What patient size criteria indicate use of the Lucas device?
Cardiac arrest in a patient with chest width less than 17.7 inches and sternum height between 6.7 and 11.9 inches.
What is a contraindication for Lucas device use related to patient size being too small?
Patient is too small for the plunger to provide adequate depth of compressions.
What is a contraindication for Lucas device use related to patient size being too large?
Patient is too large for the top part of the Lucas to click into the back plate.
What should happen to manual CPR while placing the Lucas Back Plate?
Temporarily stop manual CPR while placing the Back Plate under the patient, immediately below the armpits.
What two methods can be used to place the Lucas Back Plate under the patient?
Lift the patient's torso and slide the Back Plate under from the head, or log roll the patient and slide the Back Plate in from the side.
What should be done after placing the Back Plate, regarding CPR and the device?
Continue manual CPR and turn on the Lucas.
How is the LUCAS Upper Part removed from the bag?
By holding the handles on the Support Legs.
What should be checked before letting go of the Release Rings?
Pull the Release Rings once to make sure the Claw Locks are open.
How is it confirmed that the support legs are secured to the back plate?
By listening for a click and pulling up on it.
Where should the Suction Cup be positioned relative to the sternum?
Immediately above the end of the sternum.
What is adjusted to set the Start Position on the Lucas device?
The height of the Suction Cup.
What does the Start Position represent on the Lucas device?
The position where the device will start its two-inch compressions and return the chest for full recoil.
How is the Suction Cup pulled down to touch the patient's chest?
With two fingers, until the Pressure Pad touches the patient's chest without compressing it.
What is pressed to lock the Lucas device's start position?
Pause.
What should be checked before pressing Start on the Lucas device?
Proper position, adjusting if needed.
What should not be delayed when applying the Lucas Stabilization Strap?
Device application or medical care.
What is the purpose of the Lucas Stabilization Strap?
It is a device Stabilization Strap, not a neck Stabilization Strap.
Can the Lucas Stabilization Strap be applied over a c-collar?
Yes, it can fit and be put on the outside of a c-collar.
What is secured in the patient straps of the Lucas device?
The patient's wrists.
What provider levels can operate the Lucas device per CFD protocol?
EMT-B, EMT-A, and EMT-P.