1/31
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
|---|
No study sessions yet.
Benefits of increasing TR
Increased SNR, more number slices
Limitations of increasing TR
increased scan time, less T1 contrast
Benefits of reducing TR
decreased scan time, increased T1 contrast
Limitations of decreasing TR
less SNR, less slices
Benefits of increasing TE
increased T2 contrast
Limitations of increasing TE
reduced SNR
Benefits of reducing TE
increased SNR
Limitations of reducing TE
less T2 contrast
Benefits of increasing NSA
increased SNR and signal averaging
Limitations of increasing NSA
Increased scan time
Benefits of reducing NSA
decreased scan time
Limitations of reducing NSA
less SNR and signal averaging
Benefits of increasing slice thickness
SNR increases and coverage
Limitations of increasing slice thickness
Less slice resolution and increased partial voluming
Benefits of reducing slice thickness
more slice resolution and decreased partial voluming
Limitations of reducing slice thickness
reduced SNR and reduced anatomic coverage
Benefits of increasing FOV with fixed matrix
increased SNR, increased coverage, reduced aliasing
Limitations of increasing FOV with a fixed matrix
reduced resolution
Benefits of reducing FOV with a fixed matrix
increased resolution
Limitations of reducing FOV with a fixed matrix
reduced SNR and coverage, increased aliasing
Benefits of increasing receive bandwidth
reduced chemical shift and minimum TE
Limitations of increasing receive bandwidth
reduced SNR
Benefits of reducing receive bandwidth
increased SNR
Limitations of reducing receive bandwidth
increased chemical shift and increased minimum TE
Benefits of using a large coil
increased area of received signal
Limitations of using a large coil
less SNR, more sensitive to artifacts, more aliasing with small FOV
Benefits of using a small coil
More SNR, less sensitive to artifacts, less prone to aliasing with small FOV
Limitations of using a small coil
less area of received signal
benefits of increased phase matrix with fixed FOV
increased resolution
Limitations of increased phase matrix with fixed FOV
reduced SNR if pixels small, increased scan time
Benefits of reduced phase matrix with fixed FOV
less scan time, more SNR if pixel large
Limitations of reduced phase matrix with fixed FOV
reduced resolution