1/43
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
ADC
converts analog info to digital, making the analog info useful to the computer
used for image processing and reconstruction in CT
steps of ADC
scanning
sampling
quantization
scanning
For the picture to be digitized it is separated into small regions on a grid (matrix)
hounsfield unit
pixel
sampling
brightness of each pixel is measured
quantization
brightness value of the pixel is assigned
scan field of view
area to be imaged within the gantry
wherein raw data is collected
smaller
better: larger of smaller SFOV?
SFOV
determines the number of detectors collecting data per scan
reconstruction or display field of view
determines how much of the collected raw data is used to create an image
DFOV
equal to or less that the size of the SFOV
matrix
the computerized grid of data onto which the slice is placed for the purpose of reconstructing the density info
512×512
1024×1024
common CT matrix sizes
over the SFOV
where is the CT matrix placed
decreased
smaller matrix: increase/ decrease noise
smaller matrix
larger pixels
decreased spatial resolution
decrease noise
larger matrix
smaller pixels
increased spatial resolution
increase noise
pixel size (d)
= FOV / matrix
voxels - volume element
tissue volume
determined by multiplying the pixel size by the thickness of the CT image slice
voxel
(pixel size)² x slice thickness =
hounsfield units
the levels of brightness displayed on the monitor for each pixel
0
water CT number
1000+
dense bone CT number
-1000
air CT number
CT number
µt - µw X 1000µw =
µw
attenuation coefficient of water
µt
attenuation coefficient of tissue
higher CT numbers
light shades of gray/ white
lower CT numbers
darker shades of gray/ black
attenuation
reduction i the intensity of the beam a sit passes through an object
windowing
adjusting the gray scale
256
monitor can detect how many shades of gray
window level (wl)
HU value of the area of interest
center of range
window width (ww)
range of grays (CT numbers)
determines the number of HUs represented on an image
white
numbers above range will be?
black
numbers below range will be?
lower contrast
larger window width
high contrast
smaller window width
calculate CT numbers HU
divide widow width in half
subtract the quotient form the window level for lower limit
add the quotient to the window level to determine upper limit
slice thickness
the width of the volume of tissue being examined (voxel)
anatomical coverage
noise in image
patient dose
what does slice thickness affect?
slice thickness
thickness of the x-ray beam / slices per rotation =
magnification
permit viewer to enlarge part of the reconstructed image on the monitor for viewing or filming
mag
allows for easier examination of image
enlarges the individual pixels
post processing done to the image not to raw data
“zoom in”