Chem 241: Sensors

0.0(0)
studied byStudied by 0 people
GameKnowt Play
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
Card Sorting

1/30

encourage image

There's no tags or description

Looks like no tags are added yet.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

31 Terms

1
New cards

Do sensors always involve measuring the concentration of the analyte directly?

No- many sensors involve measuring color change, fluorescence, change in conductivity, change in mass, or a shift in electrochemical potential

2
New cards

biosensor

uses biological components (enzymes, antigens, or enzyme receptors) for a highly selective response to one analyte

3
New cards

Point-of-Care Devices

a medical device used to obtain diagnostic results at or near the "point of care," main goal is to be fast and accurate, a point of care may be at a hospital, dr's office, EMT work, or home, example is blood glucose testing, testing for infectious diseases, etc

4
New cards

ASSURED criteria

criteria for a POC device:

affordable, sensitive, specific, user-friendly, rapid, robust, equipment free, deliverable

5
New cards

ELISA

enzyme-linked immunosorbent assay, sensitivity is due to the enzyme in which 1 molecule catalyzes the same reaction multiple times, producing many molecules of the colored/fluorescent product that can be detected visually/spectroscopically

<p>enzyme-linked immunosorbent assay, sensitivity is due to the enzyme in which 1 molecule catalyzes the same reaction multiple times, producing many molecules of the colored/fluorescent product that can be detected visually/spectroscopically</p>
6
New cards

Lateral Flow Home Pregnancy Immunoassay

detects hormone hcG that is produced shortly after pregnancy in urine

gold standard for designing sensors (POC that meets ASSURED criteria)

False positives occur from cross-reactivity between molecules similar in shape to the antigen and the immobilized antibodies

<p>detects hormone hcG that is produced shortly after pregnancy in urine</p><p>gold standard for designing sensors (POC that meets ASSURED criteria)</p><p>False positives occur from cross-reactivity between molecules similar in shape to the antigen and the immobilized antibodies</p>
7
New cards

blood glucose monitor

amperometric, relates the current (I) to the concentration of glucose indirectly; has working and reference electrode, reference electrode is to provide a constant potential; all reagents on a test strip

interferences: ascorbic acid, acetaminophen

<p>amperometric, relates the current (I) to the concentration of glucose indirectly; has working and reference electrode, reference electrode is to provide a constant potential; all reagents on a test strip</p><p>interferences: ascorbic acid, acetaminophen</p>
8
New cards

amperometric measurement

measures current and relates to concentration at a constant potential

9
New cards

generations of blood glucose monitors

1st gen) hydrogen peroxide product of the reaction diffuses to the transducer and causes the electrical response, O2 is mediator that transports the electrons between the working electrode and glucose molecule (I = [H2O2] = [Glucose)

2nd gen) involves specific mediator to generate improved response, glucose oxidase as a redox mediator

3rd gen) reaction itself causes response using organic conducting materials based on charge-transfer complexes, superior selectivity

10
New cards

potentiometric measurements

measure cell potential (E in mV) or potential across an ion-selective electrode membrane with little to no current flow, ISEs fall into this category

use Nernst equation to relate cell potential to concentration

11
New cards

Ion Selective Electrodes (ISE)

not based on a REDOX reaction (no electron transfer), based on a potential difference across a membrane;

all ISEs have a permselective membrane, an inner and outer reference electrode, and a fixed concentration of the ion inside the electrode

the concentration of the analyte ion in the outer solution changes, causing the membrane potential to change

<p>not based on a REDOX reaction (no electron transfer), based on a potential difference across a membrane;</p><p>all ISEs have a permselective membrane, an inner and outer reference electrode, and a fixed concentration of the ion inside the electrode</p><p>the concentration of the analyte ion in the outer solution changes, causing the membrane potential to change</p>
12
New cards

Nernst Equation for ISEs

*Alter the sign and charge for the equation based on the charge of the ion used

<p>*Alter the sign and charge for the equation based on the charge of the ion used</p>
13
New cards

junction potentials

smaller junction potentials give a greater mobility, this is why Cl- and K+ are commonly used in ISEs

if the mobility of the ion is greater, it has the greater potential to attract the opposite charge. For example, if K+ > Na+, this means that K+ will have a more negative side of the liquid junction because it can attract more negative ions

14
New cards

combination electrode

consists of a glass electrode with a concentric reference electrode built on the same body, pH electrodes are an example

<p>consists of a glass electrode with a concentric reference electrode built on the same body, pH electrodes are an example</p>
15
New cards

pH electrode

inner solution has a fixed concentration of H+ but the analyte (outside) solution varies in H+ concentration, the binding to silica glass sites (Si-OH) determines the difference in potential, which is read as the pH.

more H+ binding = lower pH

for every factor of 10 change in H+ at 25 C, a difference of 59.16 mV is observed

<p>inner solution has a fixed concentration of H+ but the analyte (outside) solution varies in H+ concentration, the binding to silica glass sites (Si-OH) determines the difference in potential, which is read as the pH.</p><p>more H+ binding = lower pH</p><p>for every factor of 10 change in H+ at 25 C, a difference of 59.16 mV is observed</p>
16
New cards

interfering species in ISEs

adds extra term in Nernst equation, and example is alkaline/sodium error

Spectral: overlap of analyte signal

Choose another wavelength for analysis, ICP, Monochromator

17
New cards

steps for using an ISE

1. the electrolyte is placed into an aqueous solution

2. analyte ions in the exchange sample solution equilibrate with ion-exchange sites in the membrane

3. Analyte ions diffuse out of the membrane

4. A charge imbalance, also called a potential difference, develops across the interface

5. the potential difference is measured using the internal and external reference electrodes

6. the concentration of the analyte ion in the external solution is determined from the potential difference

18
New cards

sodium error

in an alkaline solution with an increased concentration of Na+, the glass electrode indicates a pH that is lower than the actual pH

this is because Na+ binds to the same exchange sites as H+, and the electrodes respond to Na+ as if H+ was present

usually for pH of 11 or greater

19
New cards

potential errors when using a pH electrode

the glass mmb is not hydrated properly, not enough equilibration time, calibrated at a different temperature than the solution, acid error, sodium error

remember that no electrode is 100% selective for one ion

20
New cards

acid error

acidic conditions, glass is saturated with H+ and cannot be protonated, giving a more basic pH than the solution truly has

21
New cards

steps for using a pH electrode

1. electrode placed in aqueous solution

2. the surface of the thin glass membrane swells as it absorbs water

3. metal ions in the hydrated gel region diffuse out of the glass into solution

4. H+ diffuses into the glass in place of the metal ions

5. The potential difference is measured across the membrane by two reference electrodes

6. The potential difference is converted to a pH value

22
New cards

ion-exchange equilibrium

established at the glass membrane when in contact with a definite H+ concentration

23
New cards

Solid State Electrodes (the F- ISE)

Doping membrane crystals with EuF2 (Eu2+) creates anion vacancies and increases conduction; used to create a potential difference against a solid-state membrane

24
New cards

ionophores

aka ligand, organic ion-exchange agent that increase the permeability of membranes to ions, provides selectivity for the analyte ion, utilized by Ca2+ and K+ in liquid membrane electrodes

25
New cards

characteristics of the organic polymer mmb in an liquid membrane electrode

hydrophobic, filled with a viscous organic solution containing an ion exchanger, sometimes a ligand specific for the analyte ion

26
New cards

Ammonia ISE

1. use a gas-permeable membrane

2. increase the pH of the external solution to convert NH4+ > NH3

3. NH3 can pass across the gas permeable membrane

4. NH4+ to NH3 ratio is changed and the pH is increased

27
New cards

Spectral Interference

Overlap of analyte signal with signals due to other elements

Solution: choose another wavelength for analysis

ICP, Monochromator

28
New cards

Physical Interference

Matrix related issues affecting droplet size or transport, fix by matching matrix in standards and samples

29
New cards

Chemical Interference

Formation of stable compounds that will not atomize

Fix: use releasing (competes with interfering species for analyte, freeing analyte into the atomization zone, or protective agents (form volatile or soluble complexes with the analyte itself)

30
New cards

Ionization Interference

Loss of neutral atoms due to high temp ionization

Na ionizing in flame, add ionization suppressors which will decrease the extent of ionization in an analyte

31
New cards

Selectivity Coefficient

Relative Response to different species of the same charge, plug into Nerst Equation to Correct

k Pot A,X = reponse to X (interferring)/ Reponse to A (ion you want to see a response to).

You want K to be small

<p>Relative Response to different species of the same charge, plug into Nerst Equation to Correct</p><p>k Pot A,X = reponse to X (interferring)/ Reponse to A (ion you want to see a response to). </p><p>You want K to be small</p><p></p>