1/125
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
The ability of the microscope to separate small details is defined as what?
resolution
Sections that are approximately 1 cell layer in thickness must be cut at what thickness?
3-4 um (microns)
Microscope lens of x90 to x100 magnification are considered:
intermediate
high dry
oil immersion
polarizing
oil immersion
When cutting sections from paraffin blocks, the most common cause of unsatisfactory sections is what?
a poor blade edge
The regular laboratory incubator maintains a temperature of about how many degrees?
37*C (degrees Celsius)
When the magnification can be changed without the need to refocus, the microscope objectives are said to be:
parfocal
binocular
achromatic
apochromatic
parfocal
If a ribbon splits when cutting paraffin sections, the trouble is most likely due to what?
nicks in the blade edge
Care must be taken when using automatic coverslippers to ensure that:
only No. 1 coverslips are applied
only full baskets of slides are loaded on the coverslipper
the mounting medium applied is correct for long-term storage
the mounting medium reservoir is cleaned daily
the mounting medium applied is correct for long-term storage
The temperature of the oven used to maintain a supply of melted paraffin for embedding tissue is most commonly about how many degrees?
60*C (degrees Celsius)
Doubly refractile particles are examined using:
polarized light
ultraviolet light
electron microscopy
dark field illumination
polarized light
Crooked paraffin ribbons may be caused by what?
nonparallel block horizontal edges
When using a microscope with a x10 ocular and a x40 objective, the total magnification is approximately what?
400
The clearance angle of the microtome blade is routinely within what range?
3* - 8* (degrees)
The ordinary refrigerator, when operating normally, has a temperature that approaches:
0*C
0*F
4*C
40*F
4*C (degrees Celsius)
When a lens for a light microscope has been corrected for 2 colors, it is said to be what?
achromatic
When using paraffin with a melting point of 55* - 57*C, the most common temperature for floating sections on a flotation bath is approximately between what range?
45* - 50*C (degrees Celsius)
Compressed or wrinkled sections may be caused by:
overchilled paraffin blocks
defects in the blade edge
the wrong blade tilt
static electricity
the wrong blade tilt
The scanning objective on the light microscope is found:
above the oculars
at the lower end of the body tube
below the substage condenser
below the iris diaphragm
at the lower end of the body tube
Various size holes are noted in sections of paraffin embedded liver on the flotation bath. With further ribboning, these holes decrease in size and disappear. The most likely cause of the holes is what?
facing the block too aggressively
Birefringent substances are best examined with which of the following microscopes?
fluorescence
electron
light
polarizing
polarizing
(used for substances that are birefringent, doubly refractile, or exhibit anisotropism)
Which microscope is used to view substances that are birefringent, doubly refractile, or exhibit anisotropism?
polarizing microscope
During microtomy, the sections lift from the blade as the block is raised. The most likely, or possible, causes are:
a dull blade
too little blade tilt
too much blade tilt
an overchilled block
defects in the blade edge
a dull blade (more likely)
too little blade tilt
Microscopic examination of an H&E-stained section reveals marked chatter, especially at the edges of the tissue. This was most likely caused by:
a dull blade
too little blade tilt
an overheated flotation bath
overdehydration of the tissue
overdehydration of the tissue
During cryotomy, sections of varying thickness are obtained. This can most likely be corrected by what?
increasing the blade tilt
When checking the pH of a staining solution, the pH meter should be calibrated using a standard solution with a pH value:
closest to that of the staining solution
approximating that of water
in any range
close to 7.0
closest to that of the staining solution
The microwave oven creates heat in staining solutions by:
convection
conduction
nonionizing radiation
electrolytic action
nonionizing radiation
Tissue components can be measured with the light microscope in a process known as what?
micrometry
A microscope that has 2 eyepieces is:
parfocal
achromatic
binocular
orthochromatic
binocular
When a lens for a light microscope has been corrected for 3 colors, it is said to be:
orthochromatic
achromatic
apochromatic
parfocal
apochromatic
The high dry objective on most microscopes has a magnification of approximately:
x45
Tissues are subjected to a series of different reagents in a closed processor by:
tissue transfer
fluid transfer
heat transfer
linear transport
fluid transfer
When processed on a short cycle, tissue must be:
cut thin during grossing
washed after fixation
fixed in neutral-buffered formalin
started in 95% alcohol on the processor
cut thin during grossing
List 3 types of microscopy most associated with histotechnology:
light microscope
polarizing microscope
fluorescence microscope
List 3 types of microtomes:
rotary microtome
clinical freezing microtome
sliding microtome (not used in routine histopathology - cuts large paraffin sections in research)
Living cells are usually examined with the dark-field microscope. True or False?
False
A 3-dimensional image is obtained with the transmission electron microscope. True or False?
False
Another name for the ocular of a microscope is the eyepiece. True or False?
True
The analyzer of the polarizing microscope is placed between the specimen and the eye. True or False?
True
Steel blades are commonly used to section glycol methacrylate. True or False?
False
The angle formed by the block face and the cutting facet of the blade is known as the bevel angle. True or False?
False
When examined microscopically, 8-um sections will show all nuclei in the same plane of focus. True or False?
False
Slow freezing of tissue leads to the formation of ice crystals. True or False?
True
An overheated cryostat motor is frequently caused by ice buildup in the cryostat chamber. True or False?
False
Commonly, an overheated cryostat motor is actually caused by dusty coils. Coils should be cleaned regularly (text pg 26)
If the histotechnologist is skilled, then the most critical factor in the laboratory becomes the care and maintenance of laboratory instrumentation. True or False?
True
Routine care of the microtome requires a thorough cleaning and oiling every 2 weeks. True or False?
False
Dust collecting on the cooling coils of the crystat may cause the motor to run more slowly. True or False?
False
The temperature of the room is very important when cutting frozen sections. True or False?
False
This used to be more important before the new designs of cryostats, which do a great job at maintaining their temperatures inside the chamber while the operator has the door open. The new cryostat design allows for the door of the cryostat to be open for longer periods of time, giving the operator more time to obtain a good frozen section.
Paraffin sections will not adhere well to clean untreated and uncharged glass slides. True or False?
TrueP
Poly-L-lysine is a common additive to the flotation bath. True or false?
False
Poly-L-lysine actually coats certain types of slides
Water in the flotation bath is a possible source of contamination on sections to be stained for microorganisms. True or False?
True
New instruments are validated by the manufacturer and may be put into use immediately. True or False?
False
Silver stains done in a microwave oven will show increased background staining. True or False?
False
Automated embedding can be done using regular cassettes. True or False?
False
Special cassettes are used for automated embedding.
Most tissue contamination from floaters occurs in the deparaffinization steps. True or False?
True
Side-by-side comparisons are the best form of instrument validation. True or False?
True
Some staining kits may have the problem of disproportionate solution volumes. True or False?
True
Match the instrument with the closest correct temperature listed. Some answers may be used more than once or not at all.
laboratory incubator
infiltrating paraffin container
flotation bath
cryostat
refrigerator
freezer
60*C
45*C
37*C
22*C
4*C
-20*C
-50*C
-70*C
laboratory incubator 37*C
infiltrating paraffin container 60*C
flotation bath 45*C
cryostat -20*C
refrigerator 4*C
freezer -20*C and -70*C
Typically used to examine Congo red stained sections
A. fluorescence microscopy
B. polarizing microscope
C. both
D. neither
B. polarizing microscope
Used to examine auramine-rhodamine stained sections
A. fluorescence microscopy
B. polarizing microscope
C. both
D. neither
A fluorescence microscopy
Requires the use of a mercury or halogen lamp
A. fluorescence microscopy
B. polarizing microscope
C. both
D. neither
A. fluorescence microscopy
Aids in the identification of crystals
A. fluorescence microscopy
B. polarizing microscope
C. both
D. neither
B. polarizing microscopy
Used to examine unstained cells
A. fluorescence microscopy
B. polarizing microscope
C. both
D. neither
D. neither
Sections may be compressed, wrinkled, or jammed
A. too little blade tilt
B. too much blade tilt
C. both
D. neither
A. too little blade tilt
Sections may have lengthwise splits
A. too little blade tilt
B. too much blade tilt
C. both
D. neither
D. neither
Sections may show microscopic chatter
A. too little blade tilt
B. too much blade tilt
C. both
D. neither
B. too much blade tilt
Ribbons may be crooked
A. too little blade tilt
B. too much blade tilt
C. both
D. neither
D. neither
Ribbons may not form
A. too little blade tilt
B. too much blade tilt
C. both
D. neither
C. both
May result from improper processing (under and over processing)
A. mushy sections
B. chatter in section
C. both
D. neither
C. both
May result from a dull blade
A. mushy sections
B. chatter in section
C. both
D. neither
B. chatter in section
May result from a nick in the blade edge
A. mushy sections
B. chatter in section
C. both
D. neither
D. neither
May result from cutting too rapidly
A. mushy sections
B. chatter in section
C. both
D. neither
B. chatter in sections
May be corrected by soaking faced blocks
A. mushy sections
B. chatter in section
C. both
D. neither
B. Chatter in sections
Usually maintained at 37*C
A. cryostat
B. flotation bath
C. both
D. neither
D. neither
The device usually maintained at 37*C is the lab incubator.
Usually maintained at -20*C
A. cryostat
B. flotation bath
C. both
D. neither
A. cryostat
Required for sectioning for immunofluorescence
A. cryostat
B. flotation bath
C. both
D. neither
A. cryostat
May cause parched earth artifact in section
A. cryostat
B. flotation bath
C. both
D. neither
B. flotation bath
A too-hot flotation bath is a cause of parched earth artifact in section
Must be kept free of debris
A. cryostat
B. flotation bath
C. both
D. neither
C. both
Cryostat kept free of debris to minimize biohazard exposure risk to operator.
Flotation bath kept free of debris to avoid tissue-to-slide cross-contamination and incorrect patient diagnosis.
Uses a barrier filter
A. fluorescence microscope
B. light microscope
C. both
D. neither
A. fluorescence microscope
Focused by varying the strength of magnetic fields
A. fluorescence microscope
B. light microscope
C. both
D. neither
D. neither
Maximum useful magnification of x1,000
A. fluorescence microscope
B. light microscope
C. both
D. neither
B. light micrioscope
Requires a halogen or mercury lamp
A. fluorescence microscope
B. light microscope
C. both
D. neither
A. fluorescence microscope
Resolution of ~0.2 um
A. fluorescence microscope
B. light microscope
C. both
D. neither
B. light microscope
Is considered an acid solution
A. pH 7.0 buffer solution
B. pH 4.0 buffer solution
C. both
D. neither
B. pH 4.0 buffer solution
Is considered a basic solution
A. pH 7.0 buffer solution
B. pH 4.0 buffer solution
C. both
D. neither
D. neither
Should be used to calibrate pH meter for adjusting the pH of the Warthin Starry staining solution
A. pH 7.0 buffer solution
B. pH 4.0 buffer solution
C. both
D. neither
B. pH 4.0 buffer solution
Should be used to calibrate pH meter for measuring the pH of neutral buffered formalin
A. pH 7.0 buffer solution
B. pH 4.0 buffer solution
C. both
D. neither
A. pH 7.0 buffer solution
Is equivalent to the pH of pure water
A. pH 7.0 buffer solution
B. pH 4.0 buffer solution
C. both
D. neither
A. pH 7.0 buffer solution
(pg w19) The type of microscope used to examine the section shown in image A was a/an:
fluorescence microscope
phase contrast microscope
light microscope
electron microscope
fluorescence microscope
(pg w19) Tissue for the procedure shown in image A must be:
fixed in formalin
fixed in acetone
fixed in B-5
frozen unfixed
frozen unfixed
(pg w19) The technique shown in image A:
uses metallic impregnation
uses a fluorescent tagged antibody
demonstrates birefringent material
demonstrates scanning electron microscopy
uses a fluorescent tagged antibody
(pg w19) The microscope used for the technique shown in image A most likely uses a/an:
electron gun
polarizer
halogen lamp
fluorescent screen
halogen lamp
(pg w19) The section shown in image A is a section of:
skin
liver
pancreas
kidney
kidney
(pg w20) The problem shown in image B could most likely be corrected by:
increasing blade tilt
increasing the clearance angle
tightening the block and blade clamps
moving the blade
tightening the block and blade clamps
(also, decreasing blade tilt)
(pg w20) One cause of the problem shown in image B is:
clamping the block too tightly
too small a clearance angle
a defect in the blade edge
too much blade tilt
too much blade tilt
(pg w20) The problem shown in image B most often occurs when sectioning:
brain
uterus
kidney
lymph node
uterus
(hard tissues)
(pg w20) The problem shown in image C is the result of:
poorly dissolved eosin
overdrying the section
mechanical damage to the section
air trapped under the section
air trapped under the section
(see how the artifact is both round and cracked - likely because extra air underneath the tissue allows both sides to dry quickly, which overdehydrates the tissue to create a parched-earth-type artifact)
(pg w20) The problem shown in image C could most likely be prevented in the future by:
allowing the flotation bath water to stand overnight
avoiding mechanical damage to the section
making certain that the eosin is completely dissolved
drying the section in a convection oven
allowing the flotation bath water to stand overnight
(pg w20) The problem shown in image C occurs during:
fixation
processing
microtomy
flotation
flotation
(pf w21) The problem shown in image D is most frequently the result of:
poor fixation
overdehydration
incomplete clearing
inadequate infiltration
overdehydration
(pg w21) The problem shown in image D could have been helped by:
soaking the faced block
increasing the clearance angle
cutting slightly faster
decreasing the flotation bath temperature
soaking the faced block