Center ray beam QC
Center ray must be within 1% of the SID
Tube distance QC
The actual SID must be with +/- 2% of the indicated SID
Coincidence (collimations) testing
The light field must align with the collimated xray field with +/- 2% of the SID
Tested semi annually
Exposure Time QC
Actual exposure time has to be within +/- 5% of the indicated exposure time (10ms)
Tested annually
AEC Testing
to test AEC capabilities. Multiple exposures with different aluminum thicknesses are done. Results are supposed to have the same exposure receptor each time.
kVp Calibration Testing
Actual KVP must be within +/- 5% or 10% of the assigned KVP
Tested annually
exposure reproducibility
If a set of exposure values (Kvp, MA, time) is selected, then it must have the same radiation results each time within +/- 5%
Tested annually
Exposure linearity
exposure per mAs should not change more than 10% between mA stations (settings)
Tested annually
Focal spot testing
Within 50%
Tested annually
Lead shielding testing
testing annually
Test cracks in lead shields due to folding
Luminance test
Testing grayscale and brightness on monitors
Equal or greater than 350 candela/M-square
Uses SMPTE or TG-18 test pattern
SMPTE test pattern
Used to check monitor performance
Used to evaluate on the screen luminance, contrast, geometric distortion, sharpness resolution