1/109
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
field uniformity
this is the ability of a scintillation camera to produce a uniform image when a source provides uniform distribution photons over a detector.
(Clinically = the ability to produce accurate images of RPx distribution in a patient)
Intrinsic field uniformity
this type of field uniformity is done WITHOUT the collimator and monitors the condition of the sodium iodide crystal and the electronics associated with detector
Extrinsic field uniformity
this type of field uniformity is done with the collimator on and assess the camera as it is in use
Sources of gamma camera non-uniformity:
Mistuning
Uncoupling of PMT from crystal
Cracked crystal
Corrupted/ deleted/ switched-off software connection cables
Collimator defect
Radioactive contamination
X-Y misalignment
Hydroscopic damage to crystal
(Film artifacts NA)
What is mis-tuning in a gamma camera?
The photopeak of the radionuclide does not coincide with the photopeak energy window of the camera (energy windows or high voltages of PMT not set correctly)
Gas filled detectors include:
• Ionization chamber-type survey meter ("cutie pie")
• Proportional counter
• GM survey meter
**dose calibrator
Dose calibrator
this is an ionization chamber that measures radioactivity in curies or Bq
(rather than cpm)
-workhorse of nuclear medicine
-anode and cathode within the chamber
-chamber contains high pressure gas
-outside of cylinder shielded by lead
Dose calibrator operates in _____ region of voltage response graph meaning that all primary electrons created in the chamber reach the anode thou gas amplification.
Ionization
A dose calibrator is a ________ instrument - the number of electrons reaching the anode per second is integrated overtime, so that ir reaches a steady reading over a second or two.
Current-mode
The dose calibrator itself cannot distinguish between ______, so it cannot distinguish between different radionuclides
Photons of different energies
Dose calibrators can measure quantities down to about ______ to within _____.
20 uCi, 5%
Dose calibrators are not able to measure the activity of most _________ directly because they cannot penetrate the liner and wall of chamber to enter the gas space
Pure-beta emitters
GM Survey meter
Device with an audible sound system that alerts the operator to the presence of ionizing radiation
-GM region of graph (Townsend avalanche to increase ionizations)
-qualitative/ semi-quantitative indicator of presence of radiation
Geiger discharge
electrons cause further ionizations and excitations, gas molecules deexcite by UV photon emission, UV photons interact with gas molecules via photoelectric, produce more electrons
GM meters operate in _____ mode
Pulse
Units of GM meter
Counts per minute
mR/ hr
Quality control
An established set of ongoing measurements and analysis designed to ensure that the performance of a procedure or instrument is within a predefined acceptable range
Constancy
QC check in which the instruments response to a long-lived radioactive source or other unchanging/ slow-changing source of radiation is measured - reading of source should be constant from one day to the next
Linearity (not gamma cam)
QC check that describes the ability o a radiation detection instrument to respond to variable quantities in a linear fashion - usually over several orders of magnitude
-activity linearity of dose calibrator
-energy linearity of scintillation detector
Linearity (gamma camera)
The ability to reproduce a radiation distribution consisting of straight lines (bar phantom)
Accuracy
QC check in which an instrument's reading is compared to the true value for the measurement - allowing for correctness of the instrument's output or reading to be evaluated
Geometry
A collection of factors that may affect a radiation measurement
-volume of source
-type of container source is in
-physical location of source relative to detector
Survey meter daily QC
-battery test
-background reading (ensure no contamination)
-constancy using long-lived source
Accuracy and calibration of survey instruments must be completed:
Annually
GM meter is considered out of calibration if accuracy reading is not within _____% of expected.
10
Dose calibrator QC requirements
Constancy - daily
Linearity - quarterly
Accuracy - annually
Geometry - installation/ repair
Dose calibrator is considered accurate if reading is within ____% of reference standard measurement
5
Daily constancy readings of a dose calibrator are considered acceptable for within +-_____%
10
Activity linearity of a dose calibrator can be evaluated in two ways:
1. Measure short-lived source (Tc-99m) at several times points over 2-3 days, then plotting the measured activity vs time on a semilogarithmic graph
2. Use a set of commercially-available lead sleeves and a single high-activity source and compare using known attenuation factors of each sleeve
If a dose calibrator proves to have nonlinearity greater than +- ___% in any activity range, it should be serviced.
10
What is a common way a manufacturer may preform geometry testing on a dose calibrator?
Incremental additions of water to a radioactive solution with measurements after each addition - 3mL syringes and 30mL vials
Limit for geometry variation of a dose calibrator is +-_____%
10
Scintillation detectors
Radiation interacts with crystal such as sodium iodide, cesium iodide, or zinc sulfide to produce very small flash of light. Electronics of instruments amplify this light pulse thousands of times in order to produce useful signal that can be processed. Most useful when detection of very small amounts of radiation is required
Scintillation detectors in NM:
Thyroid probe
Well counter
Gamma camera
PET camera
Daily QC for scintillation detectors
Calibrate high voltage supply (peaking)
Constancy/ sensitivity
Background
High voltage should be:
Nearly constant from day to day
Constancy for scintillation detectors
Long-lived source measured to ensure constant cpm each day before use
Infrequent QC for scintillation
Energy resolution - quarterly
Chi square test - quarterly
Efficiency factor - annually
Energy resolution formula
FWHM = B-A/E x 100%
Sensitivity formula
TP/(TP+FN) ??
Pulse height analysis
can be applied to a series of pulses which have been generated with a size proportional to the radiation energy.
-energy window used to identify photopeak of radionuclide being imaged
-eliminates data of lower energy (bc of scatter)
-operator chooses appropriate percentage window and centers it on the photopeak visualized on the spectrum display (sets upper and lower limits)
Spatial resolution - gamma camera
Ability of a gamma camera to reproduce the details of a nonuniform radioactive distribution
- bar phantom
-measures in mm
- ability to put counts in their correct locations
Peaking
Checking alignment of gamma cameras pulse height analyzer window with the photopeak of the gamma ray being imaged
-scatter free source
Intrinsic uniformity of gamma camera
-collimator removed
-point source placed at least 5x the latest camera dimension
-do not exceed 20K cps
Extrinsic uniformity gamma camera
-collimator on
-flat-field flood source ( water-filled or Co-57 sheet)
-quicker, no need to expose crystal
-2-6 million counts (visual)
-30-120 million counts (quantitative)
Flat-field flood source
Used to test uniformity on a daily basis
Bar phantom
Assess gamma camera spatial linearity & resolution
-5mil counts
- mm
- better than largest lines seen but not better than smallest seen
Causes of gamma camera artifact
1. Wrong energy window
2. PMT problems
3. Cracked crystal
4. Measles (water)
5. Loss of auto corrections
6. Off-peak windows
7. Poor coupling of PMTs
8. Collimator artifact
9. Fillable flood problems
10. Photographic system problems
Sensitivity
Ability of gamma camera to use gamma rays available to it
Sensitivity maps
Uniformity correction maps
Sensitivity check gamma camera
-extrinsic
-1-2 mCi Tc-99m in Petri dish diluted to 3-mm depth
-10 count measurements of 1 min each
-tests collimator more than camera
-sensitivity over time should stable to within 5%
Attenuation correction
correction for the different levels of attenuation a photon might undergo as it gets to the detector
-creating 3D image while taking into account attenuation from tissues etc
-mathematically or patient-specific (attenuation map)
Chang attenuation correction
Oldest and most simple assumes body is an ellipse and uniform attenuation/density through the slice. Attenuation through each pixel is the same. Not used in chest region
-do not consider scatter or detector response
-finds average AC for distance from edge of user-defined eclipse
Patient specific attenuation map
-specific to area being imaged
-CT scan information used to correct for attenuation of tissue is applied to SPECT emission data
-transmission based AC
-external radiation source
Transmission based attenuation correction
Patient-specific attenuation map from external radiation source (CT)
SPECT Uniformity
Intrinsic: point source, no collimators
Extrinsic: flat field flood, collimators
Center of rotation (COR)
reference data used for correcting the slight misalignment of SPECT images taken from various projections
-AOR matches COR
-point or line source
-x-axis offset measured in pixels
-sinogram for visual inspection
COR should look like
Single point of activity in transverse slices
COR artifact 360 acquisition
Donut
COR artifact 180 acquisition
Tuning fork
Causes of COR misalignment
-gravity causes normal wear
-weight of heads
COR should be checked
Weekly
3D uniformity (tomographic) SPECT
-20-cm diameter cylinder with 10-15 mCi Tc99 diluted to fill bottle
-analyzed for integral and differential uniformity
3D resolution SPECT
Jaczak phantom
-rods: resolution
-spheres: contrast
-empty portion: uniformity
SPECT artifacts
1. ray
2. ring
3. motion
4. truncation
Ray artifact
FBP reconstruction only
-very hot or cold lesion is in sufficiently corrected by applied filter
-2D
-remove if possible
Ring artifact
Any abnormalities due to errors in rotational mechanics of detectors
-carried through every projection
-"bullseye"
-flood nonuniformities or collimator defects
-COR errors, head-tilt errors
-only corrected by fixing issue
Motion artifact
Movement of 2 or more pixels during acquisition time
-depends on magnitude of movement and time when it occurs
-review cine
-motion correction programs
Truncation artifact
Transmission scanning
-part of the body is not seen in all projections
-incorrectly place counts/ count boundaries
-more likely with limited FOV collimator (cone or fan beam)
-bc this type of imaging assumes that the entire object is in the FOV from all angles
Attenuation correction in PET
-more of a problem: both photons must be detected for the event to count
-head attenuates 85%, body attenuates 95%
-can be solved exactly - does not depend on location of RPx
-depends on body's thickness
-can be directly measured with CT
Recommended Daily QC for PET systems
PMT baseline check
PMT gain adjustment
blank scan
uniform cylinder/ point source scan
blank scan
this is performed on a PET scanner daily and uses a long lived source (ex Germanium 68) to produce sinograms for the detector pairs in each ring
-air scan / reference scan
-siograms will display any equipment failures
-visual inspection and quantitative data
-non-CT systems use this for attenuation correction
reccomended weekly QC for PET systems
uniformity
well counter calibration
coincidence timing
energy window calibration
coincidence timing calibration
this QC adjusts for the timing differences in the event detection circuitry and is done weekly
-generates corrections for coinidence timing between detector blocks
-also used to measure TOF
reccomended quarterly QC for PET systems
preventative maintenace
detector efficiency/ normalization
cross calibration
reccomended annual QC for PET systems
NEMA NU 2-2007 testing:
-spatial resolution
-sensitivity
-intrinsic scatter
-scatter correction
-count rate performance
update well counter calibration & normalization factors
normalization scan
evaluation of normalization factors that are applied to every study
-most important QC procedure
-must be high quality with little noise
-rod source or Ge-68 cylinder
-scan with low dead time and no pulse pike up
-compare to older version to find statistical differences
-can take 6hr
the normalization scan corrects for several issues that arise from the ____ of PET systems.
multicrystal
multi-ring geometry
important job of normalization in PET is to:
normalize the individual detectors so that they generate a uniform response to a uniform source
cross calibration in PET
-vital to numerical results generated in clinical studies
-measure source in well counter (cpm) and compare to measurement in tomograph
= count rate per voxel
artifacts in PET
electronic malfunction
ray artifact
motion
misregistration
bed-overlap
hot-spot
partial volume
ray artifact
due to FBP processing and corrected using ramp filters
(same as SPECT)
misregistration artifact
occurs because CT and PET acquisition is not completed simultaneously - if patient moes between change in bed position or between the two scans
-AC is not applied correctly
bed overlap artifact
horizontal photpenic stripes seen at junction of bed positions if normalization correction factors are not applied correctly = redo normalization and scan again
hot spot artifact
positron emitting source is used for attenuation correction and applied correction factors are not s large as they need to be
-high dead time in areas
-WB scans around bladder
co-registration
Spatial adjustment of two images obtained by different imaging modalities (e.g., PET or SPECT and CT or MRI). Their combined image is usually visualized using a fused display.
image contrast in CT
visualization of small differences in attenuation between normal and abnormal tissues
-CT can differentiate as low as 0.5%
-contrast is based on tissue density, mA, noise
spatial resolution in CT
ability to resolve closely spaced objects
-bar phantoms
-smallest line pair that can be visualized
-affected by size and spacing of detectors, focal spot, reconstruction matrix, filter
noise in CT
quantum mottle - noise resulting from finite number of x-rays used to make exposure
-CT: directly related to number of xrays detected by each detector
-fewer xrays = greater noise
-affected by imaging parameters
uniformity in CT
how uniform image of homogenous material appears
-water phantom to evaluate for differenes in HU caused by noise
-checks for cupping and beam hardening
daily QC for CT
tube warm up
verfiy tube output & detector response to vairous kVp and mA
uniformity using water phantom
artifacts in CT
beam hardening
attenuation correction artifacts in fusion imaging
noise
scatter
pseudoenhancement
motion
partial volume
motion*
beam hardening
The phenomenon whereby low-energy photons are absorbed as the x-ray beam passes through an object, resulting in an increase in the average photon energy of the beam.
-more low energy photons are asorbed, leaving only higher energy
Radiology Information System (RIS)
A system that collects, stores, and provides information on radiological tests. Main functions include:
registraion/ patient info
scheduling
workflow/ resource management
patient & study tracking
reporting
billing
-large linked databases linked by accession numbers
Picture Archiving and Communication System (PACS)
An integrated computer system that obtains, stores, retrieves, and displays digital images (in healthcare, radiological images)
-displays from all imaging modalities
-all image data must share same format (DICOM)
-biggest challenge s storage space
intravenous infusion pump
A device used to regulate the flow or rate of intravenous fluid
ECG monitor
records electrical activity of the heart for 12 second intervals
pulse oximeter
an external monitor placed on the patient's fingertip or earlobe to measure the oxygen saturation level in the blood
defibrillator
a device that delivers an electric shock to the heart to restore its normal rhythm
glucose meter
device that measures blood glucose concentration