Analytical Chemistry Test 2

0.0(0)
studied byStudied by 12 people
0.0(0)
full-widthCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/147

encourage image

There's no tags or description

Looks like no tags are added yet.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

148 Terms

1
New cards

Explain the process of molecular absorption in terms of “states.”

Paired electrons reside at the ground state (S0). When light is absorbed, an electron will jump to a higher electron state (S1, S2, etc.). Within the excited states is various vibrational states to which the electron can move.

2
New cards

What is the effect of intermolecular forces on molecular absorbance spectra?

Intermolecular forces broaden peaks. Therefore, environment effects the peaks. A system in gas phase will produce strong peaks. A system in water will produce broad peaks. A system in cyclohexane will produce peaks in the middle.

3
New cards

What does each variable mean: A=-log(P0/P)?

A= absorption

P0=incident light

P=transmitted light

4
New cards

Why is it important to measure P0 as a blank?

to correct for scattering and reflection losses

5
New cards

Name two light sources for UV-Vis absorption and what spectral range it covers.

1) deuterium arc lamp - UV

2) tungsten lamp - vis/NIR

6
New cards

Name the different types of cuvettes, the range of spectrum they cover, and their cost. What other factors should be considered when picking a cuvette?

1) Quartz - UV, vis, NIR - $90

2) Optical glass - partial UV, vis, NIR - $40

3) Polystyrene - partial UV, Vis - $0.2

4) UV acrylic - UV, vis - $0.8

Other factors: chemical resistance and physical durability

7
New cards

One way to increase the absorbance of your sample.

Increase the path length

8
New cards

Name the four causes of deviations from Beer-Lambert’s law and how to limit each of them.

1) concentration - stay below 0.01 M

2) acids/bases - control pH using excess H+ or OH- 

3) wavelength - measure at peak and use *picture

4) “stray light” at wrong wavelength causes error at high absorbance - note absorbance limit

<p>1) concentration - stay below 0.01 M</p><p>2) acids/bases - control pH using excess H+ or OH-&nbsp;</p><p>3) wavelength - measure at peak and use *picture</p><p>4)&nbsp;“stray light” at wrong wavelength causes error at high absorbance - note absorbance limit</p>
9
New cards

Explain the effect of stray light at high absorbances.

When a sample absorbs a lot of light, less light is transmitted to the detector. Added light reaching the detector (stray light) can result in a lower absorbance value than the real absorbance.

10
New cards

Draw a single beam instrument and a double beam instrument. 

knowt flashcard image
11
New cards

Why is a chopper important in a double beam instrument?

Reduces uncertainty in P0

12
New cards

Draw a single-channel and multi-channel instrument. Note key differences.

Single-channel : sample refracts then enters sample and goes straight to detector, single beams

Multi-channel : light hits sample first them refracts and enters detector, detects and array of light rather than single beam

<p>Single-channel : sample refracts then enters sample and goes straight to detector, single beams</p><p>Multi-channel : light hits sample first them refracts and enters detector, detects and array of light rather than single beam</p>
13
New cards

What are the strengths and weaknesses of molecular absorbance techniques?

Strengths

  • wide applicability (can absorb many chemicals)

  • fast 

  • inexpensive 

Weaknesses

  • limited selectivity (overlap of broad absorbance) 

  • higher LOD than luminescence

14
New cards

Define luminescence.

Emission of light after thermal excitation.

15
New cards

Define photoluminescence and give examples.

emission of of light after excitation by light

examples: fluorescence, phosphorescence

16
New cards

Define chemiluminescence

emission of light after excitation by chemical reaction

17
New cards

Describe the different steps of fluorescence.

1) absorption = molecule absorbs light and is excited from S0 to S1/2

2) vibrational relaxation = molecule relaxes to lower vibrational states by releasing nonradioactive energy

3) internal conversion = molecules transfer energy within itself and virbrationaly relaxes to the S1 state

4) fluorescence = molecule returns to lowest energy state, S0, by emitting a photon

18
New cards

Describe the different steps of phosphorescence. 

1) absorption = molecule absorbs light and is excited from S0 to S1/2

2) vibrational relaxation = molecule relaxes to lower vibrational states by releasing nonradioactive energy

3) internal conversion = molecules transfer energy within itself and virbrationaly relaxes to the S1 state

4) intersystem crossing = transition from S1 to T1 (triplet state), facilitated by spin-orbit coupling 

5) phosphorescence = molecule returns to lowest energy state, S0, by emitting a photon, slower than fluorescence

19
New cards

what is internal and external conversion?

Internal conversion = molecule transforms energy to vibrational energy within itself and moves to lower electronic state

external conversion = molecule transfers energy to another molecule and moves to a lower electronic state

20
New cards

Draw a fluorometer and describe the selectivity of it.

knowt flashcard image
21
New cards

What is the relationship between concentration, absorbance, and fluorescence in a fluorometer?

At high concentrations, there is nonlinear fluorescence and good S/N ratio for absorbance. At low concentrations, there is linear fluorescence, poor S/N ratio for absorbance (absorbance is small), and high S/N ratio for fluorescence (no background interference). 

22
New cards
<p>What does each variable mean? What exactly is it measuring? </p>

What does each variable mean? What exactly is it measuring?

Measures the efficiency of fluorescence.

O = fluorescence quantum yield

Kf = fluorescence rate constant

Knr = sum on nonradiative rate constants

23
New cards

Why do most molecules have weak fluorescence and what can be done to make it more efficient?

Because the non-radiative rate constants are much higher than the fluorescence rate constants. You can increase the fluorescence rate constant and decrease the non-radiative rate constants.

24
New cards

what is fluorescence lifetime and how does time effect fluorescence intensity? What is needed to measure it?

The average time a molecule stays in the excited singlet state before emitting a photon. The intensity of fluorescence decreases over-time. A short light pulse and a fast detector is needed to measure fluorescence lifetime. 

25
New cards

What is dynamic (collisional) quenching and what occurs as a result of it?

When a fluorophore transfers energy to another molecule (quencher) after excitation. It decreases the fluorescence (as a result of increasing the Knf) and decreases the fluorescence lifetime.

26
New cards

What is static quenching and what effect does it have?

A fluorophore complexes with a quencher before excitation. Fluorescence decreases (as a result of a decrease in F). Fluorescence lifetime is not effected.

27
New cards

What is the relationship between the concentration of a quencher and fluorescence?

Increase in concentration decreases intensity

28
New cards

what is an efficient fluorescence method for target DNA and how does it work?

Using molecular beacons and quenching. The molecular beacon binds to to the complementary DNA sequence, separating the fluorophore from the quencher, therefore causing fluorescence.  

29
New cards

How can quenching be used to study biomolecule folding, DNA binding, and membrane penetration?

By measuring the distance between the fluorophore and the quencher

30
New cards

Name types of vibrational spectroscopy. 

Absorption and scattering

31
New cards

What are the different types of scattering and how are they different?

1) rayleigh : photon excites to virtual state and returns to same vibrational level is started at (no energy difference)

2) stokes shifted raman : photon excites to virtual state and returns to a different vibrational level (energy lost)

3) anti-stokes shifted raman : molecule starts in excited vibrational state and returns to ground vibrational state (energy is gained)

32
New cards

Relationship between wavelength of light source and scattered wavelength and raman shift.

The light scattered by the sample depends on the wavelength of the light source. However, Raman shift is independent of the wavelength of the light source.

33
New cards

How can raman and IR spectra look different?

For symmetrical molecules, raman active vibrations will be IR inactive and vice versa. This is due to different selection rules based on quantum mechanics. 

34
New cards

Difference between IR and raman?

IR = absorption, raman = scattering

35
New cards

What are some challenges with IR instrumentation and how can the Michelson interferometer help?

Challenges

  • noisy detectors

  • narrow absorbance bands (deviations from Beer’s Law)

Michelson interferometer

  • multiplex

  • resolution (mirror movement)

36
New cards

What type of wavelength separators are used in raman instrumentation and what are the advantages of each?

Interferometers = longer wavelength laser, multiplex advantage for noisy IR detectors

Spectrograph w/ array detector = short wavelength laser, less noisy UV/vis/NIR detectors

37
New cards

Is IR or Raman more useful for aqueous solutions and why?

Raman because water absorbs strongly in the IR region.

38
New cards

Advantages and disadvantages of short wavelength lasers.

Advantages

  • Raman shift is independent of source wavelength

  • scattering intensity is low

  • allows larger range of raman shift

Disadvantages

  • fluorescence background

  • photodecomposition

39
New cards

Advantages and disadvantages of long wavelength lasers.

Advantages 

  • raman shift is independent of source wavelength 

  • no fluorescence background 

  • no photodecomposition 

Disadvantages 

  • scattering intensity it high 

  • smaller range of raman shift 

40
New cards

What is linearly polarized light?

Equal mix of clockwise and counterclockwise circularly polarized light (equal amplitute, out of phase) 

41
New cards

How is linearly polarized light effected by enantiomers?

Enantiomers have different refractive indices for clockwise and counterclockwise light. Due to this, one is moving faster than the other causing a phase shift, rotating the orientation of the electric field.

42
New cards

What is a polarimeter and how does it work. Draw a diagram of one.

A polarimeter is used to measure the angle of rotation when linearly polarized light passes through an optically active substance. First, a polarizer converts light into polarized light. Then, the polarized light hits the sample. This light then travels to the analyzer (rotating polarizer) where the amount of light that has rotated is measured. The light then reaches the detector.

<p>A polarimeter is used to measure the angle of rotation when linearly polarized light passes through an optically active substance. First, a polarizer converts light into polarized light. Then, the polarized light hits the sample. This light then travels to the analyzer (rotating polarizer) where the amount of light that has rotated is measured. The light then reaches the detector. </p>
43
New cards
<p>What does each variable mean?</p>

What does each variable mean?

  • alpha = specific rotaion 

  • alpha obs = measured optical rotation 

  • pathlength (dm)

  • concentration (g/mL)

44
New cards

What two things can polarimetry measure?

enantiometric excess and optical purity 

45
New cards

What is circular dichroism? What is its effect and what problems result from it?

Chiral molecules absorb left- and right- circularly polarized light. One is absorbed more preferentially than the other, resulting in an ellipse. This makes these samples hard to measure directly. CD increases with path length and concentration.

46
New cards
<p>What does each variable mean?</p>

What does each variable mean?

theta = ellipticity

Al = absorbance of counterclockwise (left) light

Ar - absorbance of clockwise (right) light

b = pathlength (cm)

c = concentration (mol/L)

47
New cards

Draw a circular dichroism instrument. Explain how it works. 

The monochromater selects a specific wavelength for the light to absorb. The polarizer polarizes the light along a fixed direction (45°), preparing it for modulation. The photoelastic modulator rapidly alternates the polarization of light between left- and right-circular (or elliptical) polarization, essential for sensitive CD measurements. Finally, the light passes through the sample and its transmitted to the detector. 

<p class="my-2 [&amp;+p]:mt-4 [&amp;_strong:has(+br)]:inline-block [&amp;_strong:has(+br)]:pb-2">The monochromater selects a specific wavelength for the light to absorb. The polarizer polarizes the light along a fixed direction (45°), preparing it for modulation. The photoelastic modulator rapidly alternates the polarization of light between left- and right-circular (or elliptical) polarization, essential for sensitive CD measurements. Finally, the light passes through the sample and its transmitted to the detector.&nbsp;</p><p></p>
48
New cards

What is befringement?

Different refractive indices for vertical and horizonal polarization.

49
New cards

How does a photoelastic modulator work?

45 degree polarized light (in phase) contains vertical and horizontal polarization. One of these is delayed by ¼ wavelgnth resulting in circularly polarized light (out of phase). The PEM rapidly shifts birefringence (alternates the polarization of left- and right-circular polarization).

50
New cards

Advantages and Disadvantages of UV-CD

Advantages

  • fast

  • inexpensive

  • fairly sensitive

Disadvantages

  • limited structural information

51
New cards

Draw the general layout of a mass spectrometer.

knowt flashcard image
52
New cards

Steps of mass spec.

1) ionize molecule (typically -1/+1)

2) accelerate ion through electrical field

3) measure m/z

53
New cards

M/Z units

M = mass (daltons) 

Z = charge (multiples of proton charge) 

54
New cards
<p>What does each variable mean? What are both sides also equal to?</p>

What does each variable mean? What are both sides also equal to?

q = charge

V = voltage

m = mass

v = velocity

Both sides are equal to energy (E).

55
New cards
<p>What does each variable mean?</p>

What does each variable mean?

r = radius (of circular path)

m = mass

v = velocity

q = charge

B = magnetic field strength

56
New cards

What is the relationship between M/Z and electric/magnetic field?

When an ion enters an electric field, it is accelerated. When it enters a magnetic field, its motion is bent into a circle. Heavier ions bend less and therefore, have a larger radius and higher M/Z ratio. Smaller ions have small radii and lower M/Z. 

57
New cards

What is the mean free path, what does it effect, and how does it change with pressure?

Mean free path is the average distance an ion travels without collisions. It changes the velocity of the ions. An increase in pressure results in more things to collide with, decreasing the mean free path.

58
New cards
<p>What does each variable mean?</p>

What does each variable mean?

R = resolving power

m = m/z of an ion

change in m = the smallest difference in m/z the instrument can distinguish between two peaks

59
New cards

What is the significance of high resolution (resolving power) in mass spec?

The “exact mass” (molecular formula) can be determined and measurements can be made in the presence of nearby interferents. 

60
New cards

What is a magnetic sector and how does it work? Draw one. 

A magnetic sector is a type of mass analyzer. It uses a magnetic field to separate ions based on M/Z. The magnetic field bends the ions path creating a curvature. If the radius of curvature of the ion equals the radius of curvature of the instrument, the ion is detected. 

<p>A magnetic sector is a type of mass analyzer. It uses a magnetic field to separate ions based on M/Z. The magnetic field bends the ions path creating a curvature. If the radius of curvature of the ion equals the radius of curvature of the instrument, the ion is detected.&nbsp;</p>
61
New cards

How can one select a specific M/Z in mass spec? What precautions must be taken?

Keep either magnetic field strength or voltage constant while varying the other. However, varying voltage is not ideal because it effects the transmission and detection of the ions. Ideally, one would want to vary the magnetic field but this must be done slowly because it reflects a 10-fold change.

62
New cards

What is a double-focusing mass spec and why is it important? Draw one.

A double-focusing mass spec combines a magnetic and an electric sector that compensates for the initial conditions of ions. Before acceleration, ions have kinetic energy which causes changes in radius of curvature and initial direction. This results in peak broadening, limiting the resolving power. The electric sector eliminates ions with energy far from the desired. In combintaion with the magnetic sector, only one mass is focused at a time, increasing resolution.

<p>A double-focusing mass spec combines a magnetic and an electric sector that compensates for the initial conditions of ions. Before acceleration, ions have kinetic energy which causes changes in radius of curvature and initial direction. This results in peak broadening, limiting the resolving power. The electric sector eliminates ions with energy far from the desired. In combintaion with the magnetic sector, only one mass is focused at a time, increasing resolution. </p>
63
New cards

Nier-Johnson double focusing vs Mattauch-Herzog double focusing 

Nier-Johnson scans one M/Z at a time and the different sectors are not in the same plane. Mattauch-Herzog - all M/Z are scanned at once along the same plane, usually uses a spectograph. 

64
New cards

What is the field-free region length and how does it effect resolution?

The field-free region length (L) is the distance between the acceleration plates and the detector. The longer the length, the higher the resolution.

65
New cards

Describe time of flight spectroscopy.

The sample is ionized and accelerated through an electric field. The ions travel down the field-free region and reach the detector. The M/Z is measured by how long it takes the ions to reach the detector. 

66
New cards

What is a problem with time of flight spectroscopy and how can it be corrected?

The initial energy of the ions effects its total energy after acceleration, causing ions of the same M/Z to reach the detector at different times. A reflectron can be use to equate the energy of the ions. A reflectron contains electrostatic ion mirrors that reflect ions back towards the detector. Ions with greater energy will penetrate the reflectron deeper, slowing them down and synchronizing arrival times of the ions. Additionally, path length is also elongated.

67
New cards

How does quadrupole mass spectroscopy work?

Quadrupole MS filters ions along a path. Ions are injected along the center between the rods. The electric fields cause the ions to oscillate. The ions of stable oscillations make it through and the unstable oscillations will hit the rod and be filtered out.

68
New cards

What is a high pass filter in quadrupole MS?

Ions are pushed to the center by the DC field (constant voltage) and pulled towards the rods on a negative ½ cycle by the AC field. Light ions react quickly and strike the rods. Heavy ions react slowly so the AC direction switches before they can strike rod. Therefore, a high pass filter filters out light ions and heavy ions reach the detector. 

69
New cards

What is a low pass filter in quadrupole MS?

Ions are pulled towards the rods by the DC field (constant voltage) and pushed from the rods on a positive ½ cycle by the AC field. Light ions react quickly and are pulled away before they strike the ride.  Heavy ions react slowly so they strike the rod before the AC direction change. Therefore, a low pass filter filters out heavy ions and heavy and light ions reach the detector. 

70
New cards

Advantages and disadvantages of quadrupole MS

Advantages

  • high signal due to large apertures

  • can operate at high pressure due to short pathlength

  • fast scanning due to short pathlength and electric field can be switched quickly

  • fast and sensitive

Disadvantages

  • moderate resolution

71
New cards

How does ion trap mass spectroscopy work?

Ions are stored in place by AC electric field. Ions of stable oscillation will stay and ions of unstable oscillation will escape. Increasing the AC voltage ejects the unstable ions one by one in order of M/Z. 

72
New cards

What is the space charge effect and why is it important?

When multiple ions are held in one place, their charges effect one another. This limits trap capacity and effects resolution. 

73
New cards

How does FTICR work?

Ions enter an magnetic field and move in a circular orbit with lighter ions moving faster and creating a higher frequency orbit. The motion of the ions create a current (AC), outputting various frequencies. These frequencies are converted to M/Z via FT.

74
New cards

What is acquisition time and why is it important?

The amount of time the instrument takes recording the ion signal after excitation. Important because the longer the acquisition time, the higher the resolution. 

75
New cards

How does orbitrap mass spec work and how is it different/similar to FTICR?

Ions are trapped electrostatically (unlike FTICR - magnetically) and oscillate left and right depending on M/Z. The induced current on the outer elctrodes are transmitted and converted to M/Z using FT (like FTICR). 

76
New cards

Rank the mass spec analyzers from highest resolving power the lowest. 

FT-ICR > orbitrap = double focusing > TOF > ion trap > mag. sector = quad.

77
New cards

Rank the mass spec analyzers from highest mass range to lowest.

TOF > FT-ICR > orbitrap = double-foc. = ion trap > quad.

78
New cards

Rank the mass spec analyzers from highest dynamic range to lowest.

double-foc > quad > TOF = ion trap = FTICR = orbitrap

79
New cards

Rank the mass spec analyzers from highest cost to lowest. 

FT-ICR > orbitrap > double-foc. >TOF > ion trap = quad 

80
New cards

Rank the mass spec analyzers from fastest to slowest.

TOF > double-foc (spectograph) > quad. = ion trap > double-foc. = FT-ICR = orbitrap

81
New cards

Rank the mass spec analyzers from highest pressure tolerance to lowest.

ion trap > quad > double-foc. = TOF > FTICR

82
New cards

Rank the mass spec analyzers from largest to smallest.

double-foc = FT-ICR > TOF = orbitrap > quad = ion trap 

83
New cards

What is ICP and how does it work? What are its advantages and disadvantages?

ICP (inductively coupled plasma) is a type of mass spec ion source. The sample is typically introduced through nebulization, vapor generation, or laser ablation. A high temperature plasma (ionized argon gas) breaks down the sample into constituent atoms.

Advantage

  • low LOD (ppt)

Disadvantage

  • sample is completely atomized - loses structural/molecular information

84
New cards

What is glow discharge ionization and how does it work?

Type of mass spec ion source used for solid samples. An electric current is run through argon gas in which the argon is ionized and a plasma forms. The argon ions charge toward the solid (sputtering) to move the solid into the gas phase. The gas particles then enter the plasma and are ionized.

85
New cards

What is thermal ionization and how does it work? What mass analyzer does is it usually paired with? What are some of its advantages and disadvantages?

Type of mass spec ion source. The sample is placed on a filament and an electrical current is passed through. The sample is vaporized and some of the atoms are ionized. Usually paired with a magnetic sector.

Advantages

  • high resolution

  • better precision and accuracy for isotope ratio

  • inexpensive

Disadvantages

  • not applicable to all elements

  • slow

86
New cards

What is electron-impact ionization and how does it work? What mass analyzer does it usually use? What are some advantages and disadvantages of it?

Type of mass spec ion source. A heated filament emits high-energy electron and incoming gas-phase molecules. This results in the ejection of some electrons from the molecules, creating an unstable ion, leading to fragmentation. Usually paired with a quadrupole.

Advantage

  • fragmentation can help identify structure of molecule beyond molecular weight

Disadvantage

  • limited mass range (requires volatile analyte)

87
New cards

What is fragmentation and why is it important?

Fragmentation occurs when unstable ions break apart into smaller ions. Important because it reveals structural information beyond just molecular weight.

88
New cards

What is chemical ionization and how does it work? What is an advantage and disadvantage? 

Type of mass spec ion source. Reagent gas (CH4, NH3) fills ion chamber. Electrons collide with gas and gas transfers proton to analyte. 

Advantage 

  • soft ionization = reduced fragmentation = preserves analyte 

Disadvantage 

  • limited structural information 

89
New cards

What is matrix-assisted laser desorption ionization (MALDI) and how does it work? What mass analyzer is it used with? What are some of its advantages and disadvantages?

Type of mass spec ion source. Sample is deposited in matrix and a laser strikes the matrix. Energy is transferred from matrix to analyte, donating a charge (usually proton) to analyte. Almost always used with time of flight mass analyzer.

Advantages

  • reduced fragmentation

  • capable of ionizing and vaporizing high-mass molecules (proteins)

  • ion production is pulsed

  • infinite mass range

Disadvantage

  • limited structural information

90
New cards

What is electron-spray ionization and how does it work. What are some of its advantages?

Type of mass spec ion source. Sample is introduced in solution. The solution flows through metal capillary tube and voltage is applied, creating a spray of droplets holding charge. The solvent evaporates and the droplets shrink until the droplet breaks apart leaving the multiply-charged analyte.

Advantages

  • efficient for making measuring proteins using quadrupole because M/Z is so low (high mass, high charge)

  • redundant measurements = high mass accuracy

91
New cards

What is desorption electron-spray ionization and how does it work?

Type of mass spec ion source. Charged particles are sprayed onto sample surface. The molecules on surface are desorbed (lifted off) and enter gas phase. In the process, the molecules pick up charge from the droplets. 

92
New cards

What is the faraday cup? What mass analyzers are commonly used with it? What are its advantages and disadvantages?

A type of mass spec detector. Commonly used with magnetic sector and double-focusing

Advantages

  • signal only depends on charge

  • not affected by magnetic field

Disadvantage

  • slow

93
New cards

What is an electron multiplier and how does it work? What mass analyzer is it commonly used with? What are its advantages and disadvantages? 

A type of mass spec detector. It uses dynodes to amplify the electrons. Often used with quadrupole. 

Advantages 

  • fast

  • provides gain (signal amplification, sensitivity) 

Disadvantages 

  • uneven response (mass and size-sensitive) 

94
New cards

What are type of mass analyzers are used with array detectors?

Spectograph-type mass analyzers (TOF, mattauch-herzog double-focusing)

95
New cards

How does chromotography work?

An analyte is separated through interactions between the mobile and stationary phases.

96
New cards
<p>What does each variable mean?</p>

What does each variable mean?

N = # of theoretical plates 

L = length of column (cm) 

H = height of theoretical plates (cm)

97
New cards

Significance of theoretical plates

Increase in theoretical plates increases the column’s efficiency 

98
New cards
<p>What does each variable mean?</p>

What does each variable mean?

N = # of plates

TR = retention time

W = width of peak

W1/2 = full width at half max (FWHM)

99
New cards

What does retention factor measure?

How strongly analyte is retained by stationary phase

100
New cards
<p>What does each variable mean?</p>

What does each variable mean?

k = retention factor 

tR = retention time (time in mobile phase and stationary phase) 

tM = time in mobile phase