1/22
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
|---|
No study sessions yet.
Automatic Brightness Control (ABC)
circuit alters the kVp, mA or both according to changes in x-ray beam attenuation by the anatomic part due to patient position, thickness atomic number, or density
the radiologist may select the brightness level desired, and the ABC will maintain it throughout the procedure
ABC
Automatic Brightness Control
AERC
Automatic Exposure Rate Control
ABC & AERC =
2 different names for the process of Automatic Stabilization of Brightness
example - as the patient rolls onto side, the increased tissue thickness must be compensated for to maintain fluoroscopic image brightness
Two general approaches are used….
Automatic Gain Control (AGC)
Automatic Brightness Stabilization (ABS)
Automatic Gain Control (AGC)
after the image is acquired, amplification of the electronic signal (current) is increased or decreased as needed by the circuits.
if an insufficient number of x-ray photons are reaching the image intensifier, no amount of increased electronic amplification can overcome the mottle or noise that is produced
Automatic Brightness Stabilization (ABS)
more common: actual radiographic technique factors are adjusted to compensate for fluctuations in the electronic signal, the incident x-ray beam, or the measured light output.
ABS Circuits
Variable mA, Preset kVp:
the kVp is set by the operator and does not vary. The circuit automatically varies the mA up and down as needed to maintain image brightness
ABS Circuits
Variable mA with kVp following :
The circuit automatically varies the mA up and down as needed to maintain image brightness, but as an addition circuit measures if an upper boundary set for the mA is reached - When this limit is reached, the system switches to a motor-driven variable transformer that varies the kVp up and down
ABS Circuits
Variable kVp with selected mA:
mA is set by the operator and does not vary. A motor - driven variable transformer varies the kVp up and down as needed to maintain image brightness. (high mA can be set to force lower kVp’s for iodine-based studies, low mA for GI work)
ABS Circuits
Variable kVp, Variable mA:
Both kVp and mA are varied up and down as needed to maintain constant image contrast or suppress noise. There is no operator control.
Fluoroscopic Beam-On Time
The operator must be made aware of accumulated beam - on time by a manual reset, 5-minute timer.
Intermittent Fluoroscopy
A series of short visual checks, rather than continuous fluoroscopy, dramatically reduces cumulative exposure to the patient
this is the most important way to limit patient exposure during fluoroscopic studies
Controlled by the radiologist
Last Image Hold Devices
Keeping the last image displayed on the monitor, has been shown to reduce cumulative patient exposure as much as 50% - 80%
Pulsed Fluoroscopy
Pulsing of the x-ray beam spares the patient radiation exposure between frames. Shorter pulse times can also be selected, further reducing dose
Used correctly - it doesnt harm image quality
Lower frame rates than realtime, can be used in conjunction with digital memory to provide a continuous-looking image
should NEVER be above 20 fps because the human eye cant detect greater than
Frames per second
Pulsed Fluoroscopy Technique
The mA must be increased from about 2 mA to 20 mA or more
Conventional Fluoroscopy typically 0.5 - 5mA were used to reduce pateint dose and keep tube heat in control
Digital systems are higher mA, so dose in controlled with a pulsed, intermittent exposure sequence
But because it isn’t continuous it is actually less dose to the patient
Tube heat capacities are reduced
Parts of the Pulse Waveform
Pulse Height
Pulse width
Pulse interval
Pulse Height
height created by the mA setting (intensity)
Increased as more photon flux is needed
Pulse width
how long the x-ray is on for
aka the “duty cycle”
Does not affect flux gain
Shorter widths = 10ms or less
less blurry images, less dose, less tube heat
Pulse Interval
How long the x-ray is off for
Short “ones” allow fast moving anatomy to be better viewed
If long “ones” are used, use stationary imaging
Improve the fluoro blurred image by utilizing …
Fast moving anatomy may need these adjustments
Short pulse widths
Short pulse intervals
Fast frame rates
Pulse Frequency or Puse Rate
The lower the pulse number the less the patient dose
An increased pulse rate will improve temporal resolution (improve blurring or “smearing” of the image)
But increase patient dose
Typical rates …
3.75
7.5
15
30 - no dose reduction
continuous - should not be used