Serogical Pipette

The pipette pump can be automatic (electric) or manual
Pasteur Pipette

Will have a bulb at the top of the pipette
Micropipette

Has a disposable tip and variable volume pipettors
Serological Pipette

Pasteur Pipette

Pipette is disposable in DC Gold
Bulbs at the top are reusable
Used to transfer liquid, not measure liquid
Usually only holds 1-2 ml of liquid
Pasteur Pipette

Pipette is disposable in DC Gold
Bulbs at the top are reusable
Used to transfer liquid, not measure liquid
Usually only holds 1-2 ml of liquid
Micropipettes

AKA variable volume pipettors
Measure small volumes: only up to 1 ml (or 1000 ul) and are the most accurate of all pipettes
Use disposable tips
Different size tips for different size pipettes
ONE TIME USE (usually) and disposable in biohazard bin
How to use a micropipette
NEVER point up

Use the smallest pipette possible without exceeding your desired amount
Example:
If you need 3ul, use a P10 micropipette.
If you need 99 uL, use the P100, but if you need 100 uL, use the P200

what will happen if u exceed the amount the pipette is made for
Tips for using the micropipettes

Tip should be directly vertical and about 1 cm into the water
Tips For Using the Micropipettes

To aspire liquid:
You should be at the first stop position before entering the liquid
Enter tip halfway into the liquid
Release plunge to obtain liquid
To dispense liquid:
Set tip into the dispense vessel
Slowly push plunger to the first stop
Raise tip slightly above the dispensed liquid, then push to the second stop
While still on the second stop, pull whole tip out of liquid
Now, you can release the plunge and dispose of your tip