1/152
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
|---|
No study sessions yet.
Interpolation of channels
Is used to estimate the stimulation level based on surrounding channels. Can be done separately or in chunks (for kids)
Requires anchor channels
What happens to intermediate values in interpolation? Why do interpolation?
An average of the differences between channels measured is taken
Streamlines the fitting process
How do charge balanced biphasic current pulses work? Who determines the current output?
The amount of charge is the product of amplitude and pulse width
The clinician determines the current output based on loudness perception described by the pt
The greater the charge, the louder the
Percept
What happens when you begin to reach the limits of the device for volume?
If the max amplitude/height is met, the pulse width increases
The input dynamic range determines what? What happens to sound when they come into the IDR? What is the range?
Determines the intensity range of the acoustic input signal that is mapped into the audible electrical output range
Any sound below the IDR will NOT be let in
Any sound above the IDR will be compressed into the pt’s dynamic range
Range: 40-60dB
Microphone sensitivity
Adjust the range of sound that is picked up by the microphone
Increasing the microphone sensitivity results in what?
The pt hearing softer sounds
Pts often decrease sensitivity in loud situations
Pt preference varies
Automatic gain control components
Dual action
Slow-acting control with low compression ratio for everyday speech
Fast-acting control with high compression ratio for loud sounds
Linear
Peak clipping once input level exceeds criterion
Changes in volume result in changes to what?
The comfort levels as a percent of the entire dynamic range. At times the software can limit or set the flexibility of the volume control on the speech processor.
Pts perceive this as a change in loudness
What is frequency allocation? What does it affect?
Controls how frequencies are delivered to the active channels. Can be automatic.
Can affect the upper frequency limit that is allowed to be delivered
Stimulation or repetition rate
The rate at which individual channels are stimulated
Expressed as pulses per second (pps)
Total stimulation rate
The rate per channel times the number of possible channels
Pulse rates > 2000 pps
Auditory nerve firing more likely to be stochastic
Objective measures include
Electrode impedance
Electrically evoked stapedial reflexes
Electrically evoked compound action potential
Electrically evoked auditory brainstem response
Telemetry
Process for obtaining data from the implant device and sending it back to external part
Uses radio frequency link to transmit info
Tells us about the goodness of the RF link and integrity of the device
Can transmit electrophysiological data
Electrode impedance measures determine what?
Impedance measures determine whether the intracochlear and extracochlear electrodes are working appropriately
How is electrode impedance measured?
Small, known amount of current sent to the electrode
Voltage is also known, so impedance can be measured at each electrode contact
Voltage compliance
Is the maximum amount of current that can be delivered to an electrode by a power source
Voltage is determined by?
The battery
Current depends on what?
Impedance
Constant voltage and lower impedance vs constant voltage and higher impedance
Constant voltage and lower impedance means MORE current flow
Constant voltage and higher impedance means LESS current flow
CIs have what type of voltage capacity?
Finite, which determines the current it can deliver
How is electrode impedance measured?
Impedance is measured by the current that returns to the reference electrode and distribution of voltage to the selected electrode contact
Abnormal impedance may mean what?
Open circuit
Short circuit
Partial short circuit
Open circuit
An incomplete path for current to flow/discontinuous circuit. There is infinite resistance or VERY high impedance: clinical software flags values >30 kΩ
Causes:
Broken electrode contact
Broken lead wire
Air bubble
Short circuit
Low resistance between two points in a circuit that differ in potential. Results in increased current flow for a fixed voltage: characterized by very LOW impedance values <1 kΩ
Causes:
Electrode contacts coming into contact with each other (array is kinked or curled back on itself)
Partial short circuit and cause
Characterized by relatively low impedance and increased current flow BUT less than what is seen for a short circuit
Impedances lower than those for nonaffected electrodes
Often now low enough to be flagged by clinical software
Should be monitored
Caused by small tears in the silicone surrounding electrode lead
Identification for partial short circuit
Decrease in impedance of the affected electrodes over time
Decrease in impedance of affected electrode relative to impedance of unaffected electrodes
Increase in MAP levels for effected electrodes
Reduction in speech perception scores
Poor sound quality
Reduced pitch discrimination
Reduced loudness growth
Pain or nonauditory stim
What is the typical time course of impedance changes? At Surgery? Device activated? Few months of use?
Impedance lowest at surgery
2-3 weeks after surgery, impedance increases prior to initial activation
Once device stimulated, impedance typically decreases
Impedance stabilizes within a few months of use
Why can impedance increase?
Periods of non-use
Re-implantation
ME involvement and fluid; infection and inflammation
Operating outside of voltage compliance limits
Clinical uses for impedance measures (6) and possible course of action
Identify electrode failures
Disable affected electrodes
OR explant if large number of electrodes fail
Device programing
Malfunctioning electrodes should not be included in pt’s MAP
Ensure voltage compliance
If voltage compliance reached, increase pulse width, reduce upper stimulation levels; change electrode coupling
Monitor electrode function over time
Intraoperative to postoperative changes
Changes across postoperative intervals
Electrically evoked stapedial reflex (ESR)
Electrical counterpart to the AR
Is muscular contraction within the ME in response to loud sounds- stimulus is electric current
Involved afferent and efferent sensory motor neurons
Measurement of the ESR protocol
Stimulus is at 500 ms duration pulse train
Probe is placed in the contralateral ear to the CI
Select reflex decay setting
Stimulus presented via the programming software
ESR appears as a downward deflection, time-locked to each stimulus presentation
Measurement of ESR Threshold protocol
Use an ascending approach for the stimulus level
Begin with a relatively low stimulus level to avoid overstimulation
Once a clear response is seen, stimulus level should be decreased in small current step sizes
Threshold is the lowest visible repeatable response on the descending run
Measured on as many channels as possible
Clinical uses for ESRTs
Can confirm device function
Assist with speech processor programming
ESRTs tend to appear near the upper comfort levels
ESRTs present in _ % of CI pts
65-85
Setting upper stimulation levels for adults
Always measure behavioral upper stimulation levels if possible
Measure ESRTs if you want an objective measurement of loudness to compare to behavioral thresholds
Setting upper stimulation levels in children
ESRTs rec for setting upper levels
In children, behavioral measures can overestimate comfort levels
Electrically evoked compound action potential (ECAP)
Is a synchronous physiologic response from an aggregate population of nerve fibers in response to electrical stimulation
Electrical version of Wave I of the ABR
Characterized by
A negative peak (N1)
A positive peak or plateau (P2)
ECAP waveform characteristics
Amplitude increases w/ stimulus level
Amplitudes can be as large as 1-2 mV
Faster stimulation rates degrade neural response
Very few averages needed
ECAP stimulus
Measured using reverse telemetry system (NRT)
Single biphasic current pulse delivered using monopolar coupling
Low stimulation rate
Parameters set by programming software
Delivered to the implanted array via the speech processor that is connected to the programming computer
ECAP recording
Recorded using cochlear implant electrodes
Measured ECAP is the voltage change (as a function of time) produced by depolarization of the auditory nerve
Voltage change measured across an intracochlear recording electrode and an extracochlear MP ground electrode
ECAP measurement characteristics
Pts do not need to sleep, lie still, or be sedated
Not influenced by maturational effects
Measured within the cochlea
Artifact is quite large
Artifact reduction: forward-masking subtraction
Takes advantage of neural refractory properties by separating the ECAP from stimulus artifact
A-B+C-D= response
Artifact reduction: alternating polarity
Polarity of stimulus is reversed
Artifact reverses polarity
Physiological response does not change
C and A traces are averaged together
Disadvantages of alternating polarity (3)
Amplitude and latency of ECAP can differ for anodic vs cathodic leading pulses
Averaged waveform may have slightly different morphology compared to either polarity alone
Thresholds obtained may be elevated relative to those obtained with forward masking subtraction method
Threshold and growth of response
Magnitude of the ECAP increases with greater stimulus level
Peak-to-peak amplitude plotted as a function of stimulus level
Plotted as an input-output function
Amplitude growth function
Amplitude growth function: slope
Slope
Rate of ECAP growth
Calculated by applying linear regression to the data points in AGF
Amplitude growth function: threshold
Minimum amount of current needed to elicit a measurable neural response (ECAP). Two methods
AGF linear regression line used to extrapolate stimulus current that yields ECAP amplitude of 0
Used by all manufacturers
Visual detection
AGF threshold: linear regression
Typically results in lower thresholds
If noise floor is low, both methods are highly correlated
AGF functions can flatter or roll over at top of function or have long shallow tails
Need at least 3 suprathreshold measures to reasonably performed regression analysis
What is a measurable ECAP response?
Morphology and latency
Compare waveform to more robust waveform obtained at higher stimulus level
Scaling
Change amplitude scale and zoom in
Noise-floor
ECAP should be larger than the noise-floor
Clinical uses of ECAP
Indicates stimulus level that should be audible
Confirms responsiveness of auditory nerve to electrical stimulation
Confirms device function
Objective baseline
Weak to moderate correlation with T and C/M levels
Assist with sound processor programming of T’s and C/M levels
ECAp thresholds are usually above T’s regardless of pulse rate
ECAP threshold tends to occur in upper portion of the behavioral DR
Setting MAP levels: Hughes
ECAP thresholds measured on all electrodes
Behavioral T’s and C/M’s measured on a single electrode in the middle of the array
ECAP threshold function shifted up (for C/M levels) or down (for T’s) by difference btwn ECAP threshold and measured T’s and C/M’s
Setting MAP levels: Smoorenburg
Measure ECAP threshold on every electrode
T’s and C’s set to same level as ECAP threshold
T and C- levels decreased until they are inaudible
Activate live speech
T and C globally increased until a behavioral response is noted
T level
C levels increased to a level that should provide adequate stimulation without discomfort
Very subjective
Clinical use of ECAP: ability to predict MAP levels etc
ECAP thresholds alone cannot predict MAP levels with adequate accuracy
Should never substitute for behavioral measures
If ESRT cannot be measured (does help with setting upper levels), ECAP can at least provide a level of stimulation that should be audible
ECAP threshold closer to C/M levels for adults
Weak relationship b/w ECAP and behavioral measures may be due to
Orientation of electrode array
Subject differences in neural refractory properties (ECAP measures using a slow stimulation rate)
Differences in electrode arrays
Treating each electrode as an independent data point
AB first generation devices
Clarion S-series
Electrode array with 8 channels
CIS and CA processing strategies
AB second generation
Allow simultaneous stimulation
Clarion CII
Electrode array with 16 contacts
Modified CIS strategy called HiRes
AB current technology
HiRes Ultra 3D Implant System
Similar to HiRes90K
16 electrode contacts
Marvel sound processors
Naida CI M90
Naida CI M30
Sky CI M90
Neptune processors (discontinued)
First fully waterproof device
AB current arrays
HiRes Ultra electrode family
HiFocus Slim J: lateral wall, preserve cochlear structures
HiFocus Precurved Mid Scala array
Other
HiFocus Helix
AB sound processing schemes
HiRes 120 Fidelity with current steering
HiRes Optima
ClearVoice vs SoftVoice
ClearVoice: looks at each channels to determine if speech or noise to enhance the channel with speech
SoftVoice: increase audibility of soft sounds
Legacy internal device: Nucleus C122
First generation
Used 22-electrode array, straight, rested against lateral wall after insertion
Titanium case, receiver magnet holds external transmission unity in place
Body worn speech processor
Bipolar or common ground stimulation
Legacy internal device: Nucleus 24M
2nd generation
24 electrodes
Longitudinal array
Single ball electrode for monopolar stimulation
Bipolar or monopolar stimuluation
SPring processor is body worn
ESPrit processor is ear level
Current internal device: Cochlear Americas
Internal device: profile plus series
Compatible with
Cochlear nucleus Kanso
Nucleus Kanso 2
Nucleus 7
Nucleus 8
Magnet pocket
Nucleus Legacy sound processors: Spectra processor
Uses SPEAK or MPEAK
Directional mic housed in BTE
22 electrodes in a longitudinal array
Nucleus legacy sound processors: ESPrit 3 G Processor
Designed to use with the Nucleus 22 array and Nucleus 24 array
Built in T-coil and wireless Phonak microlink to use with public systems that use FM or infrared tech
Can store two programs
Info directed to 20 electrodes
CIS, SPEAK, or ACE
Uses high power zine air batteries
CA current sound processors and strategies
Sound processors
Nucleus 8
Nucleus Kanso 2
Strategies
SPEAK
ACE
CIS
NRT telemetry system for Nucleus devices and color codes meaning (green, red, brown)
Process for obtaining data from the implant device and sending it back to the external part
Uses RF link to transmit info
Tells use about the goodness of the RF link and integrity of the device
Can transmit electrophysiological data
Green: within operating range
Red: short or open circuit
Brown: Previously flagged
MED-EL 1st generation internal device, sound processor, and electrode arrays
Internal device: Combi-40+
Ceramic+non-removeable magnet
Sound processor
CIS+
OPUS 1 and 2 sound processors are backward compatible
Electrode arrays
Standard
Medium
Compressed
Split
MED-EL current system
Internal device: Synchrony 2
Titanium
MRI safe up to 3 T
Sound processors
Rondo 3
Sonnet 2
MED-EL Fine structure processing
Timing of stimulation is used to code the temporal fine structure of the sound signal in the low to mid frequencies
At the basal end, tonotopic fine structure is achieved by creating virtual channels
FSP may provide better pitch discrimination and a wider pitch range than CIS
Reasons for revision implantation
Infection
Rejection of implant by body
Extruding implant rx
Electrode array damage
Implant failure
Technology upgrade
What can a pt experience with failure?
Loud popping sounds
No sound
Sound quality changes/speech performance decreases
Intermittent sound
Loss of lock to internal device
Episodes of overly loud stimulation
Non-auditory symptoms (pain, shocking sensations, vertigo, and/or facial twitching)
3 categories of CI failures
3-8% of CI are failures
Hard failure
Soft failure
Medical failure
Hard failure
Device stops working due to either internal electronic malfunction or a structural problem
Most common indication for revision 40-80%
Revision CI can take place expeditiously
41% of hard failures have a history of what?
Head trauma
Soft failure
Occurs when a device malfunction is suspected but cannot be proven by the manufacturer’s test
Gradual decreases in performance scores (speech perception)
Non-auditory complaints for soft failure
Facial nerve stimulation
Vertigo
Tinnitus
Intermittent functioning
Soft failure consensus of 2005 defined as:
Soft failure
Uncommon occurence
Device malfunction is suspected but cannot be proven using currently available methods
Reimplantation with subsequent alleviation of symptoms strongly supports the dx but cannot conclusively confirm a device malfunction
Medical/audiological assessment of soft failure in pts
Symptoms
Auditory vs non auditory
Medical eval
Electrode position
Infection
Neural
Intracochlear
Audio eval
Drop in best performance
Failure to achieve expected benefit
Change in sensitivity
Device assessment
Internal
Faulty external hardware
Device integrity testing
Suggested checklist of soft failures for adults
Auditory
Nonauditory
Performance
Mapping
Loss of channels
Changes in impedance
Short/open circuits
Hardware
Replacement of all externals
Objective assessment
Neural response measures
Stimulus artificat
Evoked potentials
Suggested checklist of soft failures for young children
Behavioral
Teacher/therapist concerns
Other factors
Educational placement
Type and amount of therapy
Family involvement
Puberty
Present auditory symptoms, no non-auditory symptoms, and present decrease in performance
Consider revision if symptoms are severe and persistent
Present auditory symptoms, non non-auditory symptoms, no decline in performance
Monitor
No auditory symptoms, present non-auditory symptoms, no drop in performance
Consider revision if symptoms are serve and persistent
No auditory symptoms, no non-auditory symptoms, decline in performance
Remap, replace external hardware, re-eval in 1-3 months
Criteria for determining soft failure in children are determined by what outcomes?
Inner ear development
Central auditory pathways
Cognitive abilities
During post implant follow up it is difficult to provide a concise definition of clinical improvement
Soft failures commonly due to which medical outcomes?
50%
Meningitis
Congenital inner ear malformation
Asthma
Facial nerve stimulation soft failure protocol
Deactivation of 5 or more electrodes may suggest impeding device failure
When reprogramming or deactivation of electrodes does not solve FNS consider revision
Allows broader programming options
Replace an outer wall straight electrode with newer perimodiolar array
Most common causes for medical failures
Device extrusion
Trauma to site
Wound infection
Consider what for device salvage after infection?
Infectious agent
Severity of the infection
Response of the antibiotic tx
Extrusion
Revision of scalp flap and initiate antibiotics
Receiver stimulator may need to be moved to a different location on the skull
Delayed reimplantation of a new device
Electrode extrusion causes and theories
Cause is unknown but common in cases of ossification
Theories
Occurs with skull growth
Force placed on the outer wall by the electrode force it out of the cochlea
Silicone allergy
Suspect in cases of delayed device extrusion with negative cultures or non responsive to antibiotics
Custom made implants can be designed for reimplantation
Technology upgrade
May not mean improved performance
May be due to reduced plasticity of nervous system or inability to adjust completely to the new device
CI revision outcomes
Several hundred revision CIs have been performed
Revision implantation is generally well tolerated and shows stable or improved outcomes
Mean threshold and speech performance scores are either the same or improved
Children < 18 years of age that underwent revision cochlear implant surgery evaluated for open speech rec, subjective reports, and symptom resolution
Previous peak performance was more likely to be achieved or exceeded in younger than in older children
A decline in speech perception was a stronger predictor of successful outcome than chronic underperformance- children often did well after re-implantation
Suspicion of soft failure?
Contact the Clinical Application specialist from the manufacturer to complete integrity testing
Consult with CI team and Patient
Surgeon
CT scan, primary care clearance, cardiology clearance
Audiologist
Complete speech performance measures
Realistic expectations