The digital image acquisition system is best described as which combination?

Image receptor & processing computer

2
New cards

Which characteristic describes the ability to detect small amounts of radiation?

Sensitivity

3
New cards

A more sensitive imaging system generally requires what?

Less exposure

4
New cards

Dynamic range is the range of what values available for image formation?

Pixel values

5
New cards

Which process can adjust the default dynamic range?

Equalization (DRC)

6
New cards

Spatial resolution is the ability to display small pixels as what?

Sharp image detail

7
New cards

Contrast resolution describes the ability to distinguish what?

Different tissues

8
New cards

In a CR reader, sampling frequency increases when the distance between samples does what?

Decreases

9
New cards

In a DR detector, sampling frequency depends mainly on what?

Del Pitch

10
New cards

Del pitch is measured from the center of one del to what point?

The center of an adjacent del

11
New cards

What happens to spatial resolution as pixel or del size becomes smaller?

It increases

12
New cards

What is the minimum number of pixels needed to form one image detail?

Two

13
New cards

The number of resolvable details is approximately what fraction of the available pixels?

One-half

14
New cards

Which unit is commonly used to measure spatial frequency?

Line-pairs per millimeter

15
New cards

A pixel width of 0.1mm provides a maximum spatial frequency of about what?

5 LP/mm

16
New cards

A line pair contains which two components?

One light and one dark line

17
New cards

Which formula estimates maximum spatial frequency from pixel size?

1/(2xPixel Size)

18
New cards

For a hardware matrix, del or hardware-pixel size is generally what?

Fixed

19
New cards

Why is inherent DR detector sharpness consistent?

Del size is fixed

20
New cards

Typical DR del sizes in the presentation range from approximately what?

100-200 microns

21
New cards

A 100-micron del produces a spatial frequency of about what?

5 LP/mm

22
New cards

For the same model of LCD monitor, a larger screen generally has what?

More hardware pixels of the same size

23
New cards

Collimating a DR exposure changes the displayed field of view but does not change what?

Del size

24
New cards

A soft matrix differs from a hardware matrix because its pixel size can be what?

Variable

25
New cards

In a CR reader, changing the laser-beam width can change what?

Pixel size

26
New cards

For a fixed image area, increasing matrix size causes soft pixels to become what?

Smaller

27
New cards

Excessive monitor magnification can make the image appear what?

Pixelated

28
New cards

As more zoom is applied to a displayed image, sharpness generally does what?

Decreases

29
New cards

A fixed-matrix CR system uses what across diffferent IR sizes?

The same number of pixels per row

30
New cards

In a fixed-matrix CR system, which IR size generally provides better sharpness?

Smaller IR

31
New cards

Most modern CR systems maintain consistent sharpness by using what approach?

Fixed sampling

32
New cards

The soft-pixel size formula is FOV divided by what?

Matrix dimension

33
New cards

Ultimately, sharpness is determined mainly by the size of what?

Dels or pixels

34
New cards

CR and indirect-conversion DR phosphors first convert x-rays into what?

Light

35
New cards

Which efficiency describes how well a phosphor captures incoming x-rays?

Absorption efficiency

36
New cards

A reflective backing in a CR plate primarily improves which efficiency?

Emission efficiency

37
New cards

Compared with direct-conversion DR, indirect-conversion DR generally requires what?

Less technique

38
New cards

Why does the direct-conversion DR generally provide higher spatial resolution?

It avoids light dispersion

39
New cards

Which receptor type has the highest typical DQE listed at 70 kV?

Indirect-Conversion DR

40
New cards

What does a 75% fill factor mean?

75% of the del detects radiation

41
New cards

A higher fill factor generally improves contrast resolution and what?

Signal-to-noise ratio

42
New cards

If del size decreases while TFT and capacitor size stay fixed, fill factor does what?

Decreases

43
New cards

The Nyquist theorem says sampling frequency must be at least how many times image frequency?

Two times

44
New cards

Insufficient sampling of high-frequency signals produce which artifacts?

Aliasing or Moire patterns

45
New cards

Which situation can create aliasing?

Using a stationary low-frequency grid

46
New cards

Which action helps prevent grid aliasing?

Use a higher-frequency grid

47
New cards

Which artifact is more commonly associated with DR than CR?

Electronic mottle

48
New cards

A ghost image on a CR plate most likely results from what?

Incomplete erasing

49
New cards

An artifact appearing on images from several different CR cassettes most likely comes from what?

The CR reader

50
New cards

Dust on CR reader optics can delete an entire line, causing what?

Line dropout