 

minification gain (flux gain) 

19
New cards

multi-field image intensification purpose

magnify the image

20
New cards

multi-field image intensification

  • increasing the voltage of the electrostatic lenses & uses a smaller input phosphor 

  • brings focal spot closer to input phosphor and away from output phosphor 

    • magnifies the image sent to the output phosphor

21
New cards

magnification factor

ratio of diameters available to diameters used 

22
New cards

patient dose ____ when using magnification factor

increases

23
New cards

why does patient dose increase when using magnification factor? 

An image intensifier uses a smaller section of the input screen & systems ABC will increase the exposure to increase the exposure to keep the image bright 

24
New cards

advantages of magnification factor

  • better spatial resolution

  • better contrast resolution

25
New cards

disadvantage of magnification factor 

higher patient dose 

26
New cards

CCD advantages

  • high spatial resolution

  • high SNR

  • high DQE

  • no warm-up required

  • no lag or blooming

  • no spatial distortion

  • no maintenance

  • unlimited life

  • unaffected by magnetic fields

  • linear response

  • lower patient radiation dose

27
New cards

FPIR

flat panel image receptor

28
New cards

Flat panel image receptor takes the place of 

the image intensifier 

29
New cards

Flat panel image receptor is made of

  • cesium iodide (scintillator)

  • amorphous silicon (photodetector)

30
New cards

Flat panel image receptor magnification

  • done electronically 

  • No increase in patient dose

31
New cards

Flat panel image receptor advantages

  • distortion-free images

  • constant image quality over the entire image

  • improved contrast resolution over the entire image

  • high DQE at all radiation dose levels

  • rectangular image area coupled to a similar image monitor

32
New cards

multiple detector system 

  • uses 3 TFT detectors 

  • detectors switch between static and dynamic imaging

  • The tube is below the patient

33
New cards

AERC

automatic exposure rate control

34
New cards

automatic brightness control

automatically adjust the kvp and mAs based on patient attenuation to maintain consistent brightness of contrast of the image 

35
New cards

single detector system 

  • single TFT detector used for static and dynamic imaging 

  • X-ray tube above the patient 

  • allows for longer SIDs for a greater variety of exams 

  • cost-effective

  • can be used remotely by a radiologist 

36
New cards

digital fluoroscopy

  • higher mA

  • lower patient dose

  • shorter exposure time

  • pulsed exposure

37
New cards

intensification process

x rays → light → photon electrons → light

38
New cards

pulse progressive fluoroscopy 

  • uses a high-power generator 

  • Pulses the x ray production from the fluoro x ray tube synchonrously with the detector signal 

39
New cards

interrogation time

time required for the generator to come on a produce the necessary kVp and mAse

40
New cards

extinction time

time required to shut the generator down in preperation for the next pulse

41
New cards

duty cycle

the fraction of time the x ray tube is energized

42
New cards

LIH

last image hold 

43
New cards

DSA

digital subtraction angiography

44
New cards

post processing

  • window/level

  • invert

  • filtering edge enhancement

45
New cards

mobile fluoro (CR) minimum source to skin distance 

12” (30 cm) 

46
New cards

stationary fluoro minimum source to skin distance 

15” (38 cm) 

47
New cards

audible timer

5 mins

48
New cards

primary source of exposure for tech/ rad

the patient

49
New cards

lead apron & gloves

0.5 Pb/eq

50
New cards

bucky slot cover / protective curtain

0.25 mm / Pb eq

51
New cards

general purpose table top expsosure should not exceed

100 mGya/min

52
New cards

high level (IR) tabletop exposure should not exceed 

200 mGya/min

53
New cards

Quality Assurance

  • deals with people

    • radiology reports

    • repeat rates

    • image QA

54
New cards

quality control

  • deals with instrumentation & equipments

  • equipments performance evaluations

55
New cards

QC step 1

acceptance testing

56
New cards

QC step 2

periodic monitoring

57
New cards

QC step 3

maintenance  and repair 

58
New cards

acceptance testing

  • testing the new equipment once you receive it

  • not done by the people who created the equipment

  • done by a medical physicist

59
New cards

periodic monitoring

routine evaluation of the x ray producing or image processing equipment

60
New cards

maintenance and repair 

a dedicated analysis of each image to identify deficiencies and artifacts 

61
New cards

filtration

  • tested annually

  • greater than or equal to 2.5 mm Al

62
New cards

collimation

  • semiannually

  • plus / minus 2% SID

63
New cards

Focal spot size

  • annually

  • plus / minus 50%

64
New cards

Calibration of kVp

  • annually

  • plus/minus 10%

65
New cards

exposure timer accuracy 

  • annually 

  • plus/minus 5% with exposures greater than 10 ms 

66
New cards

exposure linearity

  • annually

  • plus/ minus 10% less than or equal to 10 ms

67
New cards

exposure reproducibility

  • annually

  • plus / minus 5%

68
New cards

primary test for effective FSS

slit camera 

69
New cards

Multiple Detectors in Room

  • fluoroscopic

  • upright imaging

  • recumbent imaging

70
New cards

viewing systems

  • LCD monitors (used more)

  • Plasma monitors

71
New cards

AEC testing

  • test the exposure readout using varying thicknesses of material (Al)

  • no matter the thickness the receptor readout should be the same N

,

72
New cards

SMPTE pattern

  • used for observing gross deviations in luminance 

  • determines spatial resolution & contrast resolution 

73
New cards

DICOM

digital imaging and communication

74
New cards

GSDF

  • gray scale display function

  • allows medical images to be transferred at a DICOM standard to be displayed on any DICOM device with a consistent grayscale

  • ensures images are displayed the same way on different monitors

75
New cards

TG18

  • used to check for contrast and spatial resolution display quality on the radiologist's monitor

  • evaluates for

    • geometric distortion

    • reflection

    • luminance

    • display resolution

    • display noise

76
New cards

ACR

recommends that all digital display devices utilize TG 18 patterns in formats that can be displayed on the digital monitor for evaluation  purpose

77
New cards

medical physicist

  • test as appropriate

    • annually

    • post repair

    • acceptnace testing

  • periodic review of the facility QC program

78
New cards

QC technologist

  • routinely,l as established by the department QC program

  • ability to acquire the TG 18 QC test pattern and use it on each digital display device regularly

79
New cards

staff technologist

  • daily

  • visually monitor workstations for indications of malfunction

80
New cards

geometric distortion

  • lines in pattern should appear straight

  • pincushion distortion

  • barrel like distortion

81
New cards

display noise 

any high-frequency fluctuations or patterns that interfere with the detection of the true signal on

82
New cards

conventional

  • continuous exposure

  • Vidicon TV Camera Tube

  • low mA

  • higher patient dose

83
New cards

CCD took the place of _____

Vidicon TV Camera Tube

84
New cards

digital 

  • post processing 

  • ADC

  • FPIR 

  • CCD 

  • pulsed exposure

  • high mA

  • low patient dose 

85
New cards
<p>A</p>

A

output phosphor

86
New cards
<p>B</p>

B

anode

87
New cards
<p>C</p>

C

electrostatic lenses

88
New cards
<p>D</p>

D

electrons

89
New cards
<p>E</p>

E

photocathode

90
New cards
<p>F</p>

F

input phosphor

91
New cards
<p>G</p>

G

glass envelope

92
New cards
<p>H</p>

H

focal point

93
New cards
94
New cards
95
New cards

at what stage of image intensified fluoroscopy is the number of photons the lowest

when entering the input phosphor