reproducibility
The same technique setting should always create the same exposure rate at any time.
reproducibility tolerance
+/- 5%
-test annually
radiation output
he specific amount of radiation emitted by x-ray generators per unit of x-ray tube current and time.
this is beam quantity
the different xray generators should be producing similar radiation
radiation output tolerance
Maximum variability allowed is +/- 10%.
Test annually.
filtration
Necessary to remove low-energy photons from the x-ray beam.
a pt’s skin dose can increase 90% if we don't remove the low energy photos
Best method to determine if adequate filtration exists is to measure HVL
HVL half value layer
the amount of filtration needed to reduce the intensity of the beam by half
filtration tolerance
over 70kvp, total filtration is 2.5ml aluminum equivalent
-test annually
kvp accuracy
Has a significant effect on image contrast, optical density, and patient dose.
kVp stated on the control panel should produce an x-ray beam with a comparable and consistent energy.
Wisconsin Test Cassette, Ardan and Crook’s Cassette, or a digital kVp meter.
kvp accuracy tolerance
+/- 5% or +/- 4kvp
test annually
timer accuracy
Exposure time directly affects the total quantity of radiation emitted.
-Digital x-ray timer (most accurate), or spinning top test.
timer accuracy tolerance
+/- 5% for exposure times greater than 10ms.
+/- 20% for exposure times less than 10 ms.
test annually
half wave rectified
exposure time (s) X 60
full wave rectified
exposure time (s) X 120
three phase and high frequency
arc size ÷ 360
reciprocity
Refers to the same mAs being selected but with different combinations of mA and exposure time.
reciprocity tolerance
Maximum variability allowed is +/- 10%.
Test annually or semi annually
linearity
Sequential increased in mAs should produce the same sequential increase in the exposure.
when mAs doubles, exposure should double
linearity tolerance
Maximum variability allowed is +/- 10%.
Test annually.
Actual focal spot
Actual rectangular surface on the target where the electrons strike.
effective focal spot
The projected focal spot.
effects detail in image
increase in effective focal spot decreases detail, decrease in effective increases detaik
focal spot blooming
An increase in focal spot size, usually as a result of an increase in mA.
- Increases with age, use, and use of ma stations
-ma stations cuz it increases amount of electrons
focal spot tolerance
Maximum variability allowed for a focal spot size of:
≤ .8 mm is 50%
.8 – 1.5 is 40%
≥ 1.6 is 30%
Test annually.
tools to test focal spot size
pinhole camera, resolution charts, focal spot test tool