spatial resolution
ability of the system to define small objects
low contrast resolution
ability of the system to differentiate objects of similar densities
temporal resolution
speed at which data can be acquired
what are the 3 elements for helical scanning?
constant xray output
constant table movement
constant xray tube rotation
helicla scanning is sometimes referred to as…
volumetric scanning
the newer 256 and 320 detector systems with detector widths of …
0.625 and .5 mm
describe it: describe adaptive, nonuniform array, hybrid array
variable thinner rows centrally and wider rows peripherally
MDCT provides longer and faster x coverage per gantry rotation
z-axis
MDCT systems have increased the x of the gantry, which further increases x per unit time
speed / volume coverage
binning
the action of combining slices in various combinations
what is the advantage of the dual CT
it halved the temporal resolution (data acquisition time) because each xray tube and detector array configuration only needs to rotate half of the angle
MDCT designs caused what downfall
fourth generation systems of complete ring. it would be too expensive because of the amount of detectors needed to perform how a MDCT would
what is the dual CT scanner called produced by siemens
Somatom Definition CT Scanner
what is the primary goal of the dual scanner
increase scan speed — to have the least amount of motion artifacts, specifically involuntary movement
what is unique about the fourth generation conventional CT
400 degree rotation
axial scans are typically used in protocols in which acquisition is not a major concern and x such as internal auditory canal
optimal resolution is required