Fluoroscopy is a … radiographic examination, compared to
diagnostic, which is static
dynamic
Radiographers should perform fluoroscopy only for…examinations, where
diagnosis is not involved
static
Image intensification tubes were developed in 1948 and they resulted in video camera and monitor systems which began to replace
Mirror systems
Currently used only for studies that require observation of dynamic physiological function
Fluoroscopy
Types of fluoroscopic equipment
C-arm mounted
Carriage
A fluoro carriage typically includes what?
intensification tube or x-ray tube
Fluoro x-ray tubes are similar to diagnostic x-ray tubes but how do they differ?
operate at longer periods of time w/ much lower mA
Fluoroscopic mA range
0.5-5.0 mA
A fluro tube target must be fixed at what SID?
15” (38cm)
Until when was image intensification technology developed
1948
Design to electronically amplify the brightness of an image
Image intensification tube
Modern image intensifiers are capable of increasing the image brightness to what?
5000-8000x
The input screen consists of a 0.1 - 0.2 mm layer of …. coated on the concave surface of the image intensification tube
Cesium iodide
The input screen surface may be made of glass, steel, or aluminum and ranges from … in diameter, depending on the manufacturer and use
6”-23”
What does the Input screen do?
Absorbs x-ray photons & emits light photons
What does the photocathode do?
Absorbs light photons & emits electrons
The process of converting x-rays to light to electrons is called
Photoemission
a negatively charged electrode that converts light into electrons using the photoelectric effect
Photocathode
What charge is the anode?
Positive
How much kV is the anode normally supplied w/?
+25 kV
Where is the anode positioned?
Inside glass envelop in front of output screen
What permits the accelerated electrons to pass through the anode field and onto the output screen
A hole in the center
The output screen is what type of screen?
glass fluorescent screen
What type of phosphor is the output screen?
silver-activated zinc cadmium sulfide
The electrons that strike the output screen are converted into what that exit the tube?
light photons
Is a measurement of the increase in image intensity achieved by an image intensification tube
Brightness gain
Brightness gain formula
minification gain x flux gain
Calculated as the ratio between the area of the input and output screens
Minification gain
Minification gain formula
input screen diameter2 / output screen diameter2
Flux gain formula
#of output light photons / #of input x-ray photons
Provides greater flexibility in all fluoroscopic exams and is standard in digital fluoroscopy
Multifield Image Intensification
When switching to a smaller mode, the image appears
Magnified
In magnification mode, minification gain is…, and fewer photoelectrons are incident on the output phosphor – a dimmer image results
reduced
To maintain the same level of brightness in magnification mode, the x-ray tube mA is… by the ABC which … the PT dose
increases
By increasing dose in magnification mode what is improved?
contrast & spatial resolution
Sensitive to low light (color blind), use for night vision (scotopic vision)
Rod
Are less sensitive to light, response for daylight vision, perceives (photopic vision)
Cones
Accelerate and Focus electrons and directs them toward the output screen
Electrostatic lenses
Leakage radiation must be less than …. at a distance of… from the protective housing
1 mGya/h & 1m
The control panel must indicate the conditions of exposure and must … when the x-ray tube is energized
positively indicate
The SID indicator must be accurate to within…of the actual SID.
2%
How much filtration for exposures using above 70kvp? 50-70kvp? below 50kvp
2.5mm
1.5mm
.5mm
The SSD must be not less than… on stationary fluoroscopes and not less… than on mobile fluoroscopes
38cm & 30cm
The fluoroscopic image receptor assembly serves as a primary protective barrier and must be how much lead equivalent?
2mm Pb
The fluoroscopic exposure control should be of what kind
Dead man
Where is the bucky slot moved during exam & how much of an opening is left?
moved to end of table & 5cm (12.7”)
The bucky slot should be how much lead equivalent?
0.25 mm Pb
The protective curtain should be how much lead equivalent
0.25 mm Pb
The cumulative timer produces an audible sound when time exceeds?
5min