Heirarchy of control
Elimination
Substitution
Engineering control
Administrative control
PPE
Dilution solution of common household bleach
Sodium hypochorite 1:10 (10%)
Non viscous spill contact time
15-20 minutes
viscous spills time
at least 30 minutes
Four basic waste of disposal technique
Flushing
Incineration
Landfill burial
Recycling
Biomedical waste segregation
black, green, yellow, yellow w black band, orange, red )
Black - non infectious dry
Green - non infectious wet
yellow - infectious and pathological
yellow w band - toxic chemicals waste w heavy trace of metals
orange - radioactive
red - sharps and pressurized container
Most common type of solution
%w/v
can change with temperature because it depends on the volume of the solution
Molarity
Commonly used for acid based reactions
normality
Unaffected by temperature, used when temp variation are significant
Molality
Forward pipetting
pipetting and mixing sample or reagent to another liquid
Reverse pipetting
For high viscosity
A. dispense B. aspirate
A. 45 degree angle B. 90 degree angle
Aspiration depth
1-2 mm
To deliver
Distilled water
To contain
Mercury
Polypropylene
where pipettes are made of
1st step in cleaning glasswares
pre-soak in soapy water
Acid dichromate or potassium dichromate dissolved in sulfuric or nitric acid.
Cleaning glassware
what water to rinse with cleaning glassware
type 1 or type 2
Sterilize glassware using
Dry heat oven (160-180 c for 1 ½ 2 hours)
The gold standard for clinical laboratory external qc testing
College of American Pathologist
A materianl of known value analyzed with patient samples
Control
Values assigned by the manufacturer
Assayed
QC material
a solution (usualy a serum) constituents but KNOWN, run SIMULTANEOUS WITH THE TEST, to check accuracy and precision.
Control limits are calculated from the
Mean and SD
The ideal control or reference limit is between
± 2SD
Accuracy
nearness or closeness to the true value
Precision
Ability of method to give the same result on the sample repeatedly.
Specificity
to detect an analyte without detecting others
Sensitivity
the amount of analyte in a solution to be detected even in smallest amount.
Reliability
To maintain accuracy and precision for a longer period of time.
Test validity
ability of a screening test to accurate identify diseased and non-disease individuals.
Proportion of the subject with positive test that had the disease
Positive predictive value
Measures of central tendency: (AMM)
Arithmetic mean
Median
Mode
Measures of Variation (SCV)
SD
Coefficient of variation
Variance
Most frequently used of measure of variation and help describe the normal curve
SD
QC CHARTS
Gaussian curve
Cumulative Sum Graph
Multi rule QC
Youden/twin plot (interlab qc)
LEVEY JENNINGS CHART
It will give the earliest indiciaton of systematic errors
CUSUM GRAPH
Recite the westgard control rules: (12, 13, 22, 41, r4, 10x)
12s - rejection or warning rule / for screening purposes
13s - random error
22s - systematic error
41s - last for consecutive, random error
10x - ten data on the same side ; systematic error
used in interlab qc, used to compare results from different laboratories
Youden/twin plot
Trend
QC slowly moving up or down for six consecutive days
qc values distributes themselves on one side or either side of the mean for six consecutive days.
Shift
Control values that are far from the main set of values highly deviating values
outliers
Uses a test chip to measure the analyte in the serum, which reacts with the reagent to produced a colored product.
Thin-Film analyzers
Which measures the light reflected from the colored layer?
Reflectance photometry
The total number of different tests an analyzer can perform
Test menu
Type of calibrator that uses one calibrator with a known value.
One point calibration
Keeps the analyzer clean, functional, and operating properly.
Preventive maintenance.
Uses three or more calibrators with different known values.
Multi-point calibration
It enjoys a semi-permeable membrane to separate protein from analytes, thus eliminating the need for manual deproteinization techniques.
ION SELECTIVE ELECTRODE - this describes the technology like the dialysis or direct ISE METHOD where a semi-permeable membrane separates protein from teh sample, avoiding manual protein precipitation steps.
The maximum number of tests or samples an analyzer can process in one hour
THROUGHPUT
Two consecutive controls greater than 2s above or below the mean. Accept or reject run?
Reject
One control above +2s, other below -2s from the mean. What does this indicate?
This is the R4S RULE, A random error.
Statistical test is used to compare the means of two methods?
Student’s T-test
It detect random error using previous values of individual patient
Delta check
Negative predictive value
The probability that a negative test indicates absence of disease.
Plots that will give the earliest indication of a shift or trend?
CUSUM
Plot best for detecting all types of QC errors?
Levey-Jennings chart
Plot best for comparison of precision and accuracy among laboratories
Youden / Twin plot

To determine get the COV and check the highest CV.
A. 12%
A trend in QC is most likely caused by:
find the cov: mean 76ml and sd=2.3
3%
When reffering to the qc results, what paremeter usually determines the acceptable range?
2SD