400-700 nm
visible spectrum
<400 nm
UV region
>700 nm
infrared region
wavelength inversely proportional to frequency
frequency directly proportional to energy
rel. between wavelength, frequency, and energy

ROYGBV
MIUXGC
longest wavelength → shortest wavelength in electromagnetic spectrum
spectrophotometry
measure of light transmitted by a solution
light / radiant source — source of polychromatic light
entrance slit — minimizes stray light
monochromator — isolates specific wavelength of light
exit slit — controls bandwidth / bandpass
cuvet / absorption cell / analytical cell / sample cell — holds the soln.
photodector — detects and converts transmitted light → electricity
meter / read out device → displays output
parts of spectrophotometry
Tungsten light bulb/incandescent tungsten/tungsten iodide lamp
most commonly used light / radiant source
hollow cathode lamp
light / radiant source in AAS
deuterium discharge lamp and the mercury arc lamp
most common UV light / radiant source
stray light
most common cause of loss of linearity
second order spectra
Deteriorated optics
Light dispersed by a darkened enveloped
Extraneous room light
most common cause of stray light
absorbance error
major effect of stray light
sharp cutoff filter
filter best for measuring stray light
Neutral Density filter and Dichromate sol'n
verify absorbance accuracy on linearity
grating
most commonly used monochromator
prism
produces purest light in UV range
Didymium or holmium oxide filter
mercury vapor lamp
used for quality control of monochromator
square cuvette
cuvette has more advantage
discard
scratched optical surfaces in cuvette → ???
alumina silica
most commonly used cuvet
glass cuvets
cuvets for visible range BUT absorbs UV
quartz
cuvets for UV radiation
Beer’s law
states that unknown subs.
directly proportional to absorbed light
inversely proportional to transmitted light
Cu = Au/As x CS
Cu = conc. of unknown
Au = absorbance of uknown
As = absorbance of standard
Cs = conc. of unknown
concentration of unknown: Beer’s law
Lambert law
absorbance increases exponentially w/ an increase in light path (direct relationship)
photomultiplier tube
most common type of photodetector → excellent sensitivity (never expose to light → burn out)
barrier layer cell / photocell / photovoltaic cell
simplest detector (requires no external voltage source)
photodiode
photodetector that respond to a specific wavelength → UV visible
phototube
Same as photocell, differs in that an outside voltage is required for operatio
reagent blank
blanking technique → corrects for absorbance caused by COLOR OF THE REAGENT
0 Abs (100 transmission)
in reagent blank technique, adjust spectro to ???
sample blank
blanking technique → corrects for interfering chromogens / lipemia
ultracentrifuge
used to clear serum or plasma chylomicrons
sample blank
blanking technique → subtract intrinsic absorbance of the sample caused by hemolysis, icterus, turbidity, drug interference
saline (sample + saline)
sample blank: substitute ___ for reagent
single beam spectrophotometer
simplest type of spectrophotometer: 1 measurement at a time
double beam spectrophotometer
spectrophotometer permit automatic correction of sample & reference absorbance
double beam in space
double beam spectrophotometer: use 2 photodetectors
double beam in time
double beam spectrophotometer: one photodector & alternately passed the monochromatic light using chopper