Simulation Quality Assurance/ Quality Control (1)

studied byStudied by 0 people
0.0(0)
learn
LearnA personalized and smart learning plan
exam
Practice TestTake a test on your terms and definitions
spaced repetition
Spaced RepetitionScientifically backed study method
heart puzzle
Matching GameHow quick can you match all your cards?
flashcards
FlashcardsStudy terms and definitions

1 / 22

encourage image

There's no tags or description

Looks like no one added any tags here yet for you.

23 Terms

1

What is the goal of CT simulation QA

to ensure safe and accurate operation of the CT simulation process

New cards
2

Geometrical and Spatial integrity (accurate positioning and alignment)

QA check

New cards
3

CT number accuracy (HU callibration)

QA check

New cards
4

target and critical structure localization accuracy

QA check

New cards
5

example of radiation and patient safety QC check

safety interlocks and emergency off switches and CT dosimetry

New cards
6

example of electromechanical QC check

X-ray Generator Parameters and Couch & Laser Alignment

New cards
7

example of Image quality assurance check

Resolution and contrast assessment

New cards
8

how often is the CT simulation tube warmed up

daily QA (prior to QC test)

New cards
9

how often do we check the gantry laser alignment

daily QA

New cards
10

CT tabletop orientation is checked

monthly

New cards
11

table motion accuracy is checked

montly

New cards
12

laser alignment is checked

monthly

New cards
13

this is done to verify that the CT scanner tabletop is level and orthogonal to the imaging plane

CT tabletop orientation

New cards
14

what is the tolerance for CT tabletop orientation

±2mm

New cards
15

this is done to check the vertical and longitudinal movement verification

table motion accuracy

New cards
16

what is the tolerance for table motion accuracty

±1mm

New cards
17

what is done to check the orientation of gantry, wall, and ceiling lasers with the imaging plane and verify lateral wall lasers align with gantry lasers and scan plane

laser alignment

New cards
18

what is the tolerance for all laser alignment checks

±2mm

New cards
19

what is the tolerance for table indexing and position accuracy

±1mm

New cards
20

this is done to ensure tilt indicators return to the nominal position

gantry tilt verification

New cards
21

what is the tolerance for gantry tilt verification

±1 degree or ±1mm from nominal vertical position

New cards
22

the scan volume and scan location must be within _____ tolerance for scan localization accuracy

1mm

New cards
23

the gantry laser alignment check should align with what

the center of the imaging plane

New cards
robot