the purpose of image analysis
to teach how to evaluate projections for acceptability, determine how to improve positioning and technical skills before repeat, continue to improve
histogram x axis
pixel gray shade values
histogram y axis
number of pixels of each gray shade
histogram represents the
subject contrast in the remnant radiation
exposure indicator
readings that denote the amount of radiation intensity (quantity of photons) that struck the IR
why doesn’t exposure indicator measure patient dose
it does not take into account energy level of the x-rays
EI reading is read
by the computer at the midpoint of the defined VOI
resolution
quality of the displayed projection is defined by how well details in the VOI have been resolved
a projection is evaluated for how visible the details are defined by the factors of
brightness, contrast, noise, spatial resolution, magnification, shape distortion
intensity refers to teh
total quantity of x-ray photons that expose the patient and IR
controlling factor for intensity and IRE
mAs
contrast
the ratio of percentage of difference between two adjacent brightness levels
high contrast
black and white
low contrast
very gray
subject contrast is controlled by
kVp
subject contrast is the
difference in radiation intensity in the remnant beam that demonstrates the degree of differential absorption resulting from differing absorption characteristics of the body structure
adequate penetration
demonstrated when the cortical outlines of the densest and thickest bony structures of interest are seen
if cortical outlines are not seen, how do you fix it
increase kVp
saturation
demonstrated as loss of contrast resolution where structures in VOI demonstrate pitch black, overwhelmed digital system
quantum noise is decreased by
increasing the IRE by increasing kVp/mAs
increasing kVp by factor of 15%
doubles IRE
if contrast resolution is adequate
mAs should be adjusted to offset quantum noise
signal to noise ratio (SNR)
used to compare one exposure to the next, ratio between desired signals from the undesired signals that are used to create an image
the higher the SNR is there were more
desired signals than undesired
contrast resolution
degree of difference in brightness levels between adjacent tissues on the displayed image
the higher the contrast resolution
the greater the gray shade difference
subject contrast
magnitude of signal difference in the remnant beam (the xray beam that exits the patient)
contrast resolution is primarily controlled by
bit depth
dynamic range
total span of x-ray exposure values that an image receptor can capture and faithfully represent
dynamic latitude
margin of error allows in technical exposure factors (kVp/mAs) while still producing an acceptable image quality
bit depth
the number of computer bits assigned to store the signal value of an individual pixel (or detector element, DEL)
bit depth determines the number of
gray levels each pixel can display using the formula
overexposure means that the photon intensity reaching the IR is
higher than that required to produce an acceptable projection and an excessive radiation dose was delivered to the patient