1/59
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
FLUORESCENCE SPECTROPHOTOMETRY or FLUOROMETRY
Certain molecules, particularly those with a chromophore and a rigid structure (fluorophore), can be excited by UV/visible radiation, and will then emit the radiation absorbed at a longer wavelength. The radiation emitted can then be measured
chromophores
UV-Vis spectroscopy, we only need
______ for them to absorb UV-Vis
radiation
rigid structure
fluorescence, you need a _______ or fluorophore
excited state
initial absorption of a photon by a
molecule in the sample promotes an electron to an _______
emitting
excited electron returns to the ground
electronic state by ______ a photon
FLUORESCENCE
emission arising from an "allowed" transition that typically has a short lifetime between 1 ns and 10 ns
PHOSPHORESCENCE
emission arises from a "forbidden" transition that typically has a long lifetime between 1 ms and 1 s
FLUORESCENCE
does not change in electron spin
PHOSPHORESCENCE
there is a change in electron spin
polyatomic fluorophore
once electronically excited, experiences vibrational relaxation before emitting a photon, causing a red shift or Stokes shift of the fluorescence spectrum relative to the wavelength at which it was excited
Red Shift or Stokes Shift
change in energy or shift in longer wavelength
Xenon lamps
most common, high intensity and broad wavelength range (UV to NIR)
Lasers
are the highest-intensity source; used in applications where short collection times and small amounts of sample are required
Excitation Wavelength Selector
a filter or a monochromator with a known peak transmission wavelength and bandwidth
Excitation Wavelength Selector
enables fluorescence excitation spectra to be resolved
Sampling Device
includes all optics and other equipment needed to deliver the excitation beam to the sample, collect the emission from the sample, and hold the sample in place
Sampling Device
cuvettes, microwell plates, microarrays, microscope slides, and flow systems
Emission Wavelength Selector
to ensure that the emission wavelength region being detected does not overlap with the excitation wavelength profile
Emission Wavelength Selector
important for the rejection of stray light
- filters,
- monochromators, and
- grating polychromators
______ often are used for emission wavelength selection
Detector
a photomultiplier tube (PMT) or a charge-coupled device (CCD) array
right-angle or 0°/90° geometry
used to measure dilute solutions and other transparent samples
front-face geometry
used to measure optically dense samples
front-face geometry
fluorescence is collected at an angle ≤90°
front-face geometry
i.e., epifluorescence geometry - excitation beam and collected fluorescence are both on the same side of the sample, i.e., a 0°/0° geometry
0°/180° transmitting geometry
used in microscopy
- Wavelengths
- Bandwidths
- detector
These are Instrument-Based Factors
Instrument-Based Factors (wavelengths, bandwidths, and detector)
can introduce measurement uncertainty or bias that is particularly significant when measured values are compared between instruments
intensity of the excitation beam
can change significantly with excitation wavelength or with time
Instrument-Based Factors
linear intensity range of the detection system
Instrument-Based Factors
diffraction efficiency of gratings
optically dense samples (e.g., absorbance A >0.05 at a path length of 1 cm)
does not increase linearly with concentration because of significant absorption of the excitation beam and/or emission (reabsorption) by the sample
Sample-Based Factors
Depends on the nature and characteristics of the sample
optically dense samples
reduce the amount of fluorescence that reaches the detector
photobleaching and photodegradation
fluorescence intensity of a sample may decrease with time of exposure to light because of _____ and ______
temperature
fluorescence intensity of fluorophores is _____ dependent - increase with temperature can decrease in fluorescence intensity
intensity
peak position
spectral shape
fluorophore's fluorescence _____, _____, and sometimes ______, often depend on the environment, including changes caused by the solvent used, the solution's pH, or the species to which the fluorophore is bound
Raman signal
can introduce peaks into the fluorescence spectrum - sample's solvent or matrix
- Instrument-Based Factors
- Sample-Based Factors
What are the factors that affect Quantitation
low-dose
Application:
Determination of fluorescent drugs in _____ formulations in the presence of non- fluorescent excipients.
limit tests
Application:
In carrying out ______ where the impurity is fluorescent or can be simply rendered fluorescent.
binding
Application:
Useful for studying the ______ of drugs to components in complex formulations.
bioanalysis
Application:
Widely used in ______ for measuring small amounts of drug and for studying drug-protein binding
Strength of Fluorometry
A selective detection method and can be used to quantify a strongly fluorescent compound in the presence of a larger amount of non-fluorescent material
complex molecules
Strengths of Fluorometry:
Can be used to monitor changes in _______ such as proteins, which are being used increasingly as drugs.
limited
Limitation of Fluorometry:
The technique only applies to a _____ number of molecules
- UV-absorbing species,
- heavy ions in solution, and
- affected by temperature
Limitation of Fluorometry:
Fluorescence is subject to interference by ______
Raman Spectroscopy
Change of wavelength occurs because of the molecules that deflect light and not because of relaxation
• It is not wavelength dependent
• Does not require the molecule to have a chromophore
• Energy shift in cm-1 (wavenumber) is measured instead of wavelength
The Raman effect is analogous to fluorescence except:
middle-IR
In Raman Spectroscopy: The shifts measured correspond to the wavenumbers of the bands present in the ______ spectrum of the molecule.
Lasers
are used to provide high-intensity radiation in the visible region, generally somewhere between 450 and 800 nm (NIR lasers)
Lasers
it does not excite fluorescence in molecules and has good penetration properties
weakly; strongly
bands that absorbed ____ in middle -IR region will absorb _____ in the Raman region and vice versa
FT-RAMAN AND FT-IR SPECTRA
they provide complementary information
Raman Spectroscopy
Application:
Has potential for identifying complex samples, e.g. drugs in formulations and in pack
Raman Spectroscopy
Application:
Samples such as peptide pharmaceuticals can be analysed for changes in their three-dimensional structure
Raman Spectroscopy
Application:
Provides additional fingerprint identity information complementary to middle-IR spectroscopy
Strengths of Raman Spectroscopy
Complementary to middle-IR spectroscopy but requires very little sample preparation since near-IR (NIR) radiation with its good penetration properties can be used for the analysis
Strengths of Raman Spectroscopy
Increasingly a readily available option on middle-IR FT-IR instruments.
Limitations of Raman Spectroscopy
Limitations:
- Not yet fully established as a quantitative technique
- The solvent may interfere if samples are run in solution