1/173
Fluoroscopy
Name  | Mastery  | Learn  | Test  | Matching  | Spaced  | 
|---|
No study sessions yet.
what is fluoroscopy?
an imaging technique that uses x-rays to create real-time, moving images of the internal structures of the body
what is fluoroscopy also known as?
Dynamic Images
What do dynamic images permit?
the visualization and evaluation of not only the structure, but the function of the anatomy as well
what are some of the exams done under fluoroscopy?
GI studies
urinary studies
arthrograms
orthopedic surgery
cardiovascular cases
when was fluoroscopy first invented?
in 1896
who invented fluoroscopy
thomas edison
what did the first fluoroscopy unit require?
the viewer need to look directly at the glowing screen in line with the x-ray source
Later in time, systems did not require looking directly into the source but….
only one person could view the image at a time
T/F: images could not be saved at first- diagnosis solely based off radiologist’s report
true
early fluoroscopy was _____ intensity fluorescent screen that x-ray anatomy was projected on
low
what was the screen made out of on thomas edison’s fluoroscope?
calcium tungsten (fluoresced (glowed) when exposed to x-rays)
what was the disadvantage of early fluoroscopic systems besides radiation dose?
the lake of brightness of the image
due to the lake of brightness in the images for early fluoroscopic, what had to be done?
procedures needed to be done in the dark, so radiologist needed to adapt their eyes to the dark room
what are the two types of cells in the eye that are responsible for vision?
rods and cones
what are rods?
sensitive to low light
what is the responsibility of rods?
for night vision
what are cones?
detect color and fine detail
what do cones do?
function best in bright light
early fluoro exams were performed under conditions that….
resulted in very poor visual acuity (sharpness)
T/F: conventional R/F rooms can perform both radiographic and fluoroscopic images
true
conventional rooms are convenient for…
static after pictures so the patient doesn’t need to move
conventional rooms include?
ceiling mounted x-ray tube and wall bucky
why is the tube in the table for fluoro rooms?
under the table to make it easier for the operator to reposition the image receptor
Components in the table: TUBE?
generates a continuous or pulsed x-ray beam
Components in the table: Collimator?
radiologist can decrease the field size to improve image quality and decrease patient exposure
Components in the table: Filtration
filters out low energy x-rays
Components in the table: Tabletop
in fixed systems table usually moves by button push
what is the Image intensifier?
makes low mA image brighter
some systems have a flat panel detector
what is a anti-scatter grid?
systems with an image intensifiers have a physical one attached to the image intensifier housing
systems with a flat panel detector might have a built in grid
what is the video camera or charge-coupled device? (on systems with an image intensifier)
it is what takes the intensified image and sends to the analog to digital converter (ADC)
then the adc sends the image to the monitor where it is displayed
in flat panel detector systems, the image is sent to directly to the Analog to digital converter meaning…
no intensification needed
the fluoro tube is….
radiolucent
what two ways can fluoro be initiated?
a hand switch on the carriage or a foot pedal (controlled by radiologist and both are called dead mean style)
where is the fluoro tube attached to?
the back of the table and coupled to the table
what will happen if the fluoro tube is not fixed into place over the table?
the tube will not produce x-rays
the patient should be supine and looking at what on the c-arm
a smiley face
what is high-level fluoroscopy referred to as
“ a boost”
what does high level fluoroscopy do
allow for increased radiation output to improve image quality
Info card: interventional fluoroscopy
in larger hospitals it can be found in its own modality
in smaller hospitals the technologists may be expected to perform some interventional procedures
who should be operating the c-arm
technologist or physicians (doctors) only
T/F: nurses or OR staff can operate the c-arm
false
why are c-arms used
they produce less radiation, and should only be used for larger parts
T/F: bumping into the table is best practice
false
the closer the patient is to the source…..
the higher their dose will be
what does the remote hand switch do?
allows technologist to initiate exposure- dead man style
what does the monitor display system do?
displays real-time live images
has capability to also show saved or reference images
every step away form the patient does what to the occupational dose
decreases it
what are the components of the c-arm imaging unit
the x-ray tube and either an image intensifier or a flat panel detector
what are the components of the c-arm free space
where the body part being imaged needs to go
what are the components of the c-arm
attaches tube to the image intensifier
handles for moving
measurements for positioning
what is the foot pedal?
allows doctor to initiate exposure
what are the components of the c-arm: rotational joint
allows c-arm to rotate or obliques or tolt or cranial/caudal imaging
what are the components of the c-arm: control panel
adjusts imaging settings
collimation
pulse settings
low dose
magnification mode
resets timer
T/F: pulsed per second can be adjusted in system settings
true
the SID for a c-arm is
fixed
T/F: the closer we move the source to the patient to move the image receptor is moved away from the patient with a c-arm
true
fluoroscopy uses an _____ exposure time
increase
what two ways can fluoroscopy exposure time be in
pulsed or continuous
what is focal point
just the location where the e-stream is the most tightly focused
what is the input phosphor do
converts the x-ray photons to light photons
what is the input phosphor made of
cesium iodide
why is the input phosphor made of cesium iodide
structured type of phosphor crystal
better for spatial resolution
1 photon can produce ______ light photons in the input phosphor
1000
what is the photocathode layer do?
converts light photons into electrons
how do we help lower patient dose
we use a lower mA, results in reduced brightness
what is the primary function of the image intensifier?
to brighten the image to improve visual acuity
what is visual acuity
the ability to visualize small details and small changed in density (visualize low contrast)
why was the image intensifier developed
to overcome the dim images
because we use an image intensifier, it allows us to use a lower…
mA, which results in lower patient dose
what are all the image intensifier components encased in ?
a metal housing, which protects the glass or metal envelope that maintains a vacuum around the structures
how does the image get to the viewing monitor?
from the video camera (or charged coupled device) then to a analog to digital converter
the photocathode is ____ bonded to the input phosphor
directly, which help maintain spatial resolution
where is the accelerating anode?
on the opposite end of the image intensifier
what kind of charge does the accelerating anode have
a positive charge; with a hole in the middle
T/F: the accelerating anode in the image intensifier does not spin like the anode in the x-ray tube
true
what does the anode’’ positive charge attract?
the negatively charged electrons; moving towards the anode
what does the hole in the middle of the accelerating anode do?
allows electrons to pass through and strike the output phosphor
what is flux gain
the kinetic energy gained during acceleration from photocathode to output phosphor
what does acceleration mean in “accelerating anode”
it means the speeding up of electrons, so they increase in kinetic energy
what happens when electrons gain kinetic energy?
they can produce more light photons at the output phosphor; which will intensify the signal, making an increase in image brightness
what do the electrostatic focusing lenses do
the negatively charged electrostatic focusing lenses line in the inside of the image intensifier so the electrons can make it through the hole of the anode; like the focusing cup, funnels the electrons into a concentrated stream
what does the output phosphor do
takes the electrons and converts them back into light photons
the output phosphor is the phosphor that..
the electrons strike after they pass through the anode is much smaller than the input phosphor
what is the difference between input and output phosphors
the electron stream is condensed for the output phosphor so they all can strike it
acceleration and condensing of the electrons causes what the image
more/another intensification
takes the electrons and multiplies them by 50-75 times more
info: the light photons from the output phosphor strike a CCD or video camera which then takes the light photons and converts them into a electronic signal and collect that signal
the collected signal is then sent to an analog to digital converter
what does the analog to digital converter do
takes the electronic signal and converts it to a digital signal
why are there so many steps?
each step intensifies the signal, making the final image bright enough for the monitor
T/F: light photons and electrons can be focused, x-rays cannot
true
T/F: intensifying the image allows for lower x-ray doses, protecting patients and staff
true
what does an anti-scatter grid do
reduced scatter reaching the image receptor
what does an anti-scatter grid increase
image quality, image contrast, and patient dose
where is the anti-scatter grid located
within the outer housing and is placed in front of the input phosphor
what is stationary fluoro grid
usually a linear/focused grid
what is interventional fluoro grid
uses crossed grids because of the higher beam in time and mA use
why are higher grid ratios not needed in fluoroscopy
because a low mA is used in fluoro
T/F; fluoro grids are similar to radiography but usually lower
true
T/F: Grids SHOULD be used in pediatric fluoroscopy
false
T/F: the higher image intensifier efficiency, the less radiation to produce a quality image
true
what is the flux gain equation
#of photons output phosphor produced/ #of x-ray photons striking the input phosphor