A quality management program consists of what 4 main components?
Quality Assurance, Quality Control, Quality Planning, and Continuous Quality Improvement
What three topics are assessed when determining the quality of healthcare delivery?
Structure, Process, and Outcomes
What is the term that describes a set of activities that are completed for a service or product to ensure that it will meet quality standards?
Quality Assurance
What type of testing is completed when a new device or piece of equipment arrives in the radiation oncology department to ensure that it meets the agreed-upon manufacturer’s specifications?
Acceptance
What are 3 common safety procedures used in quality assurance of external beam radiation therapy equipment?
Radiation Protection Survey, Interlocks, and Patient Monitoring Equipment
What is completed by a physicist to determine operating parameters, such as output, field size dependence, and depth dose, of a new machine?
Machine Commissioning
Which of the following is NOT recommended to be tested daily by the AAPM for linear accelerator QA?
A. Door Interlock
B. KV/MV/CBCT Uniformity and Noise
C. Laser Localization
D. Beam On Indicator
B. KV/MV/CBCT Uniformity and Noise, tested MONTHLY
Which of the following has a recommended IMRT tolerance of 2 mm?
A. Optical Distance Indicator (ODI)
B. CBCT Positioning/ repositioning
C. Jaw Position Indicators Asymmetric
D. All of the above
A. Optical Distance Indicator (ODI), the CBCT positioning/repositioning and the jaw position indicators (asymmetric) have a tolerance of 1 mm
What does TG -142 provide quality assurance recommendations for?
Medical accelerators
What two organizations have become partners to provide accreditation of radiation oncology facilities?
The American Society for Radiation Oncology (ASTRO) and American College of Radiology (ACR)
What percentage should the output constancy of all x-ray energies remain within for an IMRT treatment machine daily QA?
3%
For a remote afterloading system, what must be done before the first use to verify the geometry and positional accuracy of the applicator?
Autoradiograph
What are the 4 major sources of uncertainty that are accounted for in TPS QA?
Algorithm Inaccuracy, Inaccuracies of Data Entry, Inaccuracies of Data Output, Inaccuracies in Measured Data
How often should an ionization chamber and electrometer be submitted to an accredited dosimetry calibration laboratory (ADCL) for calibration?
Every 2 years
What is the tolerance for the alignment of a CT simulator’s gantry lasers with the center of the imaging plane?
± 2 mm