Gradients, Slice Selection, Frequency Encoding (9.1 - 9.3)

0.0(0)
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
Card Sorting

1/28

encourage image

There's no tags or description

Looks like no tags are added yet.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

29 Terms

1
New cards

NOTE

Read the note explaining Gradients

2
New cards

What is the role of gradient coils in MRI?

Gradient coils in MRI are used to vary the strength of the main magnetic field (B0B_0) along different axes (x, y, z). This variation allows the system to spatially encode the location of protons by adjusting their resonant (Larmor) frequencies.

3
New cards

How are gradient coils mathematically represented?

The gradient field is represented as:

B=(B0+Gxx+Gyy+Gzz)z^=(B0+Gr)z^ B = \left(B_0 + G_x x + G_y y + G_z z \right) \hat{z} = (B_0 + \mathbf{G} \cdot \mathbf{r}) \hat{z}

Here:

- B0B_0: Main magnetic field strength.

- Gx,Gy,GzG_x, G_y, G_z: Gradient strengths along the x, y, and z directions.

- r=(x,y,z)\mathbf{r} = (x, y, z): Position vector.

- G=(Gx,Gy,Gz)\mathbf{G} = (G_x, G_y, G_z): Gradient vector.

4
New cards

What is the significance of gradient coil switching?

Switching gradient coils on and off rapidly (0.1–1 ms) enables precise control of magnetic field variation and spatial encoding of signals. This rapid switching, called the slew rate, is typically 5–250 mT m1^{-1} s1^{-1} and must be carefully managed to avoid inducing eddy currents in the patient.

5
New cards

How do gradients contribute to MRI image formation?

Gradients allow selective excitation of protons at specific spatial locations by modifying their resonant frequencies. The spatial information is encoded in the signal phase and frequency, enabling reconstruction of the image through the Larmor frequency and transverse magnetization phase.

6
New cards

What is the purpose of slice selection in MRI?

Slice selection ensures that the received signal only comes from a specific slice of tissue. This is achieved by applying a gradient G=(Gx,Gy,Gz) G = (G_x, G_y, G_z) , which creates a Larmor frequency variation depending on the spatial location.

7
New cards

What happens to the Larmor frequency with a gradient?

The Larmor frequency becomes position-dependent:

ω(r)=γ(B0+Gr) \omega(r) = \gamma(B_0 + G \cdot r)

If we want to select a specific slice at z z , we apply a gradient G=(0,0,Gz) G = (0, 0, G_z) , making the Larmor frequency:

ω(z)=γ(B0+Gzz) \omega(z) = \gamma(B_0 + G_z z)

8
New cards

Why can’t we excite an infinitesimally thin slice in practice?

- RF pulses have finite bandwidth, so a precise sinusoidal waveform cannot be delivered.

- Signals depend on the number of protons in the slice, and a very thin slice would not provide enough signal for detection.

9
New cards

How is slice selection achieved in practice?

A range of RF frequencies is applied to excite a thick slice or slab over the frequency range [ν1,ν2][ \nu_1, \nu_2 ]. The slice thickness depends on the combination of:

- RF frequency range (bandwidth Δω=ω2ω1\Delta \omega = \omega_2 - \omega_1)

- Gradient strength (Gz G_z )

- RF center frequency (ωˉ=(ω1+ω2)/2 \bar{\omega} = (\omega_1 + \omega_2)/2 )

10
New cards

What are the three parameters for controlling slice position and thickness?

1. Gz G_z : z-gradient strength.

2. ωˉ \bar{\omega} : RF center frequency.

3. Δω \Delta \omega : RF frequency range (bandwidth).

11
New cards

What is the relationship between slice thickness and slice position, and how can you calculate RF bandwidth for slice selection?

Fundamental Equations:

1. Larmor Frequency: ω=γ(B0+Gzz)\omega = \gamma (B_0 + G_z z)

2. Slice Position: z1=ω1γB0γGzz_1 = \frac{\omega_1 - \gamma B_0}{\gamma G_z}, z2=ω2γB0γGzz_2 = \frac{\omega_2 - \gamma B_0}{\gamma G_z}

3. Mean Slice Position: zˉ=ωˉω0γGz\bar{z} = \frac{\bar{\omega} - \omega_0}{\gamma G_z}

4. Slice Thickness: Δz=ΔωγGz\Delta z = \frac{\Delta \omega}{\gamma G_z}

5. RF Bandwidth: Δν=γGzΔz\Delta \nu = \gamma G_z \Delta z

Steps for Derivation of Slice Thickness and Position:

1. Relationship for Slice Position (Mean, zˉ \bar{z} ):

- Slice position is determined using the center frequency ωˉ \bar{\omega} . The gradient strength Gz G_z and gyromagnetic ratio γ \gamma relate it to the z z -position:

z1=ω1γB0γGz,z2=ω2γB0γGz z_1 = \frac{\omega_1 - \gamma B_0}{\gamma G_z}, \quad z_2 = \frac{\omega_2 - \gamma B_0}{\gamma G_z}

- Mean position zˉ \bar{z} is calculated as the average of z1 z_1 and z2 z_2 :

zˉ=z1+z22=1γGz(ω1+ω22γB0) \bar{z} = \frac{z_1 + z_2}{2} = \frac{1}{\gamma G_z}\left(\frac{\omega_1 + \omega_2}{2} - \gamma B_0 \right)

or:

zˉ=ωˉω0γGz \bar{z} = \frac{\bar{\omega} - \omega_0}{\gamma G_z}

2. Relationship for Slice Thickness (Δz \Delta z ):

- Slice thickness is the difference between z2 z_2 and z1 z_1 :

Δz=z2z1=ω2ω1γGz \Delta z = z_2 - z_1 = \frac{\omega_2 - \omega_1}{\gamma G_z}

or:

Δz=ΔωγGz \Delta z = \frac{\Delta \omega}{\gamma G_z}

3. Calculate RF Bandwidth (Δν \Delta \nu ):

- Using the equation for Δz \Delta z , solve for RF bandwidth Δν \Delta \nu :

Δν=γGzΔz \Delta \nu = \gamma G_z \Delta z

- Substitute values:

γ=42.58×106Hz/T,Gz=40mT/m,Δz=5mm=5×103m \gamma = 42.58 \times 10^6 \, \text{Hz/T}, \, G_z = 40 \, \text{mT/m}, \, \Delta z = 5 \, \text{mm} = 5 \times 10^{-3} \, \text{m}

Δν=42.58×106×40×103×5×103=8.5kHz \Delta \nu = 42.58 \times 10^6 \times 40 \times 10^{-3} \times 5 \times 10^{-3} = 8.5 \, \text{kHz}

<p><strong>Fundamental Equations:</strong></p><p>1. <strong>Larmor Frequency</strong>: $$\omega = \gamma (B_0 + G_z z)$$  </p><p>2. <strong>Slice Position</strong>: $$z_1 = \frac{\omega_1 - \gamma B_0}{\gamma G_z}$$, $$z_2 = \frac{\omega_2 - \gamma B_0}{\gamma G_z}$$  </p><p>3. <strong>Mean Slice Position</strong>: $$\bar{z} = \frac{\bar{\omega} - \omega_0}{\gamma G_z}$$  </p><p>4. <strong>Slice Thickness</strong>: $$\Delta z = \frac{\Delta \omega}{\gamma G_z}$$  </p><p>5. <strong>RF Bandwidth</strong>: $$\Delta \nu = \gamma G_z \Delta z$$  </p><p></p><p>Steps for Derivation of Slice Thickness and Position:</p><p>1. <strong>Relationship for Slice Position (Mean, $$ \bar{z} $$):</strong>  </p><p>   - Slice position is determined using the center frequency $$ \bar{\omega} $$. The gradient strength $$ G_z $$ and gyromagnetic ratio $$ \gamma $$ relate it to the $$ z $$-position:</p><p>$$ z_1 = \frac{\omega_1 - \gamma B_0}{\gamma G_z}, \quad z_2 = \frac{\omega_2 - \gamma B_0}{\gamma G_z} $$</p><p>   - Mean position $$ \bar{z} $$ is calculated as the average of $$ z_1 $$ and $$ z_2 $$:</p><p>$$ \bar{z} = \frac{z_1 + z_2}{2} = \frac{1}{\gamma G_z}\left(\frac{\omega_1 + \omega_2}{2} - \gamma B_0 \right) $$</p><p>     or:</p><p>$$ \bar{z} = \frac{\bar{\omega} - \omega_0}{\gamma G_z} $$</p><p>2. <strong>Relationship for Slice Thickness ($$ \Delta z $$):</strong>  </p><p>   - Slice thickness is the difference between $$ z_2 $$ and $$ z_1 $$:</p><p>$$ \Delta z = z_2 - z_1 = \frac{\omega_2 - \omega_1}{\gamma G_z} $$</p><p>     or:</p><p>$$ \Delta z = \frac{\Delta \omega}{\gamma G_z} $$</p><p>3. <strong>Calculate RF Bandwidth ($$ \Delta \nu $$):</strong>  </p><p>   - Using the equation for $$ \Delta z $$, solve for RF bandwidth $$ \Delta \nu $$:</p><p>$$ \Delta \nu = \gamma G_z \Delta z $$</p><p>   - Substitute values:  </p><p>$$ \gamma = 42.58 \times 10^6 \, \text{Hz/T}, \, G_z = 40 \, \text{mT/m}, \, \Delta z = 5 \, \text{mm} = 5 \times 10^{-3} \, \text{m} $$</p><p>$$ \Delta \nu = 42.58 \times 10^6 \times 40 \times 10^{-3} \times 5 \times 10^{-3} = 8.5 \, \text{kHz} $$</p>
12
New cards

What is the trade-off in slice selection for imaging a volume?

A thicker slice results in higher signal-to-noise ratio (SNR) but poorer resolution in the slice direction. Conversely, thinner slices give better resolution but may reduce SNR and contrast-to-noise ratio (CNR).

13
New cards

What are the consequences of slice selection depending on RF pulses?

RF pulses with a specific range of frequencies are required to define slices. Perfectly defined slice edges are not achievable because RF pulses cannot perfectly deliver only the desired frequencies, leading to extra components due to the pulse shape and duration.

14
New cards

Why are small gaps left between slices in multi-slice MRI?

Small gaps help reduce overlap or interference between adjacent slices, improving the overall quality of the MRI. Optimizing RF pulse shape is also critical for high-quality images.

15
New cards
term image

The Fourier transform of a rectangular function is a sinc function. To get a perfect slice selection we need a pulse that only has frequencies in a specific range, i.e. the bandwidth, and equal magnitude at all frequencies within the bandwidth. Thus, the pulse needs to generate a rectangular function in the frequency domain, this would require a sinc pulse in the time-domain. However, a sinc function has many side lobes and thus cannot be generated in a finite duration pulse. In reality careful shaping of the RF pulse will be required to get as precise a slice selection as possible.

16
New cards

What is the equation for transverse magnetization after a pulse?

Mxy(t)=Mxy(0+)ej(2πω0tϕ)et/T2 M_{xy}(t) = M_{xy}(0^+) e^{-j(2\pi \omega_0 t - \phi)} e^{-t/T_2} (Eq. 9.4)

Magnetisation in the horizontal plane when protons are tilted.

17
New cards

What is Mxy(0+) M_{xy}(0^+) ?

Mxy(0+)=Mz(0)sinα M_{xy}(0^+) = M_z(0^-) \sin \alpha , where Mz(0) M_z(0^-) is the longitudinal magnetization before the pulse and α \alpha is the flip angle. (Eq. 9.5)

18
New cards

What is the received signal in MRI?

s(t)=AMxy(x,y,0+)ej2πω0tet/T2(x,y)dxdy s(t) = \iint A M_{xy}(x, y, 0^+) e^{-j2\pi \omega_0 t} e^{-t/T_2(x, y)} dx dy , where:

- A A : Scanner-specific gain,

- Mxy(x,y,0+) M_{xy}(x, y, 0^+) : Transverse magnetization immediately after excitation,

- ω0 \omega_0 : Larmor frequency,

- T2(x,y) T_2(x, y) : Spin-spin relaxation time. (Eq. 9.6)

19
New cards

What assumption is made for simplicity in Eq. 9.6?

The phase ϕ \phi is assumed to be zero, and the spatial variation of Mxy M_{xy} and T2 T_2 is explicitly noted.

20
New cards

What is the equation for the received signal in terms of the effective spin density, and how is it derived?

The received signal integrates over the transverse magnetization within the slice and is given as:

s(t)=ej2πω0tf(x,y)dxdy s(t) = e^{-j2\pi \omega_0 t} \iint_{-\infty}^{\infty} f(x, y) \, dx \, dy

Here, f(x,y) f(x, y) , the effective spin density, combines all the relevant parameters of the tissue and the scanner:

f(x,y)=AMxy(x,y,0+)et/T2(x,y) f(x, y) = A M_{xy}(x, y, 0^+) e^{-t/T_2(x, y)}

- A A : Scanner-specific gain, related to the receive electronics.

- Mxy(x,y,0+) M_{xy}(x, y, 0^+) : Transverse magnetization immediately after excitation.

- et/T2(x,y) e^{-t/T_2(x, y)} : Decay of the transverse magnetization due to T2 T_2 relaxation over time.

This signal reflects the sum of contributions from all protons within the selected slice.

21
New cards

What is demodulation, and why is it used in MRI?

Demodulation is the process of removing the high-frequency carrier signal (oscillating at the Larmor frequency) from the received MRI signal to extract the meaningful low-frequency components.

- In MRI, the received signal is modulated at the Larmor frequency ω0 \omega_0 .

- Demodulation shifts the signal to the "baseband," where it represents the actual spatial information.

22
New cards

What is the baseband signal?

The baseband signal is the low-frequency component of the MRI signal after demodulation, containing all the spatial information about the tissue being imaged.

- It simplifies the representation of the received signal for further processing.

- In the equation s0(t)=ej2πω0ts(t) s_0(t) = e^{j2\pi \omega_0 t}s(t) , demodulation removes ej2πω0t e^{-j2\pi \omega_0 t} , leaving only the spatially dependent signal.

23
New cards

What is the demodulated signal (baseband) we receive, and why is it necessary?

s(t)=ej2πω0tf(x,y)dxdy s(t) = e^{-j2\pi \omega_0 t} \iint_{-\infty}^{\infty} f(x, y) \, dx \, dy

The demodulated signal removes the carrier frequency ω0 \omega_0 to simplify further processing. It is written as:

s0(t)=e+j2πω0ts(t)=f(x,y)dxdy s_0(t) = e^{+j2\pi \omega_0 t} s(t) = \iint_{-\infty}^{\infty} f(x, y) \, dx \, dy

- This step removes the oscillations due to ej2πω0t e^{-j2\pi \omega_0 t} , leaving only the integral over the effective spin density.

- However, this step also eliminates all spatial information, as the dependency on x x and y y is lost.

- To recover spatial details, we need spatial encoding through gradients.

24
New cards

Why do we need a readout gradient, and what does it do?

A readout gradient is applied to encode spatial information along a specific axis (e.g., x-axis). This gradient causes the Larmor frequency to vary with the position along that axis:

ω(x)=γ(B0+Gxx)=ω0+γGxx \omega(x) = \gamma (B_0 + G_x x) = \omega_0 + \gamma G_x x

- Gx G_x : Strength of the gradient in the x-direction.

- ω0 \omega_0 : The base Larmor frequency without gradients.

The varying frequency ensures that spins at different positions along the x-axis precess at different rates, allowing their spatial location to be encoded into the signal.

25
New cards

How does the received signal change with a readout gradient?

The signal becomes spatially dependent on x x , while still integrating over y y :

s(t)=AMxy(x,y,0+)ej2π(ω0+γGxx)tet/T2(x,y)dxdy s(t) = \iint_{-\infty}^{\infty} A M_{xy}(x, y, 0^+) e^{-j2\pi (\omega_0 + \gamma G_x x)t} e^{-t/T_2(x, y)} \, dx \, dy

Breaking this down:

- AMxy(x,y,0+) A M_{xy}(x, y, 0^+) : The initial transverse magnetization of the spins within the tissue.

- ej2πω0t e^{-j2\pi \omega_0 t} : The base oscillation due to the main field B0 B_0 .

- ej2πγGxxt e^{-j2\pi \gamma G_x x t} : The additional phase due to the gradient Gx G_x , which encodes the spatial location along the x-axis.

- et/T2(x,y) e^{-t/T_2(x, y)} : The decay of the transverse magnetization due to T2 T_2 relaxation.

26
New cards

What is the final simplified form of the signal with spatial encoding?

Using the spatially dependent Larmor frequency ω(x)=ω0+γGxx \omega(x) = \omega_0 + \gamma G_x x , the signal simplifies to:

s(t)=ej2πω0tAMxy(x,y,0+)ej2πγGxxtet/T2(x,y)dxdy s(t) = e^{-j2\pi \omega_0 t} \iint_{-\infty}^{\infty} A M_{xy}(x, y, 0^+) e^{-j2\pi \gamma G_x x t} e^{-t/T_2(x, y)} \, dx \, dy

This equation incorporates spatial encoding along x x , enabling the reconstruction of spatially dependent information in the final image.

27
New cards

What is the baseband signal obtained using the readout gradient?

Old without gradient: s0(t)=e+j2πω0ts(t)=f(x,y)dxdy s_0(t) = e^{+j2\pi \omega_0 t} s(t) = \iint_{-\infty}^{\infty} f(x, y) \, dx \, dy

The baseband signal after demodulation is given by:

s0(t)=f(x,y)ej2πγGxtxdxdy s_0(t) = \iint_{-\infty}^{\infty} f(x, y) e^{-j2\pi \gamma G_x t x} \, dx \, dy

This represents the integral over the effective spin density f(x,y) f(x, y) with spatial encoding along the x-axis due to the readout gradient Gx G_x .

F(u,v)=f(x,y)e2πi(ux+vy),dx,dy F(u, v) = \int_{-\infty}^{\infty} \int_{-\infty}^{\infty} f(x, y) e^{-2\pi i (ux + vy)} , dx , dy

28
New cards

How does the baseband signal relate to a 2D Fourier Transform?

F(u,v)=f(x,y)e2πi(ux+vy),dx,dy F(u, v) = \int_{-\infty}^{\infty} \int_{-\infty}^{\infty} f(x, y) e^{-2\pi i (ux + vy)} , dx , dy

s0(t)=f(x,y)e2πi(γGxtx)dxdy s_0(t) = \iint_{-\infty}^{\infty} f(x, y) e^{-2\pi i(\gamma G_x t x)} \, dx \, dy

The baseband signal is equivalent to a 2D Fourier transform of the spin density f(x,y) f(x, y) , where:

- Spatial frequency in the x-direction: u=γGxt u = \gamma G_x t

- Spatial frequency in the y-direction: v=0 v = 0

29
New cards

What is k-space, and how is it represented in MRI?

In MRI, the Fourier space where spatial frequencies are sampled is called k-space.

- kx=u=γGxt k_x = u = \gamma G_x t : Wave number in the x-direction.

- ky=v=0 k_y = v = 0 : Wave number in the y-direction.

The concept of k-space simplifies the representation of spatial frequency data collected during the scan.


2D Fourier plot starts from origin (0,0)

Horizontal Axis (u-axis): Represents the horizontal spatial frequency.  

  • Higher values on this axis correspond to faster variations in the horizontal direction of the original image or signal.

Vertical Axis (v-axis): Represents the vertical spatial frequency.  

  • Higher values on this axis correspond to faster variations in the vertical direction of the original image or signal.

  • This means there is no vertical variation in frequency for the k-plane.