Subgroup IV: Instrument Operations and Quality Control
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai | Chat |
|---|
No analytics yet
Send a link to your students to track their progress
What tests are run on a dose calibrator?
constancy, accuracy, linearity, geometry
How often is constancy run on a dose calibrator?
daily
How often is linearity run on a dose calibrator?
quarterly
How often is accuracy run on a dose calibrator?
annually
How often is geometry run on a dose calibrator?
installation and after repair/movement
dose calibrator constancy definition
today’s readings fall within an acceptable range compared to prior readings
sources used for dose calibrator constancy
Co-57, Ba-133, Cs-137
dose calibrator constancy acceptable results
±10%
What action should you take if the dose calibrator constancy is out of range?
if consistent, explore sources of error; consider repair or replacement
dose calibrator accuracy purpose
does the dose calibrator accurately measure the amount of activity of the radioactive material placed in the chamber
What sources can be used for dose calibrator accuracy?
Co-57, Ba-133, Cs-137 (must be NIST-traceable)
dose calibrator accuracy procedure
need at least two reference sources of different gamma ray energies
decay-correct each source from the date of calibration to the date of the test
measure each source three times, take the average reading
express the results as % error relative to the expected activity
dose calibrator accuracy acceptable results
±10%
What actions should you take if the dose calibrator accuracy falls outside of the acceptable range?
repair or replace
dose calibrator linearity purpose
verify that the dose calibrator responds to varying activities in a linear fashion
What activity range should be tested for a dose calibrator linearity test?
highest amount of activity ever present in department (~50-100 mCi in a hospital) down to 30 uCi
What source is used for a dose calibrator linearity test?
Tc99m
dose calibrator linearity procedure methods
multi-day decay method, 30 hour reading is the standard
Calicheck or Lineator method (sleeve method)
dose calibrator linearity acceptable results
±10%
What actions should you take if the dose calibrator linearity falls outside of the acceptable range?
if deviation from expected values is seen in a particular range, use a correction factor for that range
dose calibrator geometry purpose
calculate a correction factor to correct for the dose calibrator’s dependence on sample geometry
What source is used for dose calibrator geometry?
Tc99m
dose calibrator geometry procedure
25 mCi in 0.5 mL in each container type, take reading
repeat with 1 mL, 2 mL, 3 mL, 4 mL, 5mL
middle volume reading is standard (3 mL for a 5 mL syringe)
calculate a correction factor
dose calibrator geometry correction factor formula
CF = (activity at any volume) / (activity at 3 mL)

dose calibrator geometry acceptable results
±10%
What actions should you take if the dose calibrator geometry falls outside of the acceptable range?
determine a correction factor that can be applied to a particular radionuclide or measurement situation that comes up frequently

Given the following data, does this dose calibrator pass the linearity test?
yes

Given the following data, does this dose calibrator pass the linearity test?
no

Given the following data, does this dose calibrator pass the geometry test?
yes