the study of the interaction between electromagnetic
radiation and matter
Spectroscopy
The type of energy that is transmitted through space at enormous velocities
Electromagnetic radiation
The length of the electrical vector at the maximum in the wave
Amplitude (A)
The linear distance between successive
maxima or minima of a wave
Wavelength (λ)
The number of oscillations of the
field per second
Frequency (n)
Postulate that electromagnetic radiation is made up
of packets of energy called
Photons/Quanta
Functional group that absorbs
light such as C=C, C=O, N=N, and N=O, having
multiple bonding
Chromophore
Functional group that does not absorb in the UV/VIS
region
Auxochromes
Unwanted light of wavelengths that differ from the target wavelength
Stray light
Influenced by instability of light source
Single-Beam Spectrophotometer
Not influenced by instability of light source
Double-Beam Spectrophotometer
Used for measurement in the ultraviolet region (190 to 350 nm)
D2 lamp
Used for measurement in the visible
region (350 to 900 nm)
Halogen lamp
Provide continuous source in the ultraviolet region
Deuterium/ Hydrogen Discharge Lamp
A spectrophotometer generally has an entrance ____ and an exit ____
slit
If the slit width is _______, the intensity of the light irradiating the sample increases but the resolution worsens.
increased
If the slit width is ______, the light intensity decreases but the resolution improves.
decreased
Wide range from 190 to 2500nm; Stable with most acids, bases, and organic solvents
Quartz/fused silica cuvette
Useful in the region of 350 - 2000 nm; Windows of good quality optical glass transmit well throughout the
visible range
Silicate glass cuvette
Disposable cells are molded from polystyrene methacrylate and other transparent plastics
Plastic cuvette
a type of semiconductor element that is used to convert light energy into electrical energy. It uses an electrical phenomenon called the "photovoltaic effect" whereby a reverse current proportional in size to the light intensity is created.
Photodiode
_______ spectroscopy is more suitable for characterization, but not for structural elucidation.
UV/Vis
_____ spectroscopy has taken a more important role in quantitation, rather than structural determination.
UV/Vis
Measured by transmittance method
Transparent liquid
Measured by transmittance or reflectance method (by integrating sphere)
Translucent liquid
Measured by reflectance method (by integrating sphere)
Opaque liquid
Measured by reflectance method (by integrating sphere or specular reflectance accessories)
Solid powder/coatings/mirrored surfaces
Only 1 calibration standard is prepared; Mainly applied to pass/ fail situation; No way to identify dilution error from the graph
Single point calibration
At least 2 calibration standards are needed; More accurate method; Easier to identify dilution error occur in one of the standards
Multipoint calibration
Perform serial dilution to obtain generally __ to __ calibration standards.
3 to 5
Perform spectrum scanning of standards to determine maximum absorption peak wavelength. Concentration of standards should fall between __ to __ absorption unit (depends on unit specs).
0 to 0.4
the derivatives of an absorbance with respect to the wavelength
Derivative spectroscopy
Used for measurement of diffuse and total reflectance, diffuse and total transmittance
Integrating Spheres (ISR)
Method for separating the solute from the stationary or mobile phase based on the difference in interaction
Chromatography
Birth of high performance liquid chromatography
1969
Localization of electrons within a molecule can cause molecules to have negative and/or positive poles.
Polarity
Stationary phase in Normal-Phase/Absorption Chromatography
High polarity
Mobile phase in Normal-Phase/Absorption Chromatography
Low polarity
Stationary phase in Reversed-Phase Chromatography
Low polarity
Mobile phase in Reversed-Phase Chromatography
High polarity
Chromatography method that separates compounds by molecular size (bulkiness)
Size Exclusion Chromatography