Q: What is the normal programmed set point for the water tank temperature on the LCA2?
180° F
Q: How do you reset the empty function and prime the lines after sanitizing the LCA2?
Normal setting; dispense 12 oz per side
Q: How long must the sanitizing solution remain in the LCA2's product tubing?
5 minutes
Q: What indicates that the silicone grommets and hoses on the LCA2 need replacing?
A: Chalky white appearance, stiffness, or inflexibility
Q: What should happen when testing the LCA2 water tank fill function while dispensing?
Refills in 3 to 5 seconds.
Q: What must be removed to access the water tank level probe for cleaning on the LCA2?
Front splash, top, and both right side panels
Q: How are scale deposits and coffee residue cleaned from the electrodes in the mix chambers?
Manually with a brush and detergent
Q: When verifying the LCA2's dispensed volume, what amounts indicate if the flow restrictor is in place or removed?
18 oz (in place); 25 oz (removed)