qualified medical physicist
to ensure proper functioning of digital display device, it is essential that a comprehensive QC program be implemented under the supervision of a ______?
the society of motion picture and television engineers (SMPTE)
Describe the format, dimension, and contrast characteristic of a pattern used to make measurements of the resolution of display systems
National Electrical Manufacturers Association (NEMA - DICOM)
they present their work as a document known as the gray scale display function (GSDF). Used for consistent grayscale appearance
GSDF
it is now mandated for all digital display devices
Video Electronics Standard Association (VESA)
they provide a set of instructions that can be used to help in the evaluation of system performance according to the compliance standard
FPDM
flat panel display measurement
DIN 2001 or Deutsches Institut fur Normung
it is developed as an acceptance testing standard to address the requirement for digital display devices
DIN 2001
It called for a joint performance evaluation of the imaging modality and the digital display device
American Association of Physicist in Medicine (AAPM TG 18)
used to perform qualitative and quantitative tests
photometer, illuminance meter, and colorimeter
what are the test tools
Photometer
the luminance response of monitors and luminance uniformity measurement
0.05 to 1000 cd/m2 , 5%, 0.01
photo meter should be able to measure luminance in the range of ___ to ____ with a better than ____ and a precision of at ___.
3%
Commission Internationale de I'Eclairage (CIE) standard photopic spectral response within the range of
Peter North
he is the first to use a near-range photometer during digital display device evaluation
Illuminance Meter
it evaluate monitor display reflection and to assess ambient light condition
Colorimeter
to measure CIE-specified color coordinates of a digital display device
Geometric DIstortion, reflection, luminance response, display resolution, and display noise
what are the digital display device quality control
Geometric Distortion
by filling the entire screen with the test pattern, one can look for pincushion and barrel-like distortion
TG 18-QC and TG 18-LPV/ LPH test patterns
Used in the Visual assessment of geometric distortion
Reflection
It appears as specular or diffuse
15 degree angle, 30-50 cm
to assess specular reflection, you must turn off the monitor and look for sources of illumination within a ------ of observation at an approx. distance of -----. Look for image of various light sources and any high-contrast pattern from the viewer's clothing or the surrounding
low contrast pattern, ambient lighting and total darkness
to evaluate diffuse reflection, one has to observe the threshold of visibility for ------ under -----and -------
Luminance Response
used to perform a qualitative evaluation of the luminance response of a digital display device
30 cm
luminance response patter should be evaluated from a distance of approx. -----?
171 cd/m2
maximum luminance value should be greater than ---?
magnifying glass
the CX pattern in the middle and in the corner can be evaluated with a ------ and compared
Display Noise
any high-frequency fluctuations or patterns that interfere with detection of the true signal
TG 18-QC
Routine QC test by the QC technologist using the ______ test patterns
specular reflection
reflection of light from a monochrome monitor
diffuse reflection
reflection with a color CRT
reading environment and digital display devices