1/126
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
a series of RF and gradient pulses applied over several time periods
pulse sequence
2 ways rephasing can be accomplished
using a 180 degree refocusing pulse
or
using the magnetic gradients
_________ pulses are also used to spatially encode the information contained within the MR signals, and to rephase and dephase transverse magnetization during certain pulse sequences
gradient
rephasing is performed by the 180 degree rephasing pulse
spin echo pulse seqeunces
(conventional spin echo, fast spin echo)
rephasing is performed by the gradients
gradient echo pulse sequences
(coherent/incoherent gradient echo)
what type of pulse sequence is the gold standard in MR imaging and produces the standard in T1 and T2 contrast
Conventional spin echo pulse sequence
True or False
The slice select gradient is turned on whenever the RF energy is being transmitted
True
(This ensures that the only protons that are excited are the ones in the current slice that we are imaging, and not the rest of the anatomy around it)
The _________ gradient comes on with the echo, as it reads and encodes varying frequencies within the echo. It stays on throughout rephasing and dephasing.
readout
The ________ of the readout gradient determines the range of frequencies encoded in the echo, known as the receiver bandwidth
amplitude
Does a steeper readout gradient encode more or less frequencies than a shallow readout gradient?
steeper gradient readout = more frequencies
shallow gradient readout = less frequencies
The timing of the ________ encoding gradient is not as specific as the slice select and readout gradient
phase
When is the phase gradient usually turned on?
after the 90 degree excitation pulse and before the 180 degree rephasing pulse
phase dispersion is only allowed by the ______ encoding gradient since _______ is a unique identifier used to locate the position of the signal
phase
true or false
a shift in phase increases incremental values as time continues
true
A positive change in frequency followed by an _______ negative change in frequency will result in no shifting of the phase
equal
the protons on the positive side of the gradient are going to spin ________ than protons on the negative side
faster
If we want to cancel the phase shift created by the application of the slice select and readout gradients, we need what?
an equal but opposite pulse from those same gradients
If a gradeint of equal amplitude, but opposite polarity is turned on, the phase will _______, leaving a net phase of 0, a rephased signal.
This is how we rephase the signal in a ___________ pulse sequence
cancel
gradient echo
With the spin echo technique, utilizing a 180 degree refocusing pulse halfway between the initial 90 degree RF and TE, the effects of what are eliminated?
T2*
the time it takes for spins to rephase after the initiation of the 180 degree pulse
TE is twice tau
tau is 1/2 TE
After the application of the 90 degree excitation pulse, what is there?
FID (T2* dephasing)
Is FID ever rephased?
NO
Are spin echo and gradient echo ever rephased?
yes, ALWAYS
How are gradient and spin echo rephased?
either with a 180 RF pulse
or
180 degree RF rephasing pulse
what is used to compensate for FID dephasing?
180 degree rephasing pulse
in a spin echo pulse sequence, what governs the decay of the signal after the 180 degree pulse?
spin spin only (True T2)
What type of dephasing occurs before tissues have time to relax by T2 and T1?
T2*
a spin echo technique follows the 90 degree pulse with a 180 degree pulse.. for what?
to bring nuclei back into phase
1. 90 degree excitation pulse
2. T2* immediately causes dephasing
3. a voltage would be induced in the receiver coil but the FID signal is not collected.
4. a 180 degree rephasing pulse brings nuclei back into phase
5. The rephased signal induces a voltage in the receive coil at time TE. This is our MR signal
timing of the spin echo pulse sequence
What is the purpose of the 180 degree rephasing pulse?
to bring the transverse magnetization back to full phase coherence
what supresses T2*?
spin echo signal
The application of the 180 degree refocusing pulse causes the spins to what?
reverse polarity
A gradient of _______ amplitude and the _______ polarity is applied on either side of the 180 degree refocusing pulse, leaving a net phase of _____
same
same
0
a positive change occurs with the application of the gradient
a 180 degree rephasing pulse is applied
another equally positive gradient is applied
the end result is no phase shift
how we balance gradients for phase in spin echo pulse sequences
During a conventional spin echo pulse sequence, if a 256 phase matrix is selected, then _____ uniquely phase encoded echoes will be produced to fill the _____ lines of kspace.
Assuming one excitation is selected, than 256 TR's are needed to complete one image. What is unique about them? (answered on next slide)
256
256
The phase gradient steps its amplitude ______ each TR, giving a unique phase to the protons in every line of k-space
down
In conventional spin echo sequences, only ______ phase encoding step is applied per TR for each slice.
The resulting spin echo then fills only _____ line of k space per TR
1
1
What is the difference between spin echo pulse sequences and fast spin echo?
Both sequences still only produce one echo per TR, per image, and k space is still filled in the same way
what's different is that we are getting an additional echo, just one, at a different weighting (different contrast)
A dual contrast conventional SE sequence contains an extra _______ pulse and generates a ________ spin echo at a ______ TE
180
2nd
2nd
What type of pulse sequence does this process describe:
2 different k-space files are opened, and one line is filled with each TR. One echo is placed in the topmost line of one k-space file, the second echo is placed in the topmost line of the second kspace file. This same phase encoding amplitude is used for both echoes
dual contrast conventional Spin echo sequence
True or False
Dual contrast techniques can also be performed with a fast spin echo sequence
true
dual contrast conventional spin echo pulse sequences generally use a _______ TR.
By using a ______ TR, these sequences are good for generating what 2 types of contrast?
long
long
PD (Long TR, short TE)
T2 contrast (Long TR, Long TE)
In a dual contrast conventional spin echo pulse sequence, the first echo will be used to generate ______ images, and the second echo will be used for _______ images
1st echo = PD images
2nd echo = T2 images
The time from the 90 degree excitation pulse to the center of the first echo
TE for first echo
The time from the 90 degree excitation pulse to the center of the second echo
TE for the second echo
In order to generate the second echo, what must be applied half way between the center of the first echo?
the second 180 degree rephasing pulse
how is fast spin echo the same conventional spin echo?
they both use the initial 90 degree RF excitation pulse, followed by a 180 degree pulse to rephase the spins, we then get a spin echo at time TE
What is the difference between conventional spin echo and fast spin echo?
with FSE, our scan times are much shorter than conventional spin echo, this is because the echoes are collected more frequently and placed in k-space differently
Many echoes are collected per TR, instead of just one echo
what is this?
fast spin echo
Each echo is placed appropriately in k-space so that the desired image contrast is achieved
what is this?
fast spin echo
FSE sequences complete the data acquisition very ________ compared to conventional spin echo, while generating images with contrast __________ to a conventional spin echo sequence
quickly
similar
in conventional spin echo, how are echoes placed in k-space?
sequentially
only one echo for one slice
in fast spin echo, how are echoes collected?
multiple echoes are collected during a single repetition, for each slice
more echoes are collected per TR
In fast spin echo, ________ echoes are placed in k space in a single TR, which means ______ repetitions are required to fill the man lines of raw data
multiple
fewer
What type of sequences use a 180 degree rephasing pulse?
conventional spin echo
fast spin echo
The number of echoes collected
echo train length
when more than one 180 degree refocusing pulses is applied after a 90 degree excitation pulse, each successive echo is of _______ amplitude
less
the 180 degree pulse can rephase dephasing caused by inhomogeneities and other dephasing mechanisms, but can it rephase spin spin dephasing?
NO!!!!! each tissue is unique and we can't control this process
For conventional spin echo and fast spin echo, what determines how many spin echoes will be needed to fill kspace for each image in the sequence?
our phase matrix
what is effective TE?
the TE that will give us the contrast we desire
in both conventional spin echo and fast spin echo, when does the 180 degree pulse have to come on?
halfway between each subsequent echo
MR echoes are temporarily stored in the K-space until they are constructed into an image in the _______ ________ using the ________ _______
array processor
Fourier transform
each echo represents ______ phase encoded line of data and contains information about the ________ slice
one
entire
Information located at the ________ of k space contains high spatial frequencies, since this is where the amplitude of the phase encoding gradient is the _________
periphery (end)
greatest
high spatial frequencies correspond to what?
spatial resolution (aka edge defintion or line pairs per mm)
information located toward the center of k-space contains ______ spatial frequencies, since this is where the amplitude of the phase encoding gradient is at the _______
low
lowest
low spatial frequencies correspond to what?
contrast information
true or false
each and every echo contributes to the entire MR image, regardless of the type of sequence that is run
true
what happens if the center of k-space is removed?
we have sharp resolution, but zero contrast defintion
what happens if the periphery of k-space is removed?
we have contrast definition but no resolution
in conventional spin echo sequences, the first echo collected is place into the top most line of k space, the second echo collected is placed in the line immediately below that and so on...
This type of "mapping" means what?
early and late echoes contribute to the images resolution, while the central echoes (the echoes collected half way through the sequence) contribute to the image contrast
In fast spin echo, the echoes are mapped out differently.
The system places the echoes that occur near/at the ____________ in the central lines of k-space, so the resulting image displays the desired contrast.
The echoes not close to the __________ will be placed into the outer lines of k-space, both positive and negative
effective TE
in fast spin echo, if the scan parameters specify a TE of 90 ms, the echo which formed at ________ will be placed in the very center of k-space to provide the appropriate contrast information
90ms
True or False
In fast spin echo, each phase encoding step applies a different slope of the phase encoding gradient to phase shift the signal by a different amount each time
true
for fast spin echo:
very steep phase encoding slopes = _____ signal amplitude
very shallow phase encoding slopes = _______ signal amplitude
low
high
In fast spin echo, the system re-orders the echoes by using a ______ phase encoding gradient to produce echoes at or near the effective TE and placing them in the center of k-space
the other echoes utilize a ______ phase encoding slope and are placed away from the effective TE
shallow
Steep
on a fast spin echo T2 weighted image, fat and water are both ________
bright
why do fat and water both appear bright on FSE T2 images?
j coupling
what is J-coupling
precessional frequencies can vary within the same molecule because the spinning protons that are side by side with each other (coupled) can lose or gain magnetization from each other. This "coupling" causes increased signal loss (called j-coupling) on a spin echo sequence
true or false
rapid rephasing of fast spin echoes suppresses the j coupling dephasing (signal loss), resulting in brighter signals for fat and water on T2 weighted images
true
what does j-coupling mean for fast spin echo?
the rapid 180 degree rephasing pulse disrupts the signal loss originally caused by j-coupling. Since we do not have signal loss in fat on a FSE sequence, we get bright (strong) signal from fat, similar to the signal we get from water
what if we don't want to see fat as bright on T2FSE sequence?
use a fat suppression technique (STIR)
What type of sequence is an inversion recovery sequence?
FSE sequence that begins with an 180 degree inversion pulse
what does the inversion period do?
creates a large contrast difference between fat and water over time & this allows for differences in the t1 recovery times , producing heavy T1 weighting
an inversion recovery sequence ALWAYS begins with what?
a 180 degree rephasing pulse
the time from the first 180 degree RF inverting pulse to the 90 degree RF excitation pulse
inversion time (TI)
Do we use gadolinium for STIR (short TI inversion recovery)?
NO
(b/c contrast agents shorten the T1 time of perfused tissues and STIR sequences reduce the signal from short T1 tissues. If contrast was given, it would be suppressed and we wouldn't see it.
what is very useful in musculoskeletal imaging?
STIR
Because STIR surpasses fatty bone marrow, what can be seen more clearly?
bony lesions
fat has a very bright signal on MR images and can make it difficult to visualize pathology in fatty areas such as
orbits
abdomen
joints
TI & TR for FLAIR
TI = very long around 2000 ms and a TR of 6-10 seconds
Applications of FLAIR
visualizing MS plaques in the brain
metastatic disease of the brain and spine
In FSE, the SAR _________ significantly because of multiple 180 degree pulses, applied in quick succession
increases
How do we decrease the SAR?
Decreasing the rephasing angle to 120-160 degrees
which are faster:
spin echo or gradient echo?
gradient echo
differences between GRE & SE
GRE utilizes short TR's (regardless of what type of contrast we want to see)
no 180 RF pulse (gradients are used to rephase it)
echo is called gradient echo
what type of flip angles are used in GRE?
Flip angles of less than 90 degrees are used, so full recovery of longitudinal magnetization occurs faster. This allows for faster repetition of the sequence
what controls contrast in a conventional GRE pulse sequence?
flip angle
what is used to rephase the signal in a GRE pulse sequence
equal but opposite gradients
why are GRE pulse sequences faster?
hydrogen protons respond to magnetic fields faster than they do with RF energy and resonance , so they rephase faster than an RF pulse