1/118
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
fluoroscopy
specialized radiographic equipment used to visualize the motion of internal structures and fluids is?
a dynamic moving image
fluoro will produce?
static
plain radiography is referred to as?
Thomas Edison in 1896
the first fluoroscope was invented by?
a c-arm
a fluoroscopic x-ray and image receptor is mounted to?
-under table unit
-over table unit
what are the two types of fluoroscopy machines?
the carriage
what is the arm called that supports the equipment over the table is called?
control panel
power drive, shutters, image selection =
older machines (like a cassette)
spot imaging is seen in?
a deadmans switch
the foot exposure switch is what type?
over the table
which fluoro type has a higher dose?
true
T/F the carriage must be engaged for fluoro to work?
longer
fluoroscopic x-ray tubes are designed to operate for _______ periods of time at a much lower mA
50-1200 mA
a diagnostic tube operates at?
.5-5 mA
fluoroscopy operates at?
SOD
patient dose is increased in fluoro die to the decreased?
700 mR
UGI approx =
1300 mR
BE approx =
body part
kVp is dependent on the _____________ being examined
high (90-120 kVp)
barium studies use a ______ kVp
low (60-80 kVp)
iodine studies use a ______ kVp
-conventional
-digital conversion (most common today)
what are the 2 types if fluoro technology?
conventional fluoroscopy
in the old days radiologists looked directly at the fluoroscopic screen either held in his/her hands or attached to the primary beam
1948
when were image intensification tubes were developed?
image intensification tubes
the image goes to a TV monitor, spot film device, or cine camera which replaced the mirror system
conventional fluoro
uses an x-ray tube, image intensifier and a television system to produce an analog image
105 mm film
photospot camera =
1000-6000
image intensification utilized the radiation that passes through the patient and enhances the image by electronically making the image ______________ times brighter than the old fluoroscopy
high
for detail to be visible the image brightness must be?
image brightness
image intensifiers were created with the principles advantage being increased?
lumen per square meter or lux
units of illumination is measured in units of?
100-1000 lux
radiographs are visualized under illumination levels of?
dark adapted
before image intensifiers the radiologists had to become ____________ by standing in a totally dark room or wearing red goggles
fluoroscopic screens
-much like x-ray screens
-converts x-rays to light
-phosphor used: zinc cadmium sulfide
zinc cadmium sulfide
fluoroscopic screens phosphors are made of?
image intensifier
tube receives the remnant x-ray beam, converts it into light and increases the light intensity =
input phosphor
the image intensifier x-rays pass through the glass envelope and hit the?
cesium iodide
the input phosphor is made out of?
visible light
the input phosphor converts x-rays to?
photocathode
the input phosphor converts x-rays to visible light which strikes the?
cesium and antimony
the photocathode is a thin metal layer made up of?
photocathode
what absorbs the light energy received from input phosphor and emits electrons?
intensity of the light reaching it
the number of electrons emitted is directly proportional to the
about 25,000 volts
what is the potential difference between the photocathode and the anode?
anode
electrons will be accelerated toward the?
electrons
electrostatic focusing lenses focus the?
-x-ray beam
-input phosphor
-photocathode
-focusing lenses
-hole of anode
-output phosphor
-monitoring system
what is the order of the image intensifier?
light is produced
once electrons strike the output phosphor what happens?
zinc cadmium sulfide crystals
the output phosphor is made out of?
monitoring system
the light is directed to a?
photocathode to the electron beam, electrons to the output screen, output to the human eye
x-ray beam to the input screen, input to the photocathode =
electrons into another light image
x-rays transform into light, back into
equivalent to one another
-1 keV x-ray photon
-1000 light photons
-50 photoelectrons
-3000 light photons
50-75
each photoelectron that hits the output phosphor produces _________ times as many light photons
flux gain
gain in brightness achieved by the II is a result of the electron acceleration and image magnification =
50
a typical II has a flux gain of about?
flux gain
ratio of the number of light photons at the output phosphor to the number of x-rays at the input phosphor
# of light photons output /
# of x-ray photons input
what is the formula for flux gain?
magnification gain
the increase in brightness caused by electrons being focused into a small area =
input diameter /
output diameter and square the answer
what is the formula for magnification gain?
magnification x flux gain
what is the formula for brightness gain?
5,000 to 30,000
the brightness gain of most II’s is?
6” (15cm) to 12” (30cm) in diameter
input phosphors range from?
1”(2.5 cm)
a typical output phosphor had a diameter of?
multi field
most II’s are what type?
examples of multi field
-dual focus
-dual field
-tri focus
-tri field image intensification
dual focus
9”/6” (23cm/15cm) or 10”/7” (25cm/17cm) =
tri focus
10/7/5” (25/17/12cm) or 9/6/4.5” (23/15/10 cm) =
9”
on a 9/6 II the maximum diameter of the input phosphor is?
magnification
the electron focal point is moved further from the output phosphor causing?
increase
mag mode will __________ magnification
resolution
mag mode increases?
less brightness
minification gain is reduced =
needs an increase
when using mag mode what needs to be done to the mA?
false
T/F mag mode decreases patient dose?
vignetting
the loss of brightness at the periphery is called?
automatic brightness control (ABC) or automatic brightness stabilization (ABS)
maintains the brightness of the image by automatically adjusting the exposure factors (mA and kVp) according to the density and contrast of the subject
-flat panel
-charged couple device (CCD)
what types of digital fluoroscopy is used today?
no cassette is required
what advantage does digital have over conventional fluoroscopy?
charge coupled device (CCD)
in digital fluoro the image intensifier output screen image is coupled via a?
analog to digital converter
a CCD converts visible light to an electrical charge that is then sent to the?
cardiac catheterizations
a CCD’s rapid discharge time virtually eliminates image lag and is particularly useful in high speed imaging procedures such as?
CCD
-still uses a II
-a layer of crystalline silicon which illuminates and creates an electrical charge
-these charges are then sorted, pixel by pixel to make the digital image
output phosphor
the CCD is mounted o0n the _______________ of the II
-smaller and lighter than the II and CCD for easier manipulation
-composed of cesium iodide
-distortion free
optical mirror system
-field of view is small
-only one person can use it at a time
cinefluoroscopy or cinecamera
-a movie camera records the images on film
-used in cardiac catheterization
-patient dose is very high
-image quality is high
35mm
16mm or 35mm are used which has a higher quality and higher dose?
framing frequencies
7.5, 15, 30, 60 and 90 =
35mm wide and 400ft
the cine camera film is smaller (_________) and longer (_____) on a roll the camera is capable of operating at 90 images per second
increase in patient dose
increased framing frequency =
30 frames per sec.
what is the usual framing frequency?
the film is in position for exposure
the tube is emitting x-rays only when?
spot film
___________- uses essentially the same technology as conventional screen film radiography
away from the II
when fluoro is on the cassette is positioned?
3-5 and 200-400
when spot filming the cassette is put into position and the tube changes from low mA (______) to high mA (_______)
true
T/F spot films produce high patient dose
the delay involved in moving the cassette into position for the exposure
what is a disadvantage of spot films?
photospot camera
spot film is also called?
spot film camera
receives image from the output phosphor so less patient dose than cassette spot film about half the dose =