What 5 things are necessary for x-ray production
1. source of electrons
2. method of accelerating the electrons
3. obstacle free path for the passage of high speed electrons
4. A target which the electrons can interact, releasing x-ray energy.
5. A tube to provide a vacuum environment to prevent rapid oxidation of the elements.
Encased in a glass envelope that forms a vacuum.
x-ray tube
what 2 sides does the x-ray tube have
Cathode and anode
purpose of the tube being cased in the metal housing
prevent stray radiation
protect the glass envelope from damage
purpose of oil in the tube
absorbs heat generated
The cathode side consists of a coiled wire filament that has what purpose
emits electrodes when heated
Filament in the cathode side is made of what metal element
tungsten (W-74)
The filament is heated by a(n):
low energy circuit
mA (milliamperage)
The amount of energy in the low energy circuit
What happens when the milliamperage is applied to the filament
heats up
released electrons
The filament is located in a concave cup called the:
focusing cup
what is focusing cup made of
molybdenum
Because of the shape of the cup, the electrons are:
forced thru a smaller area
directed towards the anode side more forcefully
The anode consist of what type of target made of tungsten placed on a cylinder base
beveled (15-23 degrees)
The anode is covered in copper for what purpose
conduct heat away from the target and into the oil
2 types of anodes
1. Stationary
2. Rotating
2 examples of stationary anode
dental
small portable x-ray units (barn/farm)
Disk shaped and rotates on an axis through the center of the tube
Rotating anode
This rotation provides:
cooler surface for the electron stream
provides a small focal spot
why do some machines not allow you to take a second exposure
The heat exceeds 3500F for a short time so you have to wait until the heat dissipates
Because of the angle (usually 20 degrees) of the anode, the x-ray beam is usually stronger on the cathode side of the film.
Heel Effect
how does the heel effect affect positioning the animal for x-rays
place the thicket part of the animal toward the cathode side or the right
place the head toward the anode side or the left
small area of the target on the anode which the electrons collide
Focal Spot
If the focal spot was the size of a pinpoint, the x-rays would have
greater clarity
halo effect on the radiograph
penumbra
As the focal spot increases, what will happen to the quality of the radiograph
the sharpness decreases causing a halo effect
the focal spot can not be the size of a pinpoint because:
it can’t withstand that much heat
Stationary machines must have a (smaller or larger) focal spot to accommodate the high temperatures
larger
Stationary machines must have a (smaller or larger) focal spot
smaller focal spot
The exposure switch/pedal has how many stages before an x-ray is taken
2 stages
First stage
activates the rotor
heats the filament and transformers
Second stage
activates the exposure of radiation
Constant firing of electrons on the cathode can:
damage the filament
cause evaporation
When the target overheats, how does this affect the glass envelope
Metal deposits that form on the inside of the glass
deposits act as a secondary anode and skew with the tragectory of the electrons
arching
What happens when there’s a crack and air is within the glass envelope
air interacts with the electrons and decreases the x-rays produced or collides and redirects the beam
Oil leaks from the metal housing causes:
overheating and will destroy the tube
195. The focusing cup of an x-ray machine is a
A. Small depression where the filament is placed
197. The filament on an x-ray tube is
D. Tungsten coil that emits electrons when heated.
200. Tube saturation can occur with a
C. Too low kVp reading
196. The target of an x-ray machine is a/an
C. Area of the anode struck by electrons during an exposure
225. With the heel effect, the x-rayy beam intensity is greater toward the
C. Cathode
210. The electron cloud is generated at the
B. Cathode
1. When positioning an animal for radiographs, it is best to place the thicker portion of the patient toward the ______end of the tube to produce a radiograph with uniform density.
C. Cathode
2. A small filament produces an image of
C. Greater detail than a large filament
3. Which of the following is a false statement regarding portable x-ray units
D. Have a rotating anode tube and single focal spot.
5. X-rays are formed on the
A. Anode