semiconductor
A material that can conduct an electric current or not, depending on the voltage applied across that material
AEC
An electronic device designed to terminate x-ray exposure after properly exposing an image receptor
line compensation
A method of adjusting the voltage to the x-ray imaging system to a constant value, in response to changes in voltage supplied by the power company
capacitor
An electrical circuit device that stores electric charge
mA meter location
Placed in the tube circuit, connected at the center of the secondary winding of the high-voltage step-up transformer in series with the x-ray tube. This reduces the possibility of shock.
diode
An electrical device that contains two electrodes
voltage ripple
The fluctuation in the voltage applied to the x-ray tube expressed as a percentage of peak kilovoltage
rectification
The process of changing alternating current and voltage into direct current and voltage
autotransformer
A type of transformer having a single winding
power
The product of voltage and current. Watt.
302.5 V
220 V is supplied across 1200 windings of the primary coil of the autotransformer. If 1650 windings are tapped, what voltage will be supplied to the primary coil of the high-voltage transformer?
71.7 volt peak
A kVp meter reads 86 kVp and the turns ratio of the high-voltage step-up transformer is 1200. What is the true voltage across the meter?
5 V
The supply voltage from the autotransformer to the filament transformer is 60 V. If the turns ratio of the filament transformer is 1/12, what is the filament voltage?
6 amps
If the current in the primary of the filament transformer was 0.5 A with a turns ratio of 1/12, what would be the filament current?
104.5 kV
The suply to a high-voltage step-up transformer with a turns ratio of 550 is 190 V. What is the voltage across the x-ray tube?
radiolucent
Allows relatively undiminished intensity of x-rays through the tabletop
90/20 table
Allow relatively undiminished intensity of x-rays through the tabletop
10. On/off control, kVp selection, mA selection, time (mAs) selection, and automatic-exposure controls
List the five major controls on the operator's console.
purpose of the autotransformer
To vary and control the amplitude of the voltage supplied to the high-voltage step-up transformer and the filament transformer
primary/secondary voltage relationship in autotransformer
Primary voltage and secondary voltage are in direct relation to the number of turns of the transformer
What the prereading kVp meter allows
The precise adjustment of the supply voltage and monitoring the kVp before the x-ray exposure
3.3
200 mA, 1/60s= what mAs?
18
600 mA, 30 ms = what mAs?
Difference between high voltage transformer and high voltage generator
The high-voltage transformer is just one component of a high-voltage generator.
direct
In order that a reverse voltage is not applied across the x-ray tube and that the tube operates most efficiently, x-ray tubes use ___ current.
rectification
Direct current is achieved in the x-ray circuit through ______.
100%
Voltage ripple of single-phase generator
14%
Voltage ripple of three-phase, six-pulse generator
4%
Voltage ripple of three-phase, twelve-pulse generator
1%
Voltage ripple of high-frequency generator
single-phase
0.7 (mA)(kVp)/1000 is the equation for computing ______ power rating
three-phase, high-frequency
(mA)(kVp) / 1000 is the equation for computing ______ or _______ power ratings.
housing cooling chart
Graph showing the cooling rate of an x-ray tube housing
leakage radiation
Radiation emitted through the x-ray tube housing (other than the primary beam)
heat unit
Measure of heat capacity (1HU = 1AVs = 1Ws =1J)
focusing cup
Shroud inside the x-ray tube surrounding the cathode to concentrate electrons on the focal spot
anode rotation speed
3400 rpm or 10,000 rpm
thoriated tungsten
Tungsten alloyed with thorium
x-ray tube current
Cathode to anode electron flow
grid-controlled x-ray tube
X-ray tube capable of high speed switching. Voltage applied to the focusing cup is the switch.
convection
Method of heat transfer by a moving fluid medium (liquid or gas)
space charge
Electron cloud in the vicinity of the filament
floor, wall, or ceiling mounted
Three methods used to support the x-ray tube
cathode and anode
Where thoriated tungsten would be used in a x-ray imaging system
saturation current
When all available electrons are projected from the cathode to the anode
heavy filaments and high capacity anodes
why are arcing and tube failure no longer a problem in modern x-ray tube design?
thermionic emission
Pass an electric current to heat a conductor and cause outer-shell electrons to be released from the conductor
tube arcing
Principal cause of tube failure
thorium
What addition to the filament material prolongs tube life?
to electrostatically shape beam
Why is the filament embedded in the focusing cup?
better spatial resolution
Why would an x-ray tube need a small focal spot?
high intensity radiation
Why would an x-ray tube need a large focal spot?
cathode
Negative side of the x-ray tube
anode
positive side of the x-ray tube
fixed and rotating
Name two type of anodes
three functions of anode
1)x-ray tube target
2)electrical and thermal conductor
3)mechanical support
high atomic number = efficient x-ray production
How does atomic number affect the selection of anode target material?
thermal conductivity = heat dissipation
How does thermal conductivity affect the selection of anode target material?
melting point = heat capacity
How does melting point affect the selection of anode target material?
induction motor
How does the anode rotate inside a glass enclosure with no mechanical connection to the outside?
anode heel effect
Higher x-ray intensity on cathode side
positioning thicker anatomy on cathode side
How can anode heel effect be used advantageously?
cracked or pitted anode, induction motor failure, and open filament
Name three causes of tube failure
increase kV
How can a space charge be removed?
detent position
X-ray tube locking-in at center and at a given SID