Chapter 11 Pro

0.0(0)
studied byStudied by 5 people
0.0(0)
full-widthCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/139

flashcard set

Earn XP

Description and Tags

powerpoint

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

140 Terms

1
New cards

40”

100cm

2
New cards

72”

180cm

3
New cards

Diagnostic –type protective tube housing:

lead lined

4
New cards

type protective tube housing is required to protect patient and imaging personnel from off focus, or

leakage radiation

5
New cards

type protective tube housing protects by restricting the emission of x-rays to the area of

primary beam

6
New cards
<p>What’s the grey out area</p>

What’s the grey out area

Second shutters 🥈

7
New cards

type protective tube housing: housing must not allow leakage radiation measured at 1 meter from the x-ray source to exceed ____mGy/hr.

0.88

8
New cards

when tube is operated at ___voltage at the ___ current that allows for continuous
operation

highest, highest

9
New cards

must be located behind a suitable protective
barrier with a radiation absorbent window that
permits observation of patient

Control panel

10
New cards

Control panel

~ Must indicate conditions of exposure
~ must ____ indicate when the x-ray tube is energized
~ the audible sound indicates tube is energized

positively

11
New cards

~ strong and support patient
~ uniform thickness
~ carbon fiber is common

Radiographic Examination Table

12
New cards

Radiographic Examination Table is ___  to absorb very little radiation

radiolucent

13
New cards

~a measure from the distance of the anode focal spot to IR
~tape measure or laser

SID indicator

14
New cards

SID indicator: distance and centering indicators must be accurate to
within ________ of the SID respectively

2% and 1%

15
New cards

primary beam is adequately collimated so it isn’t larger than the size of the IR

X-ray beam limitation device

16
New cards

X-ray beam limitation device: 

  • light localizing variable aperture rectangular collimator
         - to adjust the ____ and shape of the x-ray beam automatically or manually

size

17
New cards

X-ray beam limitation device:

  • ______
    - currently the predominant x-ray beam limitation device

collimator

18
New cards

1.Confines the useful beam
2.Limits the quantity of body tissue irradiated
3.Reduces the amount of scatter

some characteristics of X-ray beam limitation devices

19
New cards

Useful beam=primary beam=

umbra

20
New cards

is all the radiation that arises from the
interaction of an x-ray beam with the atoms of any object in the path of the beam

Scatter

21
New cards

- the most versatile device for defining the size & shape of the beam.

~It is boxed-shaped and contains the radiographic beam devices.

Collimation

22
New cards

Beam Limitation Devices/Collimation: This system consists of two sets of adjustable _________ mounted within the device at different levels, a light source to illuminate the x-ray field and permit it to be centered over the area of clinical interest, and a mirror to deflect the light beam toward the patient to be radiographed

lead shutters

23
New cards

The upper shutters are mounted as ___ as possible to the tube window to reduce the amount of off-focus radiation (x-rays are
emitted from parts of the tube other than the focal spot of the tube)
~ This ___ patient’s dose

close, reduces

24
New cards

Not all off-focus radiation can be removed because ____ of shutters
can not be placed right under the actual focal spot.

1st set

25
New cards

The ________ of collimator shutters are mounted below the level of
the light source and mirror and function to confine further the
radiographic beam to the area of interest.

2nd set

26
New cards

This set of shutters consists of two pairs of lead plates oriented at
right angles to each other

Second

27
New cards

Each set can be ____independently

adjusted

28
New cards


to minimize skin exposure to electrons produced by photon interaction with the collimator, the patient’s skin must be at least 15 cm below collimator

Skin sparing

29
New cards

Portable or mobile units are required to maintain SSD of
a minimum of ___

30 cm/ 12 inches

30
New cards

a scientific term referring to the brightness of a surface

Luminance

31
New cards

Luminance 

~ quantifies the _____of a light source
~ is determined by measuring the concentration of light over a
particular field of view 

intensity

32
New cards

The primary unit is the candela per square meter, or a __

“nit”

33
New cards

One ____ = billions photons per second being emitted
from a light source through a specific field of view.

candela

34
New cards

If the luminance of the collimator light source is ___, the localizing light beam will ____outline the margins of the radiographic beam on the area of interest on all patients

adequate, adequately

35
New cards


Coincidence between the radiographic beam and the
localizing light beam
~ when using light localizing variable aperture rectangular
collimator the physical size and alignment between the
radiographic beam and the localizing light beam is essential
to eliminate collimator ________of the body structures being radiographed

cut off

36
New cards

Both alignment and length and width dimensions of the
radiographic and light beams must correspond to within
___ of the SID

2%

37
New cards

Coincidence between the radiographic beam and the
localizing light beam- This coincidence requirements are collectively known as:

1. alignment 2. congruence

38
New cards

The maximum allowable total difference in length and
width alignments of the projected light field with the
radiographic beam at the level of the IR MUST be no more
than_________ or 40” which equals 2cm or 0.8”

2% of 100cm

39
New cards

~ radiographic collimators that are part of fixed radiographic
equipment manufactured in the US generally include this limitation
device

Positive beam limitation (automatic collimation)

40
New cards

The ____ consists of electronic sensors in the cassette holder that send
signals to the collimator housing

PBL

41
New cards

When PBL is activated, the collimators are automatically adjusted so
that the radiation field _____ the size of the IR

matches

42
New cards

The PBL feature may be deactivated by?

turning a key

43
New cards

The radiographer must ensure that collimation is adequate by
collimating the radiographic beam so that it is __________!

no larger than the IR

44
New cards

What is the purpose of filtration?

reduces exposure to the patient’s skin and superficial tissue
by absorbing most of the lower-energy photons from the
beam.

45
New cards
  • Removes long wavelength or soft x-rays (or hardens the
    beam)

  • This increases the energy/quality of the beam.

Purpose of filtration

46
New cards

A purpose of filtration: absorbs some of the photons in the beam, it ______the overall intensity (quantity) of radiation

decreases

47
New cards

Filtration: The photons that remain are ____penetrating and less
likely to be absorbed in the body tissue

more

48
New cards

The absorbed dose to the patient decreases when the correct amount and type of filtration are placed in the path of the beam

decreases*

49
New cards

Two types of filtration:

1. Inherent filtration 2. Added filtration

50
New cards

1. the glass envelope encasing the tube
2. the insulating oil
3. the glass window in the tube housing.
-1 – 3 makes up 0.5 mm thickness of aluminum
4. reflective surface of the collimator mirror
-provides almost 1mm aluminum equivalent

Inherent filtration

51
New cards

0.5 mm+1 mm=

1.5 mm of aluminum filtration

52
New cards

consists of sheets of aluminum of appropriate thickness located outside the glass window of the tube housing above the collimator shutters.

Added filtration

53
New cards

Added filtration is ______to service personnel.

accessible

54
New cards

The peak kilovoltage of a given x-ray unit determines the________ of filtration required

total amount

55
New cards

inherent + added filtration=

Total filtration

56
New cards

Total filtration of 2.5mm aluminum equivalent for fixed x-
ray units operating above ____is the required standard

70kVp

57
New cards

Stationary radiographic equipment requires a total
filtration of 1.5mm aluminum equivalent for
machines operating from

50-70kVp

58
New cards

Stationary units operating __________ only
require .5mm aluminum equivalent

below 50kVp

59
New cards

___________equipment require a
minimum of 2.5mm aluminum equivalent total permanent filtration

Mobile and fluoroscopic

60
New cards

the thickness of a designated absorber required to decrease the
intensity of the primary beam by 50% of its initial value

HVL

61
New cards

When should radiologic physicist obtain HVL measurement?

1. at least once a year
2. after an x-ray tube is replaced
3. following repairs that have been made on the housing or
collimator

62
New cards

for diagnostic x-ray beams, the HVL is expressed in

mm of Al

63
New cards

HVL is a measure of beam ____ of effective energy of the x-ray
beam

quality

64
New cards

are used for dose reduction and uniformity of
image. The body parts that may vary in thickness considerably may
be made uniform with a filter constructed of aluminum or lead-
acrylic.

Compensating Filters

65
New cards

Compensating Filters: The device partially attenuate x-rays that are directed toward the__________ area while permitting more x-ray to strike the
thicker or more dense area

thinner, or less dense

66
New cards

_____ filter may be used to improve the image of the toes on a foot
x-ray, the thickest part is over the toes and the thinner part over the
heel

Wedge

67
New cards

The __________ filter, is used in some dedicated chest
radiographic units. This filter is thin in the center to permit adequate
penetration of the mediastinum and thick laterally to reduce
exposure to aerated lung fields.

Trough, or bilateral wedge

68
New cards

consistency in output in radiation intensity for identical
generator settings from one individual exposure to
subsequent exposures

Exposure Reproducibility

69
New cards

Exposure Reproducibility: x-ray unit must have the ability to ______ certain
radiographic exposures for any given combination of kVp, mA
and time

duplicate

70
New cards

Exposure Reproducibility: variance of ______ is acceptable

5% or less

71
New cards

consistency in output radiation intensity at selected kVp
settings when settings are changed from one mA and time
combo to another

Exposure linearity

72
New cards

Exposure linearity: mA X time=

mAs

73
New cards

Exposure linearity, also the ratio difference in

mSv/mAs

74
New cards

Exposure linearity: values between two successive generator stations must be
less than

0.1

75
New cards

linearity can vary only __%

10

76
New cards

AEC is also known as

“Phototiming”

77
New cards

system of a radiographic unit is
essentially an x-ray exposure termination
device that ends the radiation when a
predetermined amount is received by the
sensors

AEC

78
New cards

are designed to produce an acceptable diagnostic image
while limiting the total amount of radiation exposure to the patient

Automatic exposure systems

79
New cards


a device made of parallel radiopaque strips alternated with low attenuation

strips of aluminum, plastic, or wood.

Radiographic grid

80
New cards

Radiographic grid: Placed between _________ to
remove scattered x-ray photons.

patient and image receptor

81
New cards

Most often this device is used when the thickness
of the body part is greater than ___.

10 cm

82
New cards

The use of grids _____patient dose, the benefit
of improved quality of the recorded image is a good
compromise

increase

83
New cards

In new DR equipment, an electronic grid is used.
-it does not require ____ in technique

an increase

84
New cards

~ x-rays pass through an object, and some photons are
scattered
~ radiographic quality is higher when scatter is not
recorded on the radiograph

function of grid

85
New cards

What does this function of grid improve?

grid is designed to block the passage of photons that
have been scattered at some angle from their original path


this improves radiographic contrast & visibility of

detail

86
New cards

Grid ratio and patient dose:
~ grid requires an increase

in mAs

87
New cards

Grid ratio and patient dose:

What happens to patient does when a grid is used?

pt. dose increases when a grid is used

88
New cards

Grid ratio and patient dose:

higher grid ratio (more lead) does what to patient dose?

increases*

89
New cards

Minimal SSD for mobile radiography
~ SSD (source –skin distance) of at ________
is imperative 40” is recommended

least 12” (30cm)

90
New cards

When should portable units be used?

portable units should only be used on patients
who can not be transported to a radiographic room

91
New cards

when SSD is small, pt entrance exposure is

higher

92
New cards

Digital radiography:
~ _____image formed by x-ray photons on a radiation detector – an
electronic image

latent

93
New cards

Digital radiography: since its produced by a ____ - digital image

computer

94
New cards

Digital radiography:

numeric values of a digital image are aligned in a fixed number of
rows and columns –

image matrix

95
New cards

each small box is called a

pixel

96
New cards

_____collectively produce a 2 dimensional representation of the
information contained in a volume of tissue

pixel”

97
New cards

In DR, the radiographic _____ appear as levels of brightness
associated with shades of gray.

densities

98
New cards

is the amount of luminance of a display monitor.

Brightness

99
New cards

The shades of gray that are displayed constitute the _____ in the
image.

contrast

100
New cards

Digital image permits adequate visualization of anatomic structures
because it has better contrast and density, and contrast can be
manipulated to improve overall ___

quality