Initial RR Setting
10-20bpm
Initial RR Setting for SIMV on VC
6-10 bpm
What are the special considerations for RR on VC?
Ve = 5-8L/min
Up to 20
What are the special considerations for RR on PC?
Up to 20
Or same as previous setting on VC mode
What are the initial settings for VT on AC/VC?
6-8ml/IBW
What are the special considerations for VT on AC/VC?
Keep Pplat < 30
Keep PIP < 40
Initial settings on Pinsp (above PEEP)
Pplat on VC
Special considerations on Pins above PEEP
PIP - PEEP (if Pplat not known)
Adjust to achieve target VT of IBW
15-25 = good start
< 35cmH2O
Initial settings for FiO2
40-60% if not known
What are the special considerations for FiO2 on VC or PC?
100% for emergencies or unstable
If patient on O2:
match their prior FiO2
Initial Settings for PEEP on VC or PC
3-5cmH2O
What are the special accommodations for PEEP on VC?
Max is 8cmH2O
What are the special accommodations for PEEP on PC?
Higher for ARDS
Same as prior settings from VC
What are the initial settings for Flow on VC?
40-60L/min
What are the special considerations for Flow on VC?
Consider I:E ratio, IT
ex: I:E = 1:2
Pressure trigger -1 to 1cmH2O
Flow trigger 1-3L/min
Initial Settings for IT/I:E ratio on AC/PC?
0.8-12 sec
I:E Ratio goal = 1:2
Special considerations for IT/I:E ratio on AC/PC
Obstructive: longer exp times
ARDS: may use IRV (Inverse ratio ventilation) to improve oxygenation → 2:1
What are the alarm settings for High Pressure?
10-15cmH2O above PIP
What are the alarm settings for Low Pressure?
10cmH2O below PIP
What are the alarm settings for High Frequency?
10-15 above observed rate
What are the alarm settings for Low Frequency?
10 below observed rate
What are the alarm settings for High Minute Volume?
3-5L above observed MV
What are the alarm settings for Low Minute Volume?
2-3 below observed Minute Ventilation
What are the alarm settings for high VT?
200-300mL above observed VT
What are the alarm settings for Low VT?
100mL below observed VT
What are the apnea alarm settings?
20 seconds
IBW for Male 5ft and over equation
106 + 6 (total inches - 60inches)
IBW for Female 5ft and over equation
105 + 5 (total inches - 60inches)
IBW for Male under 5ft equation
106 – 2 (inches under 5 feet)
IBW for Female under 5ft equation
105 – 2 (inches under 5 feet)
Initial Settings for BIPAP for IPAP
Inspiratory positive airway pressure (IPAP)= 8-12 cmH2O (up to 15)
Increase 3-5 cmH2O to improve ventilation
Initial Settings for BIPAP for EPAP
Expiratory positive airway pressure (EPAP)= 4-6 cmH2O (up to 8)
Increase 2-5 cmH2O to improve oxygenation
When EPAP is increased, remember to increase IPAP same amount to maintain Pressure Support (PS) to maintain ventilation
Remember IPAP-EPAP= PS
What is a normal vital capacity? in weaning.
65-75 mL/kg
(10 x VT)
What is an acceptable value in vital capacity?
≥ 10 mL/kg
(2 x VT)
What’s an unacceptable value in Vital Capacity?
< 10 mL/kg
(< 2 x VT)
What is a normal MIP or NIF aka diaphragm strength in weaning from the vent?
80cmH2O
What is an acceptable MIP or NIF when weaning from the vent?
20 cmH20
What is an unacceptable MIP or NIF when weaning from the vent?
< 20 cmH20
What is a normal value for RR or Frequency for weaning from the vent?
12-20/min
What is an acceptable value for RR or Frequency for weaning from the vent?
8-20 /min
What is an unacceptable value for RR or Frequency for weaning from the vent?
> 20 /min
< 8 /min
What is a normal value for a Spontaneous VT when weaning from the vent?
5-8 mL/kg
What is the acceptable value for Spontaneous VT when weaning from the vent?
≥ 5 mL/kg
What is the unacceptable value for Spontaneous VT when weaning from the vent?
< 5 mL/kg
What is a normal value for Minute Ventilation when weaning from the vent?
5-6 L/min
What is an acceptable Minute Ventilation when weaning from the vent?
< 10 L/min
What is an unacceptable Minute Ventilation when weaning from the vent?
> 10 L/min
What is the normal value for Maximum Expiratory Pressure (aka secretion management → coughing)?
160 cmH20
double MIP
What is an acceptable Maximum Expiratory Pressure (MEP) (aka cough)?
40 cmH20
What is an unacceptable Maximum Expiratory Pressure (MEP) (aka cough)?
< 40 cmH20