List three advantages that point of care testing has over central laboratory testing
More efficient workflows
Reduced turnaround time
Smaller sample volumes
List three disadvantages of point of care testing has compared to central laboratory testing
Less accuracy
Numerous operators to train
Result integration to electronic medical record (EMR) can be challenging
Defined by CLIA as a test requiring expertise beyond normal clinical automation. The test may require reagent preparation, limited pre-evaluation , and interpretation of results
High-Complexity Tests
Defined by CLIA as a test requiring basic lab knowledge and training for personnel performing the test. May require reagent preparation, limited pre-treatment of specimens, quality control, calibration, proficiency testing, some skill in troubleshooting and maintaining equipment, and some skill or judgment in interpretation of results
Moderately-Complexity Tests
The simplest complexity listing in CLIA. It involves primarily test systems approved by the FDA for home use, The requirements are that there is no reasonable risk of harm to the patient if the test is performed incorrectly. The likelihood of erroneous results is negligible; the test method is simple and uncomplicated; and is available for home use; encompasses most POCT tests; operators must follow manufacturers instructions
Waived Tests
Issued to a laboratory to perform only waived tests
Certificate of Waiver
Issued to a laboratory that enables the entity to conduct moderate or high complexity laboratory testing or both until the entity is determined by survey to be in compliance with CLIA regulations
Certificate of Registration
Issued to a laboratory in which a physician, midlevel practitioner, or dentist performs no tests other than the microscopy procedures. Permits the lab to also perform waived tests
Certificate of PPMPs
Issued to a laboratory after an inspection that finds the lab to be in compliance with all applicable CLIA requirements
Certificate of Compliance
Issues to a lab on the basis of the labs accreditation by an accreditation organization approved by the Health Care Finance Administation
Certificate of Accreditation
Eight key components of device design when evaluating analytical and technological considerations are:”
Operator Interface: Helps guide user through the operation
Bar Code Identification System: Identify Patient ID, reagent packs, calibration and other info
Sample and Reagent Delivery Mechanisms: Sensor strip, cassette, etc.
Reaction Cell: Location of Reaction
Sensor: Detects the reaction of the analyte with a binding molecule or some other type of recognition element
Control and Communication Systems: Coordinates all other systems to ensure that all the required processes for an analysis take place in the correct order
Data Management and Storage: Calibrations, QC Data and Patient results
Manufacturing Requirements: Must be reliable
What information should always accompany the test result across the POCT device, the middleware or LIS, and the electronic medical record
Identifiers for the patient, sample, operator, and device
Date and time of both the specimen collection and analysis
Type of specimen
Test requested and test result with appropriate units
Error messages and action messages